Impedance cardiography as a noninvasive technique for atrioventricular interval optimization in cardiac resynchronization therapy

被引:41
|
作者
Braun, MU [1 ]
Schnabel, A [1 ]
Rauwolf, T [1 ]
Schulze, M [1 ]
Strasser, RH [1 ]
机构
[1] Tech Univ Dresden, Dept Internal Med & Cardiol, Med Clin 2, D-01307 Dresden, Germany
关键词
impedance cardiography; Doppler flow echocardiography; cardiac resynchronization therapy; AV-interval optimization;
D O I
10.1007/s10840-005-2361-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Impedance cardiography (IC) and Doppler echocardiography (DE) are two noninvasive methods to evaluate hemodynamics in patients with dual-chamber pacemakers. The aim of the present study was to compare both techniques in respect to their ability of AV-interval optimization in cardiac resynchronization therapy (CRT) based on cardiac output (CO) measurements. Methods and Results: Twenty-four patients (64 +/- 8 years) with congestive heart failure (EF < 35%; NYHA III-IV) and LBBB (> 150 ms) were evaluated at baseline and 1 month after implantation of a CRT-D. The optimal AV interval was defined by IC and subsequently by transaortic flow DE as the interval corresponding to the highest CO measured at different AV intervals, varying from 60 to 200 ms (with 20 ms increments). For standardization and comparison of both techniques, a fixed atriobiventricular pacing rate of 90 beats/min was used. Absolute values of COmax were higher by IC (5.8 +/- 0.9 l/min) as compared to DE (4.6 +/- 0.9 l/min, p < 0.01). The optimal AV interval as determined by IC varied interindividually from 80-180 ms (mean: 121 +/- 18 ms). In DE, the range was also 80-180 ms with the mean optimal AV interval of 128 +/- 23 ms. Thus, there was a strong correlation for AV-interval optimization in CRT between both methods (r = 0.74; p < 0.001). Conclusion: In CRT, AV-interval optimization based on CO values determined by IC correlates closely to those measured by transaortic flow DE. Impedance cardiography as an easy and cost-effective technique for AV-interval optimization is a promising alternative for routine management of heart failure patients on a beat-to-beat analysis during CRT follow-up.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 50 条
  • [31] CARDIAC OUTPUT MEASURED BY NONINVASIVE IMPEDANCE CARDIOGRAPHY AND PICU OUTCOMES
    Sumbel, Lydia
    Wats, Anchal
    Bhalala, Utpal
    CRITICAL CARE MEDICINE, 2020, 48
  • [32] Role of impedance cardiography in predicting clinical near-term outcomes of patients with cardiac resynchronization therapy
    Martinelli Filho, MMartino
    Peixoto, G. L.
    Siqueira, S. F.
    Lensi, M. M.
    Trindade, R. H. L.
    Nishioka, S. A. D.
    Teixeira, R. A.
    Pedrosa, A. A. A.
    Costa, R.
    EUROPEAN JOURNAL OF HEART FAILURE, 2014, 16 : 155 - 155
  • [33] Doppler echocardiographic methods for optimization of the atrioventricular delay during cardiac resynchronization therapy
    Waggoner, Alan D.
    Fuentes, Lisa de las
    Davila-Roman, Victor G.
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2008, 25 (09): : 1047 - 1055
  • [34] The effects of QRS duration and atrial pacing on atrioventricular optimization of cardiac resynchronization therapy
    Gold, Michael R.
    Niazi, Imran
    Giudici, Michael
    Yu, Yinghong
    Arcot-Krishnamurthy, Shantha
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 2A - 2A
  • [35] Prospective comparison of echocardiographic atrioventricular delay optimization methods for cardiac resynchronization therapy
    Kerlan, JE
    Sawhney, NS
    Waggoner, AD
    Chawla, MK
    Garhwal, S
    Osborn, JL
    Faddis, MN
    HEART RHYTHM, 2006, 3 (02) : 148 - 154
  • [36] Atrioventricular intrinsic conduction and cardiac resynchronization therapy
    Wang Ru-xing
    Guo Tao
    CHINESE MEDICAL JOURNAL, 2011, 124 (03) : 436 - 439
  • [37] Atrioventricular junction ablation in cardiac resynchronization therapy
    Foley, Paul
    Smith, Russell
    Leyva, Francisco
    BRITISH JOURNAL OF HOSPITAL MEDICINE, 2009, 70 (02) : 112 - 112
  • [38] NONINVASIVE DETERMINATION OF CARDIAC-OUTPUT BY MEANS OF IMPEDANCE CARDIOGRAPHY - COMPARATIVE EVALUATION WITH A THERMAL DILUTION TECHNIQUE
    NECHWATAL, W
    BIER, P
    EVERSMANN, A
    KONIG, E
    BASIC RESEARCH IN CARDIOLOGY, 1976, 71 (05) : 542 - 552
  • [39] Impedance cardiography is superior to echocardiographic methods for pacing interval optimization
    Turcott, Robert G.
    Witteles, Ronald M.
    Wang, Paul J.
    Vagelos, Randall H.
    Fowler, Michael B.
    Ashley, Euan A.
    JOURNAL OF CARDIAC FAILURE, 2008, 14 (06) : S65 - S65
  • [40] NONINVASIVE ASSESSMENT OF CARDIAC-OUTPUT BY IMPEDANCE CARDIOGRAPHY IN THE NEWBORN CANINE
    GOTSHALL, RW
    MILES, DS
    CRITICAL CARE MEDICINE, 1989, 17 (01) : 63 - 65