Comparison of Vascular Responses Following New-Generation Biodegradable and Durable Polymer-Based Drug-Eluting Stent Implantation in an Atherosclerotic Rabbit Iliac Artery Model

被引:36
|
作者
Nakazawa, Gaku [1 ]
Torii, Sho [1 ]
Ijichi, Takeshi [1 ]
Nagamatsu, Hirofumi [1 ]
Ohno, Yohei [1 ]
Kurata, Fumi [1 ]
Yoshikawa, Ayako [1 ]
Nakano, Masataka [1 ]
Shinozaki, Norihiko [1 ]
Yoshimachi, Fuminobu [1 ]
Ikari, Yuji [1 ]
机构
[1] Tokai Univ, Sch Med, Dept Cardiol, 143 Shimokasuya, Isehara, Kanagawa 2591193, Japan
来源
关键词
atherosclerosis; endothelialization; polymer; stent; BARE-METAL STENTS; MYOCARDIAL-INFARCTION; SPIRIT III; THROMBOSIS; RESTENOSIS; PATHOLOGY; TRIAL; PREDICTORS; MICRORNAS; HUMANS;
D O I
10.1161/JAHA.116.003803
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Incomplete endothelialization is the primary substrate of late stent thrombosis; however, recent reports have revealed that abnormal vascular responses are also responsible for the occurrence of late stent failure. The aim of the current study was to assess vascular response following deployment of biodegradable polymer-based Synergy (Boston Scientific) and Nobori (Terumo) drug-eluting stents and the durable polymer-based Resolute Integrity stent (Medtronic) in an atherosclerotic rabbit iliac artery model. Methods and Results-A total of 24 rabbits were fed an atherogenic diet, and then a balloon injury was used to induce atheroma formation. Synergy, Nobori, and Resolute Integrity stents were randomly implanted in iliac arteries. Animals were euthanized at 28 days for scanning electron microscopic evaluation and at 90 days for histological analysis. The percentage of uncovered strut area at 28 days was lowest with Synergy, followed by Resolute Integrity, and was significantly higher with Nobori stents (Synergy 1.1 +/- 2.2%, Resolute Integrity 2.0 +/- 3.9%, Nobori 4.6 +/- 3.0%; P< 0.001). At 90 days, inflammation score was lowest for Synergy (0.27 +/- 0.45), followed by Nobori (0.62 +/- 0.59), and was highest for Resolute Integrity (0.89 +/- 0.46, P< 0.001). Foamy macrophage infiltration within neointima (ie, neoatherosclerosis) was significantly less with Synergy (0.62 +/- 0.82) compared with Nobori (0.85 +/- 0.74) and Resolute Integrity (1.39 +/- 1.32; P=0.034). Conclusions-The biodegradable polymer-coated thin-strut Synergy drug-eluting stent showed the fastest stent strut neointimal coverage and the lowest incidence of neoatherosclerosis in the current animal model.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Early and chronic phased local coagulative responses following bioresorbable-polymer drug-eluting stent implantation
    Yamaguchi, K.
    Wakatsuki, T.
    Okushi, Y.
    Suto, K.
    Matsumoto, K.
    Takahashi, T.
    Kadota, M.
    Kawabata, Y.
    Matsuura, T.
    Ise, T.
    Kusunose, K.
    Yagi, S.
    Yamada, H.
    Soeki, T.
    Sata, M.
    EUROPEAN HEART JOURNAL, 2021, 42 : 1245 - 1245
  • [42] Prognosis of biodegradable polymer drug-eluting stents versus second-generation durable polymer drug-eluting stents in acute myocardial infarction patients following percutaneous coronary intervention
    Choe, J. C.
    Cha, K. S.
    Shin, J. Y.
    Lee, S. H.
    Ahn, J.
    Park, J. S.
    Lee, H. W.
    Oh, J. H.
    Choi, J. H.
    Lee, H. C.
    Hong, T. J.
    Jeong, M. H.
    EUROPEAN HEART JOURNAL, 2018, 39 : 273 - 274
  • [43] Comparison of coronary angioscopic findings 8 months after stent implantation between two kinds of biodegradable polymer-coated and one durable polymer-coated drug-eluting stent
    Yasumura, Keisuke
    Shutta, Ryu
    Matsunaga-Lee, Yasuharu
    Nakamura, Daisuke
    Yano, Masamichi
    Yamato, Masaki
    Egami, Yasuyuki
    Sakata, Yasushi
    Nishino, Masami
    Tanouchi, Jun
    CORONARY ARTERY DISEASE, 2021, 32 (02) : 91 - 95
  • [44] Thrombogenicity and Early Vascular Healing Response in Metallic Biodegradable Polymer-Based and Fully Bioabsorbable Drug-Eluting Stents
    Koppara, Tobias
    Cheng, Qi
    Yahagi, Kazuyuki
    Mori, Hiroyoshi
    Sanchez, Oscar David
    Feygin, Julia
    Wittchow, Eric
    Kolodgie, Frank D.
    Virmani, Renu
    Joner, Michael
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2015, 8 (06)
  • [45] Randomized Ivus Comparison Between a Durable and a Biodegradable-Polymer Second Generation Drug-Eluting Stents: The Bioactive Trial
    Costa, Jose de Ribamar
    Almeida, Breno
    Costa, Ricardo
    Chamie, Daniel
    Perin, Marco
    Abizaid, Andrea
    Sousa, Amanda
    Abizaid, Alexandre
    CIRCULATION, 2014, 130
  • [46] Very early neointimal coverage of new biodegradable polymer drug-eluting stent compared with durable polymer everolimus-eluting stent evaluated by optical frequency domain imaging
    Kobayashi, Norihiro
    Ito, Yoshiaki
    Yamawaki, Masahiro
    Araki, Motoharu
    Sakai, Tsuyoshi
    Sakamoto, Yasunari
    Mori, Shinsuke
    Tsutsumi, Masakazu
    Nauchi, Masahiro
    Honda, Yohsuke
    Tokuda, Takahiro
    Makino, Kenji
    Shirai, Shigemitsu
    Hirano, Keisuke
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2018, 34 (04): : 515 - 522
  • [47] Very early neointimal coverage of new biodegradable polymer drug-eluting stent compared with durable polymer everolimus-eluting stent evaluated by optical frequency domain imaging
    Norihiro Kobayashi
    Yoshiaki Ito
    Masahiro Yamawaki
    Motoharu Araki
    Tsuyoshi Sakai
    Yasunari Sakamoto
    Shinsuke Mori
    Masakazu Tsutsumi
    Masahiro Nauchi
    Yohsuke Honda
    Takahiro Tokuda
    Kenji Makino
    Shigemitsu Shirai
    Keisuke Hirano
    The International Journal of Cardiovascular Imaging, 2018, 34 : 515 - 522
  • [48] In vivo comparison of a polymer-free Biolimus A9-eluting stent with a biodegradable polymer-based Biolimus A9 eluting stent and a bare metal stent in balloon denuded and radiated hypercholesterolemic rabbit iliac arteries
    Waksman, Ron
    Pakala, Rajbabu
    Baffour, Richard
    Seabron, Rufus
    Hellinga, David
    Chan, Rosanna
    Su, Shih-Horng
    Kolodgie, Frank
    Virmani, Renu
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 80 (03) : 429 - 436
  • [49] Association between angiographic and intravascular ultrasound optimizations after new-generation drug-eluting stent implantation and clinical outcomes
    Cha, Jung-Joon
    Kim, Daehoon
    Kim, Byeong-Keuk
    Hong, Sung-Jin
    Ahn, Chul-Min
    Kim, Jung-Sun
    Ko, Young-Guk
    Choi, Donghoon
    Hong, Myeong-Ki
    Jang, Yangsoo
    CORONARY ARTERY DISEASE, 2021, 32 (06) : 541 - 548
  • [50] New Drug-Eluting Stents An Overview on Biodegradable and Polymer-Free Next-Generation Stent Systems
    Abizaid, Alexandre
    Costa, J. Ribamar, Jr.
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (04) : 384 - 393