Experimental Investigation of the Spiral Structure of a Magnetic Capsule Endoscope

被引:2
|
作者
Yang, Wanan [1 ]
Li, Zhen [2 ]
He, Yong [1 ]
Dai, Houde [3 ]
Wu, Zhouxing [3 ]
机构
[1] Yibin Univ, Sch Comp & Informat Engn, Yibin, Peoples R China
[2] CNPC, Geophys Prospecting Co, Chuanqing Drilling Engn Co Ltd, Beijing, Peoples R China
[3] Chinese Acad Sci, Fujian Inst Res Struct Matter, Quanzhou Inst Equipment Mfg, Beijing 100864, Peoples R China
基金
中国国家自然科学基金;
关键词
Wireless Capsule Endoscope; Magnet Ring; Rotational Magnetic Field; Spiral Line; LOCOMOTION; DEVICES;
D O I
10.5772/63930
中图分类号
TP24 [机器人技术];
学科分类号
080202 ; 1405 ;
摘要
Fitting a wireless capsule endoscope (WCE) with a navigation feature can maximize its functional benefits. The rotation of a spiral-type capsule can be converted to translational motion. The study investigated how the spiral structure and rotational speed affected the capsule's translation speed. A hand-held instrument, including two permanent magnets, a stepper motor, a controller and a power supplier, were designed to generate rotational magnetic fields. The surfaces of custom-built permanent magnet rings magnetized radially were mounted in spiral lines with different lead angles and diameters, acting as mock-up capsules. The experimental results demonstrate that the rotational speed of the magnetic field and the spiral have significant effects on the translational speed of a capsule. The spiral line with a larger lead angle and the rotating magnetic field with a higher speed can change the capsule's rotation into a translational motion more efficiently in the intestine.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Predictive Tilt Compensation for Robot Assisted Magnetic Capsule Endoscope
    Mahmood, Salman
    Schurr, Marc O.
    Schostek, Sebastian
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 3697 - 3702
  • [32] The software implementation of magnetic localization and orientation detection for capsule endoscope
    Ma, Tongxing
    Hu, Chao
    Meng, Max Q. -H.
    2007 IEEE/ICME INTERNATIONAL CONFERENCE ON COMPLEX MEDICAL ENGINEERING, VOLS 1-4, 2007, : 35 - 38
  • [33] A Locomotion Mechanism with External Magnetic Guidance for Active Capsule Endoscope
    Wang, Xiaona
    Meng, Max Q. -H.
    Chen, Xijun
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 4375 - 4378
  • [35] A magnetic stereo-actuation mechanism for active capsule endoscope
    Wang, Xiaona
    Meng, Max Q. -H.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 2811 - 2814
  • [36] Fabrication of Magnetic Microactuators for Cytology Brush Built into Capsule Endoscope
    Yamashita, Maki
    Honda, Takashi
    ELECTRONICS AND COMMUNICATIONS IN JAPAN, 2015, 98 (04) : 27 - 34
  • [37] Small Antenna with Magnetic Sheet for Wireless Capsule Endoscope Communication
    Yokoyama, Yutaro
    Saito, Kazuyuki
    Peng, Yunxiao
    Ito, Koichi
    2020 INTERNATIONAL WORKSHOP ON ANTENNA TECHNOLOGY (IWAT), 2020,
  • [38] Design of a Reconfigurable Sensor Array for Localization of Magnetic Capsule Endoscope
    Liu, Hongxi
    Su, Jingran
    Song, Shuang
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2025, 74
  • [39] An Improved Magnetic Localization and Orientation Algorithm for Wireless Capsule Endoscope
    Hu, Chao
    Yang, Wanan
    Chen, Dongmei
    Meng, Max Q. -H.
    Dai, Houde
    2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 2055 - 2058
  • [40] Research on Sensors Arrangement for Capsule Endoscope Microrobot Magnetic Localization
    Xu, Jiansheng
    2013 25TH CHINESE CONTROL AND DECISION CONFERENCE (CCDC), 2013, : 5208 - 5211