Comparison between small radiation therapy electron beams collimated by Cerrobend and tubular applicators

被引:6
|
作者
Di Venanzio, Cristina [1 ]
Marinelli, Marco [1 ]
Tonnetti, Alessia [1 ]
Verona-Rinati, Gianluca [1 ]
Bagala, Paolo [2 ]
Falco, Maria Daniela [2 ]
Guerra, Antonio Stefano [3 ]
Pimpinella, Maria [3 ]
机构
[1] Univ Roma Tor Vergata, Dipartimento Ingn Ind, Ist Nazl Fis Nucl, I-00133 Rome, Italy
[2] Tor Vergata Univ Gen Hosp, Dept Diagnost Imaging Mol Imaging Intervent Radio, Rome, Italy
[3] ENEA INMRI CR Casaccia, Ist Nazl Metrol Radiaz Ionizzanti, Rome, Italy
来源
关键词
Cerrobend; tubular applicators; diamond; radiation therapy; small electron beams; DIAMOND DETECTOR; ENERGY;
D O I
10.1120/jacmp.v16i1.5186
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to compare the dosimetric properties of small field electron beams shaped by circular Cerrobend blocks and stainless steel tubular applicators. Percentage depth dose curves, beam profiles, and output factors of small-size circular fields from 2 to 5 cm diameter, obtained either by tubular applicators and Cerrobend blocks, were measured for 6, 10, and 15 MeV electron beam energies. All measurements were performed using a PTW microDiamond 60019 premarket prototype. An overall similar behavior between the two collimating systems can be observed in terms of PDD and beam profiles. However, Cerrobend collimators produce a higher bremsstrahlung background under irradiation with high-energy electrons. In such irradiation condition, larger output factors are observed for tubular applicators. Similar dosimetric properties are observed using circular Cerrobend blocks and stainless steel tubular applicators at lower beam energies. However, Cerrobend collimators allow the delivery of specific beam shapes, conformed to the target area. On the other hand, in high-energy irradiation conditions, tubular applicators produce a lower bremsstrahlung contribution, leading to lower doses outside the target volume. In addition, the higher output factors observed at high energies for tubular applicators lead to reduced treatment times.
引用
收藏
页码:329 / 335
页数:7
相关论文
共 50 条
  • [1] Radiotherapy electron beams collimated by small tubular applicators: characterization by silicon and diamond diodes
    Bagala, P.
    Di Venanzio, C.
    Falco, M. D.
    Guerra, A. S.
    Marinelli, Marco
    Milani, E.
    Pimpinella, M.
    Pompili, F.
    Prestopino, G.
    Santoni, R.
    Tonnetti, A.
    Verona, C.
    Verona-Rinati, G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (22): : 8121 - 8133
  • [2] Dosimetric properties of magnetically collimated electron beams for radiation therapy
    Ma, LJ
    MEDICAL PHYSICS, 2004, 31 (11) : 2973 - 2977
  • [3] Dosimetric Characterization of Small Radiotherapy Electron Beams Collimated by Circular Applicators with the New Microsilicon Detector
    Russo, Serenella
    Bettarini, Silvia
    Grilli Leonulli, Barbara
    Esposito, Marco
    Alpi, Paolo
    Ghirelli, Alessandro
    Barca, Raffaella
    Fondelli, Simona
    Paoletti, Lisa
    Pini, Silvia
    Scoccianti, Silvia
    APPLIED SCIENCES-BASEL, 2022, 12 (02):
  • [4] Dosimetric properties of magnetically collimated electron beams for radiation therapy
    Ma, L
    MEDICAL PHYSICS, 2004, 31 (06) : 1797 - 1797
  • [5] Monte Carlo and experimental investigations of multileaf collimated electron beams for modulated electron radiation therapy
    Lee, MC
    Jiang, SB
    Ma, CM
    MEDICAL PHYSICS, 2000, 27 (12) : 2708 - 2718
  • [6] Scattered radiation from applicators in clinical electron beams
    van Battum, LJ
    van der Zee, W
    Huizenga, H
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (15): : 2493 - 2507
  • [7] SMALL ELECTRON-BEAMS IN RADIATION-THERAPY
    MCGINLEY, PH
    MCLAREN, JR
    BARNETT, BR
    RADIOLOGY, 1979, 131 (01) : 231 - 234
  • [8] Study of Dosimetric Characteristics of Clinical Electron Beams Collimated by Square Applicators using PlaneParallel Ionization Chambers
    Singh, Sant Pal
    Kulshreshtha, Asita
    Srivastava, Anoop Kumar
    Srivastava, Sumit Kumar
    Kumar, Ambrish
    JOURNAL OF PHARMACEUTICAL NEGATIVE RESULTS, 2022, 13 (04) : 883 - 891
  • [9] In-house machined copper as an alternative to cerrobend for custom cutouts in electron radiation therapy
    Hough, Cameron
    McGeachy, Phil
    MEDICAL PHYSICS, 2022, 49 (08) : 5630 - 5630
  • [10] Comparison of photon MLC and cut-out collimated electron beams for standard electron treatments
    Mueller, S.
    Fix, M. K.
    Henzen, D.
    Frei, D.
    Frauchiger, D.
    Lossl, K.
    Stampanoni, M. F. M.
    Manser, P.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 849 - 850