The first trial of phase contrast imaging for digital full-field mammography using a practical molybdenum X-ray tube

被引:79
|
作者
Tanaka, T
Honda, C
Matsuo, S
Noma, K
Oohara, H
Nitta, N
Ota, S
Tsuchiya, K
Sakashita, Y
Yamada, A
Yamasaki, M
Furukawa, A
Takahashi, M
Murata, K
机构
[1] Shiga Univ Med Sci, Dept Radiol, Otsu, Shiga 5202192, Japan
[2] Konica Minolta Med & Graphic Inc, Ctr Res & Dev, Tokyo, Japan
关键词
phase contrast; digital mammography; computed radiography; photothermographic printer; 25-micron pixel size;
D O I
10.1097/01.rli.0000165575.43381.48
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: The image quality of a newly developed full-field digital phase contrast mammography (PCM) system and of a conventional screen-film (SF) mammography system were compared via images of a phantom and receiver operating characteristic (ROC) analysis of clinical images. Methods: Magnified (1.75x) PCM images were scanned (sampling rate, 43.75 mu m) and then reduced to original-sized, 25-micron pixel images printed on photothermographic film. Along with corresponding SF images, the phantom images were evaluated subjectively, and the clinical images of 38 patients were subjected to ROC analysis of mass and microcalcification. Results: In the image quality of a phantom, the PCM exceeded the SF. In both mass and microcalcification, the ROC analysis Az values of the PCM clinical images surpassed those of the SF images. Conclusion: The PCM provides better images than the SF. Clinical trials suggest superior detection of both mass and microcalcification by full-field digital PCM over conventional SF mammography.
引用
收藏
页码:385 / 396
页数:12
相关论文
共 50 条
  • [1] X-ray scattering in full-field digital mammography
    Nykänen, K
    Siltanen, S
    [J]. MEDICAL PHYSICS, 2003, 30 (07) : 1864 - 1873
  • [2] Advantages of magnification in digital phase-contrast mammography using a practical X-ray tube
    Honda, Chika
    Ohara, Hiromu
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : S69 - S72
  • [3] Adaptive image interpolation for full-field digital x-ray mammography
    Liu, H
    Wang, G
    Xu, F
    Fajardo, LL
    [J]. APPLIED OPTICS, 1999, 38 (01) : 253 - 257
  • [4] Phase contrast digital mammography using molybdenum X-ray: Clinical implications in detectability improvement
    Freedman, MT
    Lo, SCB
    Honda, C
    Makariou, E
    Sisney, G
    Pien, E
    Ohara, H
    Ishisaka, A
    Shimada, F
    [J]. MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 533 - 540
  • [5] X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study
    Bernhardt, Philipp
    Mertelmeier, Thomas
    Hoheisel, Martin
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4337 - 4349
  • [6] Halo suppression in full-field x-ray Zernike phase contrast microscopy
    Vartiainen, Ismo
    Mokso, Rajmund
    Stampanoni, Marco
    David, Christian
    [J]. OPTICS LETTERS, 2014, 39 (06) : 1601 - 1604
  • [7] Evaluation of Digital Breast Tomosynthesis as Replacement of Full-Field Digital Mammography Using an In Silico Imaging Trial
    Badano, Aldo
    Graff, Christian G.
    Badal, Andreu
    Sharma, Diksha
    Zeng, Rongping
    Samuelson, Frank W.
    Glick, Stephen J.
    Myers, Kyle J.
    [J]. JAMA NETWORK OPEN, 2018, 1 (07) : e185474
  • [8] Monoenergetic X-ray beam in digital full field mammography
    Diekmann, F
    Diekmann, SD
    Fischer, T
    Arkadiev, V
    Lawaczeck, R
    Hamm, BK
    [J]. RADIOLOGY, 2002, 225 : 417 - 417
  • [9] Effect of using tungsten-anode X-ray tubes on dose and image quality in full-field digital mammography
    Oduko, Jennifer M.
    Young, Kenneth C.
    Gundogdu, Ozcan
    Alsager, Abdulaziz
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 522 - +
  • [10] Contrast transfer functions for Zernike phase contrast in full-field transmission hard X-ray microscopy
    Yang, Yang
    Cheng, Yin
    Heine, Ruth
    Baumbach, Tilo
    [J]. OPTICS EXPRESS, 2016, 24 (06): : 6063 - 6070