Changes in intracuff pressure of a cuffed endotracheal tube during prolonged surgical procedures

被引:27
|
作者
Kako, Hiromi [1 ,2 ]
Goykhman, Anatoliy [3 ]
Ramesh, Archana S. [1 ]
Krishna, Senthil G. [1 ,2 ]
Tobias, Joseph D. [1 ,2 ,4 ]
机构
[1] Nationwide Childrens Hosp, Dept Anesthesiol & Pain Med, Columbus, OH 43205 USA
[2] Ohio State Univ, Dept Anesthesiol & Pain Med, Columbus, MS USA
[3] Ohio Univ, Heritage Coll Osteopath Med, Athens, GA USA
[4] Ohio State Univ, Dept Pediat, Columbus, MS USA
关键词
Airway management; Intraoperative monitoring; Intubation; Endotracheal; Pediatrics; TRACHEAL TUBES; CHILDREN; ANESTHESIA; INFANTS;
D O I
10.1016/j.ijporl.2014.11.017
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Background: With the introduction of redesigned cuffed endotracheal tubes (ETTs), there has been an increasing trend toward their use in pediatric patients. Despite improvements in design, an unintended and prolonged hyperinflation of the cuff can compromise tracheal mucosal perfusion. The current study prospectively monitors changes in intracuff pressure continuously in pediatric patients undergoing prolonged surgical procedures. Methods: The study was conducted on pediatric patients who were scheduled to undergo prolonged surgical procedures (more than 4 h) with a cuffed ETT. After placement of the cuffed ETT, the cuff was inflated using the air-leak test with a CPAP of 20 cmH(2)O in the anesthesia circuit. After inflation, the inflating port of the pilot balloon was connected to the transducer of the invasive pressure monitoring device using our previously described technique to continuously measure the intracuff pressure. Measurements were recorded every 15 min for the first 1 h, and then every 30 min throughout the surgical procedure. Results: The study cohort included 30 patients who ranged in age from 1.2 to 17.6 years and in weight from 9.4 to 113.4 kg. There were 16 boys and 14 girls. The size of the cuffed ETT ranged from 3.5 mm to 8.0 mm ID. The baseline intracuff pressure at the time of inflation was 17.6 +/- 8.8 cmH(2)O. The absolute change in the intraoperative intracuff pressure when compared to the baseline intracuff pressure ranged from -25.8 to +16.3 cmH(2)O. In 9 patients (30%), the decrease of the intracuff pressure was >= 10 cmH(2)O. In 6 patients (20%), the increase of the intracuff pressure was >= 10 cmH(2)O. In 5 of 30 patients (17%), the absolute intracuff pressure was greater than 30 cmH(2)O at least once intraoperatively. In no patient, did the intracuff pressure remain the same as the baseline throughout the procedure. Conclusion: We noted significant variations in the intracuff pressure during prolonged surgical procedures. These unintended changes, both increases and decreases, may impact the perioperative course of patients. Our study suggests the need for continuously monitoring intracuff pressure if a cuffed ETT is used in children for prolonged surgical procedures. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:76 / 79
页数:4
相关论文
共 50 条
  • [21] The effect of mouth prop on endotracheal tube intracuff pressure in children during dental rehabilitation under general anaesthesia
    Canpolat, Dilek Gunay
    Cantekin, Kenan
    Bayram, Adnan
    Yildirim, Mustafa Denizhan
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2018, 32 (01) : 141 - 145
  • [22] The effect of mouth prop on endotracheal tube intracuff pressure in children during dental rehabilitation under general anaesthesia
    Dilek Günay Canpolat
    Kenan Cantekin
    Adnan Bayram
    Mustafa Denizhan Yıldırım
    Journal of Clinical Monitoring and Computing, 2018, 32 : 141 - 145
  • [23] An in vitro and in vivo validation of a novel color-coded syringe device for measuring the intracuff pressure in cuffed endotracheal tubes
    Kamata, Mineto
    Kako, Hiromi
    Ramesh, Archana S.
    Krishna, Senthil G.
    Tobias, Joseph D.
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (07): : 11356 - 11359
  • [24] The relationship between head and neck position and endotracheal tube intracuff pressure in the pediatric population
    Kako, Hiromi
    Krishna, Senthil G.
    Ramesh, Archana S.
    Merz, Meredith N.
    Elmaraghy, Charles
    Grischkan, Jonathan
    Jatana, Kris R.
    Ruda, James
    Tobias, Joseph D.
    PEDIATRIC ANESTHESIA, 2014, 24 (03) : 316 - 321
  • [25] TRACHEAL STENOSIS ASSOCIATED WITH A LOW-PRESSURE CUFFED ENDOTRACHEAL-TUBE
    BRADBEER, TL
    JAMES, ML
    SEAR, JW
    SEARLE, JF
    STACEY, R
    ANAESTHESIA, 1976, 31 (04) : 504 - 507
  • [26] Comparison of changes in the balloon pressure of the endotracheal intraoperative tube: coelioscopy versus other procedures
    Lagant, C.
    Jolly, I.
    Smeyers, M.
    Biz, P.
    Sukami, M.
    Rubeo-Lisa, G.
    Combe, A.
    Fournier, C.
    Keita, H.
    ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2013, 32 : A412 - A412
  • [27] Intracuff pressure of endotracheal tubes with or without 'pressure compensating cuff' during ventilation with nitrous oxide
    Sala-Blanch, X
    Masià, R
    Correa, J
    Taurá, P
    BRITISH JOURNAL OF ANAESTHESIA, 1999, 82 : 31 - 31
  • [28] The effect on intracuff pressure of various nitrous oxide concentrations used for inflating an endotracheal tube cuff
    Karasawa, F
    Ohshima, T
    Takamatsu, I
    Ehata, T
    Fukuda, I
    Uchihashi, Y
    Satoh, T
    ANESTHESIA AND ANALGESIA, 2000, 91 (03): : 708 - 713
  • [29] The TaperGuard™ Endotracheal Tube Intracuff Pressure Increase Is Less Than That of the Hi-Lo™ Tube During Nitrous Oxide Exposure: A Model Trachea Study
    Tsuboi, Sayaka
    Miyashita, Tetsuya
    Yamaguchi, Yoshikazu
    Yamamoto, Yoshiko
    Sakamaki, Kentaro
    Goto, Takahisa
    ANESTHESIA AND ANALGESIA, 2013, 116 (03): : 609 - 612
  • [30] Changes in cuff pressure of endotracheal tube during laparoscopic and open abdominal surgery
    Zeynep Baysal Yildirim
    Ali Uzunkoy
    Ali Cigdem
    Suleyman Ganidagli
    Abdullah Ozgonul
    Surgical Endoscopy, 2012, 26 : 398 - 401