Comparing the Netra smartphone refractor to subjective refraction

被引:14
|
作者
Tousignant, Benoit [1 ]
Garceau, Marie-Christine [1 ]
Bouffard-Saint-Pierre, Nikola [1 ]
Bellemare, Marie-Maxime [1 ]
Hanssens, Jean-Marie [1 ]
机构
[1] Univ Montreal, Sch Optometry, Montreal, PQ, Canada
关键词
diagnostic techniques; instrumentation; ocular; ophthalmological instrumentation; refractive errors; telemedicine; vision tests; AUTOREFRACTOR; IMPAIRMENT; MAGNITUDE; SCREENER; ERRORS;
D O I
10.1111/cxo.13003
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Among technologies targeting mobile eye care, EyeNetra is a smartphone-based subjective refraction system. This study compared the results from this system with those of professional subjective refraction. Participant visual comfort and preference of results were also measured. Methods Thirty-six optometry-naive participants (n = 36 eyes, aged 18-35 years), were randomly subjected to three refraction methods: professional subjective refraction, unassisted Netra (participants alone) and refined Netra (sphere results refined by a practitioner). Using a randomised, double-blind design, refraction results were mounted in a trial frame and distance logMAR visual acuities were measured. Subjective appreciation and visual comfort were assessed by questionnaire. Overall preference was ranked. Results Unassisted Netra yielded a median myopic overcorrection of 0.60 D (interquartile range [IQR] 0.25 to 0.94) compared to professional subjective refraction. Median equivalent sphere with unassisted Netra (-1.40 D, IQR -3.10 to -0.90) was significantly more myopic than refined Netra (-0.70 D, IQR -1.60 to -0.30) and then subjective refraction (-0.80 D, IQR -1.60 to -0.30) (all p-values < 0.01). Median visual acuity with professional subjective refraction (-0.16, IQR -0.22 to -0.09) was superior than unassisted Netra (-0.08, IQR -0.20 to 0.03) (p < 0.01). Subjective refraction was ranked first in preference of trial framed results by 72 per cent of participants; median preference rank favoured professional subjective refraction to both Netra results (all p < 0.01). For all questionnaire items, visual comfort was higher with subjective refraction than with unassisted Netra (all p < 0.04). Conclusion The Netra device - especially when used without professional assistance and compared to subjective refraction - induces significant myopic overcorrection and lower levels of visual acuity, subjective preference and visual comfort.
引用
收藏
页码:501 / 506
页数:6
相关论文
共 50 条
  • [41] Subjective Refraction in Eyes With Multifocal IOLs
    Mohammadi, S-Farzad
    Rahman-A, Nazanin
    Mazouri, Arash
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (03) : 161 - 161
  • [42] PHYSIOLOGICAL INDEXES FOR AUTOMATED SUBJECTIVE REFRACTION
    HUNG, G
    MARG, E
    [J]. AMERICAN JOURNAL OF OPTOMETRY AND ARCHIVES OF AMERICAN ACADEMY OF OPTOMETRY, 1972, 49 (10): : 886 - &
  • [43] Subjective refraction - binocular alignment.
    Krause, Kunibert
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2010, 227 (07) : R93 - R106
  • [44] Assessment of objective and subjective eccentric refraction
    Lundström, L
    Gustafsson, J
    Svensson, I
    Unsbo, P
    [J]. OPTOMETRY AND VISION SCIENCE, 2005, 82 (04) : 298 - 306
  • [45] Five days without a smartphone Smartphone use and subjective well-being: Results from a study comparing normal use with a deprivation condition
    Kohout, Susann
    Schumann, Christina
    [J]. SCM STUDIES IN COMMUNICATION AND MEDIA, 2020, 9 (03): : 394 - 420
  • [46] Comparison Between Aberrometry-Based Binocular Refraction and Subjective Refraction
    Carracedo, Gonzalo
    Carpena-Torres, Carlos
    Serramito, Maria
    Batres-Valderas, Laura
    Gonzalez-Bergaz, Anahi
    [J]. TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2018, 7 (04):
  • [47] Evaluation of an auto-refractor for over-refraction with multifocal contact lenses patients
    Giner, Anna
    Aldaba, Mikel
    Arjona, Montserrat
    Pujol, Jaume
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [48] Correlation of Nidek OPD-scan objective refraction with subjective refraction
    Yeung, IYL
    Mantry, S
    Cunliffe, IA
    Benson, MT
    Shah, S
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2004, 20 (05) : S734 - S736
  • [49] The Reliability and Acceptability of RDx-Based Tele- Controlled Subjective Refraction Compared with Traditional Subjective Refraction
    Huang, Jie
    Li, Xiaoning
    Yan, Tao
    Wen, Longbo
    Pan, Lun
    Yang, Zhikuan
    [J]. TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2022, 11 (01):
  • [50] Agreement and Repeatability of Central and Peripheral Refraction by One Novel Multispectral-Based Refractor
    Lu, Weicong
    Ji, Rongyuan
    Ding, Wenzhi
    Tian, Yuyin
    Long, Keli
    Guo, Zhen
    Leng, Lin
    [J]. FRONTIERS IN MEDICINE, 2021, 8