Novel Bending Fatigue Testing of Small Medical Device Cables

被引:0
|
作者
Zhang, Haitao [1 ]
Li, Bernard [1 ]
机构
[1] Medtronic, 7000 Cent Ave NE, Minneapolis, MN 55432 USA
关键词
Fatigue; Cable; Medical device; Testing; Neuromodulation; DEEP BRAIN-STIMULATION; MOVEMENT-DISORDERS; EFFICACY; SAFETY;
D O I
10.1007/978-3-319-21455-9_16
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Neuromodulation is among one of the fastest-growing areas of the medical device industry. Neuro stimulators are being used to provide therapies to patients with chronic pain, movement disorders, overactive bladders, and psychiatric disorder symptoms. In these therapies, well-controlled stimulation current is delivered to the target nerves through implantable leads. The implantable leads must have sufficient fatigue durability to last for many years of implantation. The implantable leads are composed of insulation polymer jackets and conductor coils/cables. Conductor cables are generally very small and made from MP35N alloys (Co-35Ni-20Cr-10Mo). These conductor cables are more susceptible to fatigue failure than the insulation polymer jackets. A novel bending fatigue testing method was developed and used to test small medical device cables. 1x7 standard MP35N and low Ti MP35N cables are tested. The generalized S-N curve, in the form of curvature amplitude as a function of cycle to failure was obtained. The results show that low Ti MP35N cable has better fatigue performance than the standard MP35N cable. A SEM image of fractured surface shows that cable fatigue is dominated by crack initiation. The crack initiation is induced by shear deformation. This is a typical crack initiation mechanism under high plasticity deformation.
引用
收藏
页码:133 / 140
页数:8
相关论文
共 50 条
  • [41] Medical Device Feedthrough Fatigue Characterization
    Popp, J.
    Taylor, J.
    Hendrickson, J.
    MEDICAL DEVICE MATERIALS VI, 2013, : 86 - 91
  • [42] Design of 4-DOF adjusting device for Air Duct Rotary bending fatigue testing machine Tailstock
    Zhang Lingyun
    Piao Xiaodong
    PROCEEDINGS OF THE 2016 INTERNATIONAL FORUM ON MECHANICAL, CONTROL AND AUTOMATION (IFMCA 2016), 2017, 113 : 376 - 379
  • [43] DETERMINATION OF DYNAMIC ERRORS WHEN TESTING FATIGUE IN BENDING
    GRAF, ME
    RUBAN, PM
    INDUSTRIAL LABORATORY, 1958, 24 (11): : 1532 - 1535
  • [44] DESIGN AND DEVELOPMENT OF A UNIVERSAL BENDING FATIGUE TESTING MACHINE
    ARORA, PR
    RAGHAVAN, MR
    INDIAN JOURNAL OF TECHNOLOGY, 1973, 11 (12): : 713 - 719
  • [45] Fatigue Testing of TBC on Structural Steel by Cyclic Bending
    Musalek, Radek
    Kovarik, Ondrej
    Medricky, Jan
    Curry, Nicholas
    Bjorklund, Stefan
    Nylen, Per
    JOURNAL OF THERMAL SPRAY TECHNOLOGY, 2015, 24 (1-2) : 168 - 174
  • [46] FATIGUE-TESTING MACHINE FOR COMBINED BENDING AND TORSION
    SHAWKI, GSA
    MASHHOUR, AA
    EXPERIMENTAL MECHANICS, 1974, 14 (05) : 196 - 201
  • [47] Small-sized specimen design with the provision for high-frequency bending-fatigue testing
    Ghadimi, Hamed
    Jirandehi, Arash P.
    Nemati, Saber
    Guo, Shengmin
    FATIGUE & FRACTURE OF ENGINEERING MATERIALS & STRUCTURES, 2021, 44 (12) : 3517 - 3537
  • [48] Flex bending fatigue testing of wires, foils, and ribbons
    Lavvafi, H.
    Lewandowski, J. R.
    Lewandowski, J. J.
    MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 2014, 601 : 123 - 130
  • [49] Fatigue Testing of TBC on Structural Steel by Cyclic Bending
    Radek Musalek
    Ondrej Kovarik
    Jan Medricky
    Nicholas Curry
    Stefan Bjorklund
    Per Nylen
    Journal of Thermal Spray Technology, 2015, 24 : 168 - 174
  • [50] Various fatigue testing of polysilicon microcantilever beam in bending
    Hong Hocheng
    Hung, Jeng-Nan
    MICROPROCESSES AND NANOTECHNOLOGY 2007, DIGEST OF PAPERS, 2007, : 356 - 357