Comparison of Vascular Responses Following New-Generation Biodegradable and Durable Polymer-Based Drug-Eluting Stent Implantation in an Atherosclerotic Rabbit Iliac Artery Model

被引:36
|
作者
Nakazawa, Gaku [1 ]
Torii, Sho [1 ]
Ijichi, Takeshi [1 ]
Nagamatsu, Hirofumi [1 ]
Ohno, Yohei [1 ]
Kurata, Fumi [1 ]
Yoshikawa, Ayako [1 ]
Nakano, Masataka [1 ]
Shinozaki, Norihiko [1 ]
Yoshimachi, Fuminobu [1 ]
Ikari, Yuji [1 ]
机构
[1] Tokai Univ, Sch Med, Dept Cardiol, 143 Shimokasuya, Isehara, Kanagawa 2591193, Japan
来源
关键词
atherosclerosis; endothelialization; polymer; stent; BARE-METAL STENTS; MYOCARDIAL-INFARCTION; SPIRIT III; THROMBOSIS; RESTENOSIS; PATHOLOGY; TRIAL; PREDICTORS; MICRORNAS; HUMANS;
D O I
10.1161/JAHA.116.003803
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Incomplete endothelialization is the primary substrate of late stent thrombosis; however, recent reports have revealed that abnormal vascular responses are also responsible for the occurrence of late stent failure. The aim of the current study was to assess vascular response following deployment of biodegradable polymer-based Synergy (Boston Scientific) and Nobori (Terumo) drug-eluting stents and the durable polymer-based Resolute Integrity stent (Medtronic) in an atherosclerotic rabbit iliac artery model. Methods and Results-A total of 24 rabbits were fed an atherogenic diet, and then a balloon injury was used to induce atheroma formation. Synergy, Nobori, and Resolute Integrity stents were randomly implanted in iliac arteries. Animals were euthanized at 28 days for scanning electron microscopic evaluation and at 90 days for histological analysis. The percentage of uncovered strut area at 28 days was lowest with Synergy, followed by Resolute Integrity, and was significantly higher with Nobori stents (Synergy 1.1 +/- 2.2%, Resolute Integrity 2.0 +/- 3.9%, Nobori 4.6 +/- 3.0%; P< 0.001). At 90 days, inflammation score was lowest for Synergy (0.27 +/- 0.45), followed by Nobori (0.62 +/- 0.59), and was highest for Resolute Integrity (0.89 +/- 0.46, P< 0.001). Foamy macrophage infiltration within neointima (ie, neoatherosclerosis) was significantly less with Synergy (0.62 +/- 0.82) compared with Nobori (0.85 +/- 0.74) and Resolute Integrity (1.39 +/- 1.32; P=0.034). Conclusions-The biodegradable polymer-coated thin-strut Synergy drug-eluting stent showed the fastest stent strut neointimal coverage and the lowest incidence of neoatherosclerosis in the current animal model.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Arterial Responses Following Drug-Eluting Stent Implantation in an Atherosclerotic Rabbit Iliac Model
    Melder, Robert J.
    Kakazawa, Gaku
    Pruitt, Sean
    Kolodgie, Frank D.
    Virmani, Renu
    Wilcox, Josiah N.
    AMERICAN JOURNAL OF CARDIOLOGY, 2008, 102 (8A): : 163I - 163I
  • [2] Comparison of vascular response between durable and biodegradable polymer-based drug-eluting stents in a porcine coronary artery model
    Nakazawa, Gaku
    Shinke, Toshiro
    Ijichi, Takeshi
    Matsumoto, Daisuke
    Otake, Hiromasa
    Torii, Sho
    Hiranuma, Noritoshi
    Ohsue, Tsuyoshi
    Otsuka, Fumiyuki
    Shite, Junya
    Hirata, Ken-ichi
    Ikari, Yuji
    EUROINTERVENTION, 2014, 10 (06) : 717 - 723
  • [3] Comparison of intermediate-term vascular response to new-generation biodegradable polymer and durable polymer-based drug-eluting stents: optical coherence tomographic study
    Kawamura, A.
    Nishino, M.
    Matsuhiro, Y.
    Nakamura, Z.
    Yasumura, K.
    Yasumoto, K.
    Tanaka, A.
    Matsunaga, Y.
    Nakamura, D.
    Yano, M.
    Yamato, M.
    Egami, Y.
    Syutta, R.
    Tanouchi, Z.
    EUROPEAN HEART JOURNAL, 2019, 40 : 2015 - 2015
  • [4] Vascular Healing Following Comparator Drug-Eluting Stents Implantation in an Atherosclerotic Rabbit Iliac Model - A Model of More Predictive in Human Responses to Drug-Eluting Stents
    Nakazawa, Gaku
    Wilcox, Josiah N.
    Melder, Robert
    Pruitt, Sean
    Kolodgie, Frank D.
    Virmani, Renu
    CIRCULATION, 2008, 118 (18) : S746 - S746
  • [5] Vascular Response to Drug-eluting Stent with Biodegradable Versus Durable Polymer
    Katayama, Yosuke
    Kubo, Takashi
    Akasaka, Takashi
    Aoki, Hiroshi
    Nishiguchi, Tsuyoshi
    Satogami, Keisuke
    Kameyama, Takeyoshi
    Yamano, Takashi
    Yamaguchi, Tomoyuki
    Matsuo, Yoshiki
    Ino, Yasushi
    Kitabata, Hironori
    Tanaka, Atsushi
    Hozumi, Takeshi
    Kimura, Takeshi
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2017, 69 (16) : S47 - S47
  • [6] Rotational atherectomy and new-generation drug-eluting stent implantation
    Hachinohe, Daisuke
    Kashima, Yoshifumi
    Kanno, Daitaro
    Kobayashi, Ken
    Sugie, Takuro
    Kaneko, Umihiko
    Tadano, Yutaka
    Watanabe, Tomohiko
    Shitan, Hidemasa
    Fujita, Tsutomu
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2018, 91 (06) : 1026 - 1034
  • [7] Comparisons of early vascular reactions in biodegradable and durable polymer-based drug-eluting stents in the porcine coronary artery
    Ijichi, Takeshi
    Nakazawa, Gaku
    Torii, Sho
    Nagamatsu, Hirofumi
    Yoshikawa, Ayako
    Souba, Junko
    Isobe, Atsushi
    Hagiwara, Hitomi
    Ikari, Yuji
    PLOS ONE, 2019, 14 (01):
  • [8] Evaluation of Polymer-Based Comparator Drug-Eluting Stents Using a Rabbit Model of Iliac Artery Atherosclerosis
    Nakazawa, Gaku
    Nakano, Masataka
    Otsuka, Fumiyuki
    Wilcox, Josiah N.
    Melder, Robert
    Pruitt, Sean
    Kolodgie, Frank D.
    Virmani, Renu
    CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2011, 4 (01) : 38 - U75
  • [9] Pathology of Coronary Artery After Third-generation Abluminal Biodegradable-polymer Drug-eluting Stent Implantation in Atherosclerotic Porcine Model
    Migita, Suguru
    Kitano, Daisuke
    Li, Yuxin
    Koyama, Yutaka
    Onishi, Akira
    Fuchimoto, Daiichiro
    Suzuki, Shunichi
    Nakamura, Yoshiyuki
    Hirayama, Atsushi
    Okumura, Yasuo
    Hao, Hiroyuki
    CIRCULATION, 2020, 142
  • [10] In vitro and in vivo characterisation of biodegradable polymer-based drug-eluting stent
    Yazdani, Saami K.
    Vorpahl, Marc
    Nakano, Masataka
    Su, Shih-Horng
    Kolodgie, Frank D.
    Virmani, Renu