Tracheal Intubation of a Difficult Airway Using Airway Scope, Airtraq, and Macintosh Laryngoscope: A Comparative Manikin Study of Inexperienced Personnel

被引:32
|
作者
Liu, Liangji [1 ]
Tanigawa, Koichi [1 ]
Kusunoki, Shinji [1 ]
Tamura, Tomoko [1 ]
Ota, Kohei [1 ]
Yamaga, Satoshi [1 ]
Kida, Yoshiko [1 ]
Otani, Tadatsugu [1 ]
Sadamori, Takuma [1 ]
Takeda, Taku [1 ]
Iwasaki, Yasumasa [1 ]
Hirohashi, Nobuyuki [1 ]
机构
[1] Hiroshima Univ, Dept Emergency & Crit Care Med, Hiroshima 7348551, Japan
来源
ANESTHESIA AND ANALGESIA | 2010年 / 110卷 / 04期
关键词
VIDEO LARYNGOSCOPE; PENTAX-AWS; PENTAX-AWS(R); PERFORMANCE;
D O I
10.1213/ANE.0b013e3181d2aad7
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: The Airway Scope (AWS) (Pentax-AWS, Hoya Corp., Tokyo, Japan) and the Airtraq (ATQ) (Prodol, Vizcaya, Spain) have similarities in the novel structures of their blades. In this study, we evaluated the ease of use of the AWS and ATQ compared with the Macintosh laryngoscope (ML) by inexperienced personnel in a simulated manikin difficult airway. METHODS: Twenty-four fifth-year medical students with no previous experience in tracheal intubation participated in this study. We used an advanced patient simulator (SimMan (R), Laerdal Medical, Stavanger, Norway) to simulate difficult airway scenarios including cervical spine rigidity, limited mouth opening, and pharyngeal obstruction. The sequences in selecting devices and scenarios were randomized. Success rates for tracheal intubation, and the time required for visualization of the glottis, tracheal intubation, and inflation of the lungs, and the number of optimization maneuvers and dental click sounds were analyzed. The 3 different intubation devices were tested in 4 different scenarios by 24 students. RESULTS: Both the AWS and ATQ had very high success rates of tracheal intubation compared with the ML (AWS 100%*; ATQ 98%*; and ML 89%; *P < 0.05 AWS, ATQ versus ML). The time to intubation with the AWS was significantly shorter than with the ATQ and ML (AWS 11 6 seconds; ATQ 16 12 seconds; and ML 16 11 seconds; *P < 0.05 AWS versus ATQ, ML). The number of optimization maneuvers with the AWS was significantly lower than with the ATQ and ML. There were significantly more audible dental click sounds with the ML than with the AWS and ATQ. CONCLUSION: Both the AWS and ATQ may be suitable devices for difficult intubation by inexperienced personnel in this manikin simulated scenario. Further studies in a clinical setting are necessary to confirm these findings. (Anesth Ana Ig 2010;110:1049-55)
引用
收藏
页码:1049 / 1055
页数:7
相关论文
共 50 条
  • [1] TRACHEAL INTUBATION OF A DIFFICULT AIRWAY USING IN-DIRECT LARYNGOSCOPES, AIRWAY SCOPE®, AIRTRAQ® AND GLIDESCOPE:® A COMPARATIVE MANIKIN STUDY OF INEXPERIENCED PERSONNEL
    Tamura, Tomoko
    Tanigawa, Koichi
    Hirohashi, Nobuyuki
    Iwasaki, Yasumasa
    Takeda, Taku
    Tsumura, Ryu
    Otani, Tadatsugu
    Sadamori, Takuma
    CRITICAL CARE MEDICINE, 2010, 38 (12) : U130 - U130
  • [2] Tracheal intubation for a difficult airway using Airway scope®, KingVision® and McGRATH®: a comparative manikin study of inexperienced personnel
    J Itai
    Y Tanabe
    T Nishida
    T Inagawa
    Y Torikoshi
    Y Kida
    T Tamura
    K Ota
    T Otani
    T Sadamori
    K Une
    R Tsumura
    Y Iwasaki
    N Hirohashi
    K Tanigawal
    Critical Care, 17 (Suppl 2):
  • [3] Comparison of Tracheal Intubation Using Airway Scope and Macintosh Laryngoscope in Manikin Study
    Xue, Fu S.
    Xiong, Jun
    Wang, Qiang
    Yuan, Yu J.
    Liao, Xu
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2010, 69 (04): : 1007 - 1008
  • [4] Tracheal intubation in daylight and in the dark: a randomised comparison of the Airway Scope®, Airtraq®, and Macintosh laryngoscope in a manikin
    Ueshima, H.
    Asai, T.
    ANAESTHESIA, 2010, 65 (07) : 684 - 687
  • [5] Randomized comparison of the airway scope and the macintosh laryngoscope for the tracheal intubation in manikin study
    Yuasa, Haruyuki
    Yokoyama, Hiroyuki
    Yonemoto, Naohiro
    Kada, Akiko
    Kasahara, Yoichiro
    Koga, Yoshihisa
    Nonogi, Hiroshi
    CIRCULATION, 2007, 116 (16) : 944 - 944
  • [6] Performance of a Pentax-Airway scope and a Macintosh laryngoscope for difficult intubation in a manikin study
    Xue, F. S.
    Liao, X.
    Yuan, Y. J.
    Liu, J. H.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2011, 55 (07) : 900 - 901
  • [7] Performance of a Pentax-Airway scope and a Macintosh laryngoscope for difficult intubation in a manikin study Reply
    Kariya, T.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2011, 55 (07) : 901 - 901
  • [8] Tracheal intubation by inexperienced medical residents using the Airtraq and Macintosh laryngoscopes - a manikin study
    Maharaj, Chrisen H.
    Chonghaile, Martina Ni
    Higgins, Brendan D.
    Harte, Brian H.
    Laffey, John G.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2006, 24 (07): : 769 - 774
  • [9] Airway Scope versus Macintosh laryngoscope: a manikin study
    Hirabayashi, Yoshihiro
    EMERGENCY MEDICINE JOURNAL, 2007, 24 (05) : 357 - 358
  • [10] Tracheal intubation with restricted access: a randomised comparison of the Pentax-Airway Scope and Macintosh laryngoscope in a manikin
    Asai, T.
    ANAESTHESIA, 2009, 64 (10) : 1114 - 1117