Noise performance of a precision pulsed electromagnet power supply for magnetic resonance imaging

被引:6
|
作者
Matter, Nathaniel I. [1 ]
Chronik, Blaine [2 ]
Pauly, John M. [1 ]
Macovski, Albert [1 ]
Conolly, Steven M. [1 ,3 ]
Scott, Greig C. [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Magnet Resonance Syst Res Lab, Stanford, CA 94305 USA
[2] Univ Western Ontario, Dept Phys, London, ON N6A 3K7, Canada
[3] Univ Calif Berkeley, Dept Bioengn, Berkeley, CA 94720 USA
关键词
control systems; electromagnet power supplies; magnetic fields; magnetic resonance imaging (MRI); nuclear magnetic resonance (NMR); power electronics;
D O I
10.1109/TMI.2007.903253
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Prepolarized magnetic resonance imaging (PMRI) uses two pulsed electromagnets to achieve high-field image quality with the benefits of low-field data acquisition. The principal challenge with all resistive MRI systems is the implementation of a highly precise magnet current supply. The noise current through the magnet is fundamentally limited by the current transducer used to provide feedback and the voltage reference used to generate the demand signal. Field instability in the main field magnet can both corrupt the received data and degrade the robustness of Carr-Purcell-Meiboom-Gill (CPMG) echo trains, which are paramount to efficient imaging in PMRI. In this work, we present the magnet control system that achieved sufficient field stability for PMRI at 0.5/0.13 T, identify the dominant sources of noise in the control system, examine the imaging artifacts that can occur if the field stability is insufficient, and identify how the design can be improved for better field stability, should it be required for future implementations of PMRI.
引用
收藏
页码:75 / 86
页数:12
相关论文
共 50 条
  • [21] Measurement of noise protection in functional magnetic resonance imaging
    Liener, K
    Wunderlich, A
    Ehret, G
    Bachor, E
    LARYNGO-RHINO-OTOLOGIE, 2005, 84 (02) : 108 - 112
  • [22] Electrical Noise in the Intraoperative Magnetic Resonance Imaging Setting
    Dzwonczyk, Roger
    Fujii, Jeffrey T.
    Simonetti, Orlando
    Nieves-Ramos, Ricardo
    Bergese, Sergio D.
    ANESTHESIA AND ANALGESIA, 2009, 108 (01): : 181 - 186
  • [23] Spatially variable Rician noise in magnetic resonance imaging
    Maximov, Ivan I.
    Farrher, Ezequiel
    Grinberg, Farida
    Shah, N. Jon
    MEDICAL IMAGE ANALYSIS, 2012, 16 (02) : 536 - 548
  • [24] Acoustic noise concerns in functional magnetic resonance imaging
    Moelker, A
    Pattynama, PMT
    HUMAN BRAIN MAPPING, 2003, 20 (03) : 123 - 141
  • [25] Acoustic noise in magnetic resonance imaging: An ongoing issue
    McNulty, J. P.
    McNulty, S.
    RADIOGRAPHY, 2009, 15 (04) : 320 - 326
  • [26] NOISE-REDUCTION IN MAGNETIC-RESONANCE IMAGING
    BROSNAN, T
    WRIGHT, G
    NISHIMURA, D
    CAO, QH
    MACOVSKI, A
    SOMMER, FG
    MAGNETIC RESONANCE IN MEDICINE, 1988, 8 (04) : 394 - 409
  • [27] Noise removal in magnetic resonance diffusion tensor imaging
    Chen, B
    Hsu, EW
    MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (02) : 393 - 401
  • [28] Acoustic noise during functional magnetic resonance imaging
    Ravicz, ME
    Melcher, JR
    Kiang, NYS
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (04): : 1683 - 1696
  • [29] Reducing the impact of power supply noise on microprocessor performance
    Rahal-Arabi, T
    Wong, KL
    Ma, M
    Barkatullah, J
    Taylor, G
    Electrical Performance of Electronic Packaging, 2004, : 307 - 310
  • [30] Noise analysis in smash imaging at ultrahigh field magnetic resonance imaging
    Wassenaar, PA
    Schmalbrock, P
    Clymer, BD
    SEVENTH IASTED INTERNATIONAL CONFERENCE ON SIGNAL AND IMAGE PROCESSING, 2005, : 104 - 109