X-Ray Filter Design and Its Evaluation in Dual-Energy X-Ray Absorptiometry (DXA)

被引:1
|
作者
Kim, Kwang Hyun [1 ]
机构
[1] Jungwon Univ, Dept Med Engn, Goesan Gun, Chungcheongbuk, South Korea
关键词
Dual-energy X-ray; DXA; K-edge filter; X-ray beam hardening; RADIOGRAPHY; TOMOGRAPHY;
D O I
10.1109/TNS.2010.2050335
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper reports a new X-ray beam geometry to acquire two separated polychromatic X-rays as well as two separated fan beams using parallel different filters built in a single collimator. This newly designed X-ray spectrum was generated by a single X-ray shot but had the effect of double X-ray exposure. The two separated filters were used side by side. The first filter for the low energy spectra was 0.5 mm thick erbium and the second filter for the high energy spectra was composed of two layers of a 0.8 mm of copper and 0.4mmof rhodium. This combination of filters showed excellent performance of X-ray beam hardening regardless of the bone thickness in the soft tissue under DXA conditions.
引用
收藏
页码:2155 / 2158
页数:4
相关论文
共 50 条
  • [21] Dual-energy X-ray absorptiometry in early life
    Lapillonne, A
    Braillon, PM
    Delmas, PD
    Salle, BL
    HORMONE RESEARCH, 1997, 48 : 43 - 49
  • [22] SIMULATION STUDIES OF DUAL-ENERGY X-RAY ABSORPTIOMETRY
    SORENSON, JA
    DUKE, PR
    SMITH, SW
    MEDICAL PHYSICS, 1989, 16 (01) : 75 - 80
  • [23] Fundamentals and pitfalls of bone densitometry using dual-energy X-ray absorptiometry (DXA)
    Nelson B. Watts
    Osteoporosis International, 2004, 15 : 847 - 854
  • [24] Dual-energy X-ray absorptiometry scan: Guidelines
    Haymann, Jean-Philippe
    PROGRES EN UROLOGIE, 2010, 20 (04): : F119 - F122
  • [25] DUAL-ENERGY X-RAY ABSORPTIOMETRY - INTEREST IN ORTHOPEDICS
    MAUGARS, Y
    BERTHELOT, JM
    DELECRIN, J
    TRECANT, M
    PASSUTI, N
    DACULSI, G
    PROST, A
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1995, 81 (04): : 326 - 332
  • [26] DUAL-ENERGY X-RAY ABSORPTIOMETRY IN NEUROMUSCULAR DISEASES
    KANDA, F
    FUJII, Y
    TAKAHASHI, K
    FUJITA, T
    MUSCLE & NERVE, 1994, 17 (04) : 431 - 435
  • [27] Technical standardization for dual-energy X-ray absorptiometry
    Leib, E
    JOURNAL OF CLINICAL DENSITOMETRY, 2004, 7 (01) : 27 - 36
  • [28] Detection of glass contamination in horticultural peat with dual-energy X-ray absorptiometry (DXA)
    Ayalew, G
    Holden, NM
    Grace, PA
    Ward, SM
    COMPUTERS AND ELECTRONICS IN AGRICULTURE, 2004, 42 (01) : 1 - 17
  • [29] LIMITATIONS OF DUAL-ENERGY X-RAY ABSORPTIOMETRY - REPLY
    ROUBENOFF, R
    KEHAYIAS, JJ
    DAWSONHUGHES, B
    HEYMSFIELD, SB
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1995, 61 (02): : 399 - 400
  • [30] Indications and reporting for dual-energy X-ray absorptiometry
    Leib, E