Airtraq Laryngoscope Versus the Conventional Macintosh Laryngoscope During Pediatric Intubation Performed by Nurses: A Randomized Crossover Manikin Study With Three Airway Scenarios

被引:5
|
作者
Szarpak, Lukasz [1 ]
Karczewska, Katarzyna [2 ]
Czyzewski, Lukasz [3 ]
Truszewski, Zenon [1 ]
Kurowski, Andrzej [4 ]
机构
[1] Med Univ Warsaw, Dept Emergency Med, Lindleya 4 St, PL-02005 Warsaw, Poland
[2] Mazovian Reg Hosp, Anesthesiol & Intens Care Unit, Radom, Poland
[3] Med Univ Warsaw, Dept Nephrol Nursing, Warsaw, Poland
[4] Cardinal Wyszynski Natl Inst Cardiol, Dept Anesthesiol, Warsaw, Poland
关键词
pediatric; endotracheal intubation; airtraq; macintosh; difficult airway; EMERGENCY CARDIOVASCULAR CARE; HEART-ASSOCIATION GUIDELINES; TRACHEAL INTUBATION; CARDIOPULMONARY-RESUSCITATION; LIFE-SUPPORT; ENDOTRACHEAL INTUBATION; MILLER LARYNGOSCOPE; CHEST COMPRESSION; NOVICE DOCTORS; CARDIAC-ARREST;
D O I
10.1097/PEC.0000000000000741
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives We hypothesized that the Airtraq laryngoscope (Airtraq LLC, Bonita Springs, Fla) is beneficial for intubation of pediatric manikins while performing cardiopulmonary resuscitation (CPR). In the present study, we evaluated the effectiveness of the Macintosh (MAC) laryngoscope (HEINE Optotechnik, Munich, Germany) and Airtraq in 3 simulated CPR scenarios. Methods A randomized crossover simulation trial was designed. Eighty-three nurses intubated the trachea of a PediaSIM CPR training manikin (FCAE HealthCare, Sarasota, Fla) using the MAC and Airtraq in a normal airway scenario, normal airway with chest compression scenario, and difficult airway with chest compression scenario. The participants were directed to perform a maximum of 3 attempts in each scenario. The success rate, time to intubation, Cormack & Lehane grade, dental compression, and the ease of intubation were measured. Results All participants performed successful intubation with the Airtraq in all 3 scenarios. In all scenarios, the success rate was significantly higher and the time to intubation was significantly shorter with the Airtraq than with the MAC. Glottic visualization using the Cormack-Lehane scale was also better when using Airtraq in all scenarios. Conclusions In this manikin study, we found that the Airtraq can be used successfully for the intubation of pediatric manikins with normal and difficult airways by medical staff without previous experience in pediatric intubation. Moreover, intubation can be achieved without interrupting chest compression. The use of the Airtraq compared with the MAC led to faster time to intubation. Nevertheless, we recommend that the performance of the Airtraq and the MAC during CPR should be further evaluated in a clinical setting.
引用
收藏
页码:735 / 739
页数:5
相关论文
共 50 条
  • [1] Randomized comparison of the airway scope and the macintosh laryngoscope for the tracheal intubation in manikin study
    Yuasa, Haruyuki
    Yokoyama, Hiroyuki
    Yonemoto, Naohiro
    Kada, Akiko
    Kasahara, Yoichiro
    Koga, Yoshihisa
    Nonogi, Hiroshi
    CIRCULATION, 2007, 116 (16) : 944 - 944
  • [2] A comparative evaluation of conventional Macintosh laryngoscope and the Airtraq (R) in different intubation scenarios
    Ali, Qazi Ehsan
    Amir, Syed Hussain
    Siddiqui, Obaid A.
    Mahopatra, Partha S.
    SRI LANKAN JOURNAL OF ANAESTHESIOLOGY, 2012, 20 (01): : 3 - 6
  • [3] Airway Scope versus Macintosh laryngoscope: a manikin study
    Hirabayashi, Yoshihiro
    EMERGENCY MEDICINE JOURNAL, 2007, 24 (05) : 357 - 358
  • [4] Comparing Performance of Airtraq Versus Macintosh Laryngoscope for Pediatric Intubation by Novices
    Li, Hui-Xian
    Liu, Ya-Yang
    Yang, Gui-Zhen
    Xue, Fu-Shan
    PEDIATRIC EMERGENCY CARE, 2018, 34 (10) : E196 - E196
  • [5] Tracheal intubation in daylight and in the dark: a randomised comparison of the Airway Scope®, Airtraq®, and Macintosh laryngoscope in a manikin
    Ueshima, H.
    Asai, T.
    ANAESTHESIA, 2010, 65 (07) : 684 - 687
  • [6] IS AIRTRAQ VIDEO LARYNGOSCOPE A BETTER ALTERNATIVE TO CONVENTIONAL MACINTOSH DIRECT LARYNGOSCOPE DURING ROUTINE INTUBATION? A COMPARATIVE STUDY
    Mahmood, Syed Fazal
    Padmanabha, S.
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2015, 4 (93): : 15903 - 15904
  • [7] Tracheal Intubation of a Difficult Airway Using Airway Scope, Airtraq, and Macintosh Laryngoscope: A Comparative Manikin Study of Inexperienced Personnel
    Liu, Liangji
    Tanigawa, Koichi
    Kusunoki, Shinji
    Tamura, Tomoko
    Ota, Kohei
    Yamaga, Satoshi
    Kida, Yoshiko
    Otani, Tadatsugu
    Sadamori, Takuma
    Takeda, Taku
    Iwasaki, Yasumasa
    Hirohashi, Nobuyuki
    ANESTHESIA AND ANALGESIA, 2010, 110 (04): : 1049 - 1055
  • [8] Randomized crossover comparison of the AirTraq Avant® to the Macintosh laryngoscope for intubation with during simulated resuscitation by novice physicians
    Truszewski, Zenon
    Czyzewski, Lukasz
    Marchese, Giuseppe
    Buljan, Domagoj
    Szarpak, Lukasz
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (08): : 1708 - 1709
  • [9] Comparison of Tracheal Intubation Using Airway Scope and Macintosh Laryngoscope in Manikin Study
    Xue, Fu S.
    Xiong, Jun
    Wang, Qiang
    Yuan, Yu J.
    Liao, Xu
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2010, 69 (04): : 1007 - 1008
  • [10] Evaluation of intubation using the Airtraq® or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy -: a manikin study
    Maharaj, CH
    Higgins, BD
    Harte, BH
    Laffey, JG
    ANAESTHESIA, 2006, 61 (05) : 469 - 477