Airtraq Laryngoscope Versus the Conventional Macintosh Laryngoscope During Pediatric Intubation Performed by Nurses: A Randomized Crossover Manikin Study With Three Airway Scenarios

被引:5
|
作者
Szarpak, Lukasz [1 ]
Karczewska, Katarzyna [2 ]
Czyzewski, Lukasz [3 ]
Truszewski, Zenon [1 ]
Kurowski, Andrzej [4 ]
机构
[1] Med Univ Warsaw, Dept Emergency Med, Lindleya 4 St, PL-02005 Warsaw, Poland
[2] Mazovian Reg Hosp, Anesthesiol & Intens Care Unit, Radom, Poland
[3] Med Univ Warsaw, Dept Nephrol Nursing, Warsaw, Poland
[4] Cardinal Wyszynski Natl Inst Cardiol, Dept Anesthesiol, Warsaw, Poland
关键词
pediatric; endotracheal intubation; airtraq; macintosh; difficult airway; EMERGENCY CARDIOVASCULAR CARE; HEART-ASSOCIATION GUIDELINES; TRACHEAL INTUBATION; CARDIOPULMONARY-RESUSCITATION; LIFE-SUPPORT; ENDOTRACHEAL INTUBATION; MILLER LARYNGOSCOPE; CHEST COMPRESSION; NOVICE DOCTORS; CARDIAC-ARREST;
D O I
10.1097/PEC.0000000000000741
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives We hypothesized that the Airtraq laryngoscope (Airtraq LLC, Bonita Springs, Fla) is beneficial for intubation of pediatric manikins while performing cardiopulmonary resuscitation (CPR). In the present study, we evaluated the effectiveness of the Macintosh (MAC) laryngoscope (HEINE Optotechnik, Munich, Germany) and Airtraq in 3 simulated CPR scenarios. Methods A randomized crossover simulation trial was designed. Eighty-three nurses intubated the trachea of a PediaSIM CPR training manikin (FCAE HealthCare, Sarasota, Fla) using the MAC and Airtraq in a normal airway scenario, normal airway with chest compression scenario, and difficult airway with chest compression scenario. The participants were directed to perform a maximum of 3 attempts in each scenario. The success rate, time to intubation, Cormack & Lehane grade, dental compression, and the ease of intubation were measured. Results All participants performed successful intubation with the Airtraq in all 3 scenarios. In all scenarios, the success rate was significantly higher and the time to intubation was significantly shorter with the Airtraq than with the MAC. Glottic visualization using the Cormack-Lehane scale was also better when using Airtraq in all scenarios. Conclusions In this manikin study, we found that the Airtraq can be used successfully for the intubation of pediatric manikins with normal and difficult airways by medical staff without previous experience in pediatric intubation. Moreover, intubation can be achieved without interrupting chest compression. The use of the Airtraq compared with the MAC led to faster time to intubation. Nevertheless, we recommend that the performance of the Airtraq and the MAC during CPR should be further evaluated in a clinical setting.
引用
收藏
页码:735 / 739
页数:5
相关论文
共 50 条
  • [31] Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study
    Komasawa, Nobuyasu
    Ueki, Ryusuke
    Nomura, Hajime
    Itani, Motoi
    Kaminoh, Yoshiroh
    JOURNAL OF ANESTHESIA, 2010, 24 (02) : 306 - 308
  • [32] Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study
    Nobuyasu Komasawa
    Ryusuke Ueki
    Hajime Nomura
    Motoi Itani
    Yoshiroh Kaminoh
    Journal of Anesthesia, 2010, 24 : 306 - 308
  • [33] Comparison of the AP Advance™ Video Laryngoscope and C-MAC® Video Laryngoscope with the Macintosh 3 Blade for intubation of the difficult airway: a manikin study
    Davies, O.
    Kourteli, E.
    Cecconi, M.
    Garewel, D.
    Mir, F.
    BRITISH JOURNAL OF ANAESTHESIA, 2016, 116 (06) : E899 - E899
  • [34] A manikin study to compare video-optical intubation stylet versus Macintosh laryngoscope used by novice in normal and simulated difficult airway intubation
    Tsui, K. L.
    Hung, C. Y.
    Kam, C. W.
    HONG KONG JOURNAL OF EMERGENCY MEDICINE, 2008, 15 (03) : 133 - 138
  • [35] Comparison of Miller laryngoscope and UEScope videolaryngoscope for endotracheal intubation in four pediatric airway scenarios: a randomized, crossover simulation trial
    Jacek Smereka
    Marcin Madziala
    Dominika Dunder
    Elzbieta Makomaska-Szaroszyk
    Lukasz Szarpak
    European Journal of Pediatrics, 2019, 178 : 937 - 945
  • [36] Comparison of Miller laryngoscope and UEScope videolaryngoscope for endotracheal intubation in four pediatric airway scenarios: a randomized, crossover simulation trial
    Smereka, Jacek
    Madziala, Marcin
    Dunder, Dominika
    Makomaska-Szaroszyk, Elzbieta
    Szarpak, Lukasz
    EUROPEAN JOURNAL OF PEDIATRICS, 2019, 178 (06) : 937 - 945
  • [37] Comparison of Macintosh laryngoscope, non-channelled (C-MAC video-laryngoscope) and channelled laryngoscope (Airtraq) for intubation in lateral position - A prospective randomized controlled study
    Sultana, Farheen
    Nikhar, Sapna Annaji
    Durga, Padmaja
    TRENDS IN ANAESTHESIA AND CRITICAL CARE, 2021, 40 : 30 - 34
  • [38] Comparison of a New Video Intubation Stylet and McGrath® MAC Video Laryngoscope for Intubation in an Airway Manikin with Normal Airway and Cervical Spine Immobilization Scenarios by Novice Personnel: A Randomized Crossover Study
    Park, Jin-Woo
    An, Sungmin
    Park, Seongjoo
    Nahm, Francis Sahngun
    Han, Sung-Hee
    Kim, Jin-Hee
    BIOMED RESEARCH INTERNATIONAL, 2021, 2021
  • [39] A manikin study comparing the performance of traditional Macintosh laryngoscope, GlideScope®, Airtraq®, and video-optical intubation stylet in endotracheal intubation used by emergency doctors in simulated difficult airway intubation: A pilot study
    Li, Yu On
    Wong, Oi Fung
    Ko, Shing
    Ma, Hing Man
    Lit, Chau Hung Albert
    Shih, Yau Ngai
    HONG KONG JOURNAL OF EMERGENCY MEDICINE, 2023, 30 (03) : 179 - 188