Three dimensional subsurface elemental identification of minerals using confocal micro-X-ray fluorescence and micro-X-ray computed tomography

被引:16
|
作者
Cordes, Nikolaus L. [1 ]
Seshadri, Srivatsan [2 ]
Havrilla, George J. [3 ]
Yuan, Xiaoli [4 ]
Feser, Michael [2 ]
Patterson, Brian M. [1 ]
机构
[1] Los Alamos Natl Lab, Div Mat Sci & Technol, Polymers & Coatings Grp, Los Alamos, NM 87545 USA
[2] Carl Zeiss Xray Microscopy Inc, Pleasanton, CA 94588 USA
[3] Los Alamos Natl Lab, Div Chem, Los Alamos, NM 87545 USA
[4] Univ Queensland, Julius Kruttschnitt Mineral Res Ctr, Brisbane, Qld 4068, Australia
基金
美国国家科学基金会;
关键词
Confocal micro-X-ray fluorescence; Micro-X-ray computed tomography; Subsurface particle analysis; PLASMA-MASS SPECTROMETRY; OBJECTS; QUANTIFICATION; SYSTEMS; COPPER; LAYERS; XRF;
D O I
10.1016/j.sab.2014.12.006
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Current non-destructive elemental characterization methods, such as scanning electron microscopy-based energy dispersive spectroscopy (SEM-EDS) and micro-X-ray fluorescence spectroscopy (MXRF), are limited to either elemental identification at the surface (SEM-EDS) or suffer from an inability to discriminate between surface or depth information (MXRF). Thus, a non-destructive elemental characterization of individual embedded particles beneath the surface is impossible with either of these techniques. This limitation can be overcome by using laboratory-based 3D confocal micro-X-ray fluorescence spectroscopy (confocal MXRF). This technique utilizes focusing optics on the X-ray source and detector which allows for spatial discrimination in all three dimensions. However, the voxel-by-voxel serial acquisition of a 3D elemental scan can be very time-intensive (similar to 1 to 4 weeks) if it is necessary to locate individual embedded particles of interest. As an example, if each point takes a 5 s measurement time, a small volume of 50 x 50 x 50 pixels leads to an acquisition time of approximately 174 h, not including sample stage movement time. Initially screening the samples for particles of interest using micro-X-ray computed tomography (micro-CT) can significantly reduce the time required to spatially locate these particles. Once located, these individual particles can be elementally characterized with confocal MXRF. Herein, we report the elemental identification of high atomic number surface and subsurface particles embedded in a mineralogical matrix by coupling micro-CT and confocal MXRF. Synergistically, these two X-ray based techniques first rapidly locate and then elementally identify individual subsurface particles. Published by Elsevier B.V.
引用
收藏
页码:144 / 154
页数:11
相关论文
共 50 条
  • [1] Elemental Imaging of Nail Sample by Confocal Micro-X-ray Fluorescence Analysis
    Urata, Taisei
    Matsuyama, Tsugufumi
    Inoue, Fumiyuki
    Tsuji, Kouichi
    [J]. BUNSEKI KAGAKU, 2023, 72 (06) : 217 - 225
  • [2] A method for calibrating the confocal volume of a confocal three-dimensional micro-x-ray fluorescence setup
    周鹏
    马欣然
    张爽
    孙天希
    刘志国
    [J]. Chinese Physics B, 2020, 29 (02) : 222 - 227
  • [3] Confocal micro-X-ray fluorescence analysis for difference identification of ceramic samples
    Mori, Kazuaki
    Hourai, Takayuki
    Matsuyama, Tsugufumi
    Zhuo, Shangjun
    Tsuji, Kouichi
    [J]. ANALYTICAL SCIENCES, 2024, 40 (03) : 367 - 373
  • [4] Confocal micro-X-ray fluorescence analysis for difference identification of ceramic samples
    Kazuaki Mori
    Takayuki Hourai
    Tsugufumi Matsuyama
    Shangjun Zhuo
    Kouichi Tsuji
    [J]. Analytical Sciences, 2024, 40 : 367 - 373
  • [5] Performances of capillary X-ray optics for confocal three-dimensional micro-X-ray fluorescence technology
    Sun, Tianxi
    Liu, Zhiguo
    Li, Yude
    Zhu, Guanghua
    Lin, Xiaoyan
    Luo, Ping
    Pan, Qiuli
    Liu, Hui
    Teng, Yuepeng
    Yang, Ke
    Ding, Xunliang
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 606 (03): : 829 - 832
  • [6] A method for calibrating the confocal volume of a confocal three-dimensional micro-x-ray fluorescence setup
    Zhou, Peng
    Ma, Xin-Ran
    Zhang, Shuang
    Sun, Tian-Xi
    Liu, Zhi-Guo
    [J]. CHINESE PHYSICS B, 2020, 29 (02)
  • [7] Micro-X-Ray Fluorescence and the Old Masters
    Caruso, Francesco
    Saverwyns, Steven
    Van Bos, Marina
    Martino, Delia Francesca Chillura
    Ceulemans, Anne-Emmanuelle
    de Valck, Joris
    Caponetti, Eugenio
    [J]. APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2012, 107 (01): : 197 - 202
  • [8] Confocal micro-x-ray fluorescence spectrometer for light element analysis
    Smolek, S.
    Pemmer, B.
    Foelser, M.
    Streli, C.
    Wobrauschek, P.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2012, 83 (08):
  • [9] Quantification of large scale micro-X-ray fluorescence elemental images
    Worley, CG
    Havrilla, GJ
    Dunn, PS
    [J]. APPLIED SPECTROSCOPY, 2001, 55 (11) : 1448 - 1454
  • [10] Micro-X-ray Fluorescence in Food Forensics
    St Jeor, V. L.
    Muroski, A. R.
    McGuire, C.
    Lape, A.
    [J]. 10TH INTERNATIONAL CONFERENCE ON X-RAY MICROSCOPY, 2011, 1365 : 411 - 414