Wavelength-encoded multi-channel spectral domain optical coherence tomography (SD-OCT) for spatiotemporally resolved imaging

被引:0
|
作者
Mekonnen, Taye [1 ]
Kourmatzis, Agisilaos [2 ]
Cheng, Shaokoon [1 ]
机构
[1] Macquarie Univ, Sch Engn, Sydney, NSW 2109, Australia
[2] Univ Sydney, Sch Aerosp Mech & Mechatron Engn, Sydney, NSW 2006, Australia
关键词
optical coherence tomography; interferometer; multi-channel; spectral decomposition; CONFIGURATION; LAYERS; SCANS;
D O I
10.1088/1361-6501/ab85d6
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In conventional spectral domain optical coherence tomography (SD-OCT), a beam is focused to one spot on the surface of the sample to acquire depth information beneath that spot. Adjacent spots are serially scanned to obtain a tomographic image of the sample; thus, available OCT techniques can take significant time to scan large objects. This paper illustrates an SD-OCT system involving spectral-slicing of a broadband spectrum to enable the simultaneous acquisition from multiple lateral positions of the sample. Using a wavelength encoding of spatial information, where each lateral location is associated with a specific portion of a source spectrum, multiple lateral spots are interrogated concurrently to improve the imaging speed of SD-OCT. A triple-channel SD-OCT system was devised using a broadband source of centre wavelength 840 nm and full width half maximum (FWHM) bandwidth of 50 nm. Both sample and reference arm beams were split into three channels using two dichroic splitters of 50% cut-on wavelength at 825 nm and 848 nm, resulting in a FWHM spectral distribution of 13.5 nm, 19 nm and 17.5 nm in the three channels, respectively. The imaging performance of the technique was demonstrated by concurrently acquiring tomographic images of a 3D-printed phantom. Despite the relative drop in the per-channel axial resolution and depth sensitivity, this new approach presents a straightforward way of interrogating a sample simultaneously in multiple locations, provides the additional advantage of allowing different delay lines for each spectral section and offers the potential of such a system to be used in a wide range of applications that demand a small product foot print.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] DEPTH RESOLVED IMAGING OF NON BIOLOGICAL SAMPLE BY SPECTRAL DOMAIN OPTICAL COHERENCE TOMOGRAPHY (SD-OCT)
    Panakkal, Hareesh
    Rao, Yogesh
    Jayasree, V. K.
    Makkar, Roshan
    2015 INTERNATIONAL CONFERENCE ON MICROWAVE, OPTICAL AND COMMUNICATION ENGINEERING (ICMOCE), 2015, : 247 - 249
  • [2] Modelling for Spectral Domain Optical Coherence Tomography (SD-OCT) System
    Choudhari, Suyog
    Patil, Mukesh
    Makkar, Roshan
    ADVANCES IN OPTICAL SCIENCE AND ENGINEERING, 2017, 194 : 591 - 597
  • [3] Spectral Domain Optical Coherence Tomography (SD-OCT) for Salt Quantification in Blood
    Ullah, H.
    Andleeb, F.
    Batool, Z.
    Nazir, A.
    LASERS IN ENGINEERING, 2020, 47 (1-3) : 77 - 94
  • [4] Intraoperative Spectral Domain Optical Coherence Tomography (SD-OCT) Imaging of Complex Pediatric Retinal Disease
    Berrocal, A. M.
    Houston, S. K.
    Pina, Y.
    Murray, T. G.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [5] Compressive SD-OCT: the application of compressed sensing in spectral domain optical coherence tomography
    Liu, Xuan
    Kang, Jin U.
    OPTICS EXPRESS, 2010, 18 (21): : 22010 - 22019
  • [6] Spectral Domain Optical Coherence Tomography (SD-OCT) Findings in Refsum's Disease
    Kaya, Cagdas
    Somfai, Gabor M.
    Gerding, Heinrich
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2017, 234 (04) : 527 - 529
  • [7] Glaucoma progression analysis by Spectral-Domain Optical Coherence Tomography (SD-OCT)
    Renard, J-P
    Fenolland, R.
    Giraud, J-M
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2019, 42 (05): : 499 - 516
  • [8] En face spectral domain optical coherence tomography (SD-OCT) for corneal dystrophies
    Ghouali, Wajdene
    Hassani, Rachid Tahiri Joutei
    Liang, Hong
    Brasnu, Emmanuelle
    Labbe, Antoine
    Baudouin, Christophe
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [9] Histopathologic Correlation with Spectral-Domain Optical Coherence Tomography (SD-OCT), a Novel Digital Imaging Technique
    Smith, M.
    Kagemann, L.
    Schuman, J. S.
    Fine, J. L.
    LABORATORY INVESTIGATION, 2012, 92 : 394A - 394A
  • [10] Histopathologic Correlation with Spectral-Domain Optical Coherence Tomography (SD-OCT), a Novel Digital Imaging Technique
    Smith, M.
    Kagemann, L.
    Schuman, J. S.
    Fine, J. L.
    MODERN PATHOLOGY, 2012, 25 : 394A - 394A