Image-Quality Perception as a Function of Dose in Digital Radiography

被引:6
|
作者
Lehnert, Thomas [1 ]
Naguib, Nagy N. N. [1 ]
Korkusuz, Huedayi [1 ]
Bauer, Ralf W. [1 ]
Kerl, J. Matthias [1 ]
Mack, Martin G. [1 ]
Vogl, Thomas J. [1 ]
机构
[1] Clin Johann Wolfgang Goethe Univ Frankfurt, Dept Diagnost & Intervent Radiol, D-60590 Frankfurt, Germany
关键词
digital radiography; filtration; image quality; musculoskeletal imaging; skin entrance dose; RADIATION;
D O I
10.2214/AJR.10.6269
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this article is to determine the degree to which the skin entrance dose could be lowered, by adjusting exposure parameters and filtration, and the subsequent effect on readers' confidence levels of digital radiographs. MATERIALS AND METHODS. The study was prospectively performed on a cadaver. Digital radiographs of bones were evaluated and scored on a 9-point scale separately by four radiologists who were blinded to the types of filtration and doses used. The study entailed three phases: phase 1, random dose and filter; phase 2, fixed filter and varying radiation doses (100%, 75%, 50%, and 25% of the standard recommended dose); and phase 3, fixed dose and varying filtration (no filtration, aluminum filter, and aluminum-copper filter). Skin entrance dose was measured using a dosimeter placed on the skin. Differences in scores were tested using a Friedman test. RESULTS. The mean scores given to images with 100%, 75%, 50%, and 25% of the recommended standard dose were 6.18, 6.1, 5.11, and 4.07, respectively. No significant difference was noted between 100%- and 75%-dose images (p = 0.1). A significant difference (p < 0.0001) was noted when we compared the 100%- and 75%-dose images with the 50%- and 25%-dose images. The mean scores given for no filtration, aluminum filtration, and aluminum-copper filtration were 5.67, 5.43, and 5.18, respectively. No significant difference between no filtration and aluminum filtration (p = 0.411) was noted. A significant difference was detected between no filtration and aluminum-copper filtration (p = 0.012). The combination of an aluminum filter and a 75% standard dose achieved a 31.1% reduction in skin entrance dose. CONCLUSION. It is possible to achieve a 31.1% reduction in skin entrance dose for imaging bony structures by using 75% of the standard dose and aluminum filtration without significantly affecting image quality.
引用
收藏
页码:1399 / 1403
页数:5
相关论文
共 50 条
  • [41] Digital chest radiography: an update on modern technology, dose containment and control of image quality
    Schaefer-Prokop, Cornelia
    Neitzel, Ulrich
    Venema, Henk W.
    Uffmann, Martin
    Prokop, Mathias
    EUROPEAN RADIOLOGY, 2008, 18 (09) : 1818 - 1830
  • [42] Digital chest radiography: an update on modern technology, dose containment and control of image quality
    Cornelia Schaefer-Prokop
    Ulrich Neitzel
    Henk W. Venema
    Martin Uffmann
    Mathias Prokop
    European Radiology, 2008, 18 : 1818 - 1830
  • [43] Digital radiography: optimization of image quality and dose using multi-frequency software
    Precht, H.
    Gerke, O.
    Rosendahl, K.
    Tingberg, A.
    Waaler, D.
    PEDIATRIC RADIOLOGY, 2012, 42 (09) : 1112 - 1118
  • [44] Towards image quality, beam energy and effective dose optimisation in digital thoracic radiography
    J. H. Launders
    A. R. Cowen
    R. F. Bury
    P. Hawkridge
    European Radiology, 2001, 11 : 870 - 875
  • [45] Dose and image quality optimization in neonatal radiography
    Dougeni, E. D.
    Delis, H. B.
    Karatza, A. A.
    Kalogeropoulou, C. P.
    Skiadopoulos, S. G.
    Mantagos, S. P.
    Panayiotakis, G. S.
    BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (958): : 807 - 815
  • [46] The practice of chest radiography using different digital imaging systems: Dose and image quality
    Moey S.-F.
    Shazli Z.A.
    Sayed I.S.
    Iranian Journal of Medical Physics, 2018, 15 (02) : 101 - 107
  • [47] Digital radiography: optimization of image quality and dose using multi-frequency software
    H. Precht
    O. Gerke
    K. Rosendahl
    A. Tingberg
    D. Waaler
    Pediatric Radiology, 2012, 42 : 1112 - 1118
  • [48] Dose and perceived image quality in chest radiography
    Veldkamp, Wouter J. H.
    Kroft, Lucia J. M.
    Geleijns, Jacob
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (02) : 209 - 217
  • [49] IMAGE QUALITY CRITERIA APPLIED TO DIGITAL RADIOGRAPHY
    BUSCH, HP
    FAULKNER, K
    MALONE, JF
    RADIATION PROTECTION DOSIMETRY, 1995, 57 (1-4) : 139 - 140
  • [50] Slot-Scan Digital Radiography of the Lower Extremities: a Comparison to Computed Radiography with Respect to Image Quality and Radiation Dose
    Lee, Kwang Hwi
    Kwon, Jong Won
    Yoon, Young Cheol
    Choi, Sang-Hee
    Jung, Jee Young
    Kim, Ji Hye
    Lee, Sang Jun
    KOREAN JOURNAL OF RADIOLOGY, 2009, 10 (01) : 51 - 57