Comparison between small radiation therapy electron beams collimated by Cerrobend and tubular applicators

被引:6
|
作者
Di Venanzio, Cristina [1 ]
Marinelli, Marco [1 ]
Tonnetti, Alessia [1 ]
Verona-Rinati, Gianluca [1 ]
Bagala, Paolo [2 ]
Falco, Maria Daniela [2 ]
Guerra, Antonio Stefano [3 ]
Pimpinella, Maria [3 ]
机构
[1] Univ Roma Tor Vergata, Dipartimento Ingn Ind, Ist Nazl Fis Nucl, I-00133 Rome, Italy
[2] Tor Vergata Univ Gen Hosp, Dept Diagnost Imaging Mol Imaging Intervent Radio, Rome, Italy
[3] ENEA INMRI CR Casaccia, Ist Nazl Metrol Radiaz Ionizzanti, Rome, Italy
来源
关键词
Cerrobend; tubular applicators; diamond; radiation therapy; small electron beams; DIAMOND DETECTOR; ENERGY;
D O I
10.1120/jacmp.v16i1.5186
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to compare the dosimetric properties of small field electron beams shaped by circular Cerrobend blocks and stainless steel tubular applicators. Percentage depth dose curves, beam profiles, and output factors of small-size circular fields from 2 to 5 cm diameter, obtained either by tubular applicators and Cerrobend blocks, were measured for 6, 10, and 15 MeV electron beam energies. All measurements were performed using a PTW microDiamond 60019 premarket prototype. An overall similar behavior between the two collimating systems can be observed in terms of PDD and beam profiles. However, Cerrobend collimators produce a higher bremsstrahlung background under irradiation with high-energy electrons. In such irradiation condition, larger output factors are observed for tubular applicators. Similar dosimetric properties are observed using circular Cerrobend blocks and stainless steel tubular applicators at lower beam energies. However, Cerrobend collimators allow the delivery of specific beam shapes, conformed to the target area. On the other hand, in high-energy irradiation conditions, tubular applicators produce a lower bremsstrahlung contribution, leading to lower doses outside the target volume. In addition, the higher output factors observed at high energies for tubular applicators lead to reduced treatment times.
引用
收藏
页码:329 / 335
页数:7
相关论文
共 50 条
  • [41] Is EBT3 Energy Independent for Radiation-Therapy Photon & Electron Beams?
    Rubinstein, A.
    Tailor, R.
    Zhao, Z.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Accurate calculation of radiation damping parameters in the interaction between very intense laser beams and relativistic electron beams
    Popa, Alexandru
    LASER AND PARTICLE BEAMS, 2014, 32 (03) : 477 - 486
  • [43] The use of intensity-modulated radiation therapy photon beams for improving the dose uniformity of electron beams shaped with MLC
    Mosalaei, Homeira
    Karnas, Scott
    Shah, Sheel
    Van Doodewaard, Sharon
    Foster, Tim
    Chen, Jeff
    MEDICAL DOSIMETRY, 2012, 37 (01) : 76 - 83
  • [44] A comparison between spectra of runaway electron beams in SF6 and air
    Zhang, Cheng
    Tarasenko, Victor
    Gu, Jianwei
    Baksht, Evgenii
    Wang, Ruexue
    Yan, Ping
    Shao, Tao
    PHYSICS OF PLASMAS, 2015, 22 (12)
  • [45] High-dose spatially-fractionated GRID radiation therapy (SFGRT): A comparison of outcomes of treatment delivered through cerrobend GRID versus MLC GRID
    Neuner, G. A.
    Vander Walde, N.
    Ha, J. K.
    Yu, C. X.
    Mohiuddin, M.
    Regine, W. F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S488 - S488
  • [46] COMPARISON OF LARGE AND SMALL COMPUTERS FOR RADIATION-THERAPY TREATMENT PLANNING
    FENN, JO
    WALLACE, KM
    FREY, GD
    TRAVID, EA
    RADIOLOGY, 1974, 113 (02) : 429 - 433
  • [47] Dosimetric comparison between proton beam therapy and photon radiation therapy for locally advanced non-small cell lung cancer
    Wu, Chen-Ta
    Motegi, Atsushi
    Motegi, Kana
    Hotta, Kenji
    Kohno, Ryosuke
    Tachibana, Hidenobu
    Kumagai, Motoki
    Nakamura, Naoki
    Hojo, Hidehiro
    Niho, Seiji
    Goto, Koichi
    Akimoto, Tetsuo
    JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2016, 46 (11) : 1008 - 1014
  • [48] Comparison of molecular alterations and proliferative activity between diminutive and small tubular adenoma of the colorectum
    Nando, Yoshiki
    Watari, Jiro
    Ito, Chiyomi
    Tomita, Toshihiko
    Ohda, Yoshio
    Oshima, Tadayuki
    Fukui, Hirokazu
    Miwa, Hiroto
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2016, 31 : 144 - 145
  • [49] Investigation of Combining Modulated Electron Beams with Intensity Modulated Photons for Radiation Therapy of Breast Cases
    Eldib, A.
    Li, J.
    Jin, L.
    Ma, C.
    MEDICAL PHYSICS, 2012, 39 (06) : 3838 - 3838
  • [50] THE ADVANTAGES OF EMPLOYING MIXED HIGH ENERGY X-RAY AND ELECTRON BEAMS IN RADIATION THERAPY
    GALE, NH
    INNES, GS
    BRITISH JOURNAL OF RADIOLOGY, 1960, 33 (388): : 261 - 264