Electronic measurement of cell antigen expression in whole blood

被引:6
|
作者
Civelekoglu, Ozgun [1 ]
Liu, Ruxiu [1 ]
Usanmaz, Can F. [1 ]
Chu, Chia-Heng [1 ]
Boya, Mert [1 ]
Ozkaya-Ahmadov, Tevhide [1 ]
Arifuzzman, A. K. M. [1 ]
Wang, Ningquan [1 ]
Sarioglu, A. Fatih [1 ,2 ,3 ]
机构
[1] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, Parker H Petit Inst Bioengn & Biosci, Atlanta, GA 30332 USA
[3] Georgia Inst Technol, Inst Elect & Nanotechnol, Atlanta, GA 30332 USA
基金
美国国家科学基金会;
关键词
FLOW-CYTOMETRY; PROGENITOR CELLS; DISEASE; PRINCIPLES; MOLECULES; HISTOGRAM; NETWORKS; DENSITY; MARKERS; RISK;
D O I
10.1039/d1lc00889g
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Membrane antigens are phenotypic signatures of cells used for distinguishing various subpopulations and, therefore, are of great interest for diagnosis of diseases and monitoring of patients in hematology and oncology. Existing methods to measure antigen expression of a target subpopulation in blood samples require labor-intensive lysis of contaminating cells and subsequent analysis with complex and bulky instruments in specialized laboratories. To address this long-standing limitation in clinical cytometry, we introduce a microchip-based technique that can directly measure surface expression of target cells in hematological samples. Our microchip isolates an immunomagnetically-labeled target cell population from the contaminating background in whole blood and then utilizes the differential responses of target cells to on-chip magnetic manipulation to estimate their antigen expression. Moreover, manipulating cells with chip-sized permanent magnets and performing quantitative measurements via an on-chip electrical sensor network allows the assay to be performed in a portable platform with no reliance on laboratory infrastructure. Using our technique, we could successfully measure expressions of the CD45 antigen that is commonly expressed by white blood cells, as well as CD34 that is expressed by scarce hematopoietic progenitor cells, which constitutes only similar to 0.0001% of all blood cells, directly from whole blood. With our technology, flow cytometry can potentially become a rapid bedside or at-home testing method that is available around the clock in environments where this invaluable assay with proven clinical utility is currently either outsourced or not even accessible.
引用
收藏
页码:296 / 312
页数:17
相关论文
共 50 条
  • [31] A COMPARISON OF 3 FLUOROPHORES FOR USE IN AN OPTICAL BIOSENSOR FOR THE MEASUREMENT OF PROSTATE-SPECIFIC ANTIGEN IN WHOLE-BLOOD
    DANIELS, PB
    FLETCHER, JE
    ONEILL, PM
    STAFFORD, CG
    BACARESEHAMILTON, T
    ROBINSON, GA
    SENSORS AND ACTUATORS B-CHEMICAL, 1995, 27 (1-3) : 447 - 451
  • [32] Donor white blood cell differential is the single largest determinant of whole blood gene expression patterns
    O'Connell, Grant C.
    Wang, Jing
    Smothers, Christine
    GENOMICS, 2023, 115 (06)
  • [33] An optimized whole blood assay for flow cytometric measurement of ZAP-70 protein expression
    Forman, M
    Cobb, J
    Scibelli, P
    Smith, C
    Mills, R
    Shankey, TV
    CYTOMETRY PART B-CLINICAL CYTOMETRY, 2005, 67B (01) : 42 - 42
  • [34] THE MEASUREMENT OF AMMONIA IN WHOLE BLOOD, ERYTHROCYTES, AND PLASMA
    SELIGSON, D
    HIRAHARA, K
    JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1957, 49 (06): : 962 - 973
  • [35] MEASUREMENT OF PLATELET MICROPARTICLES IN DONATED WHOLE BLOOD
    Tamura, S.
    Akino, M.
    Katsumata, M.
    Homma, C.
    Azuma, H. A.
    Kato, T.
    Ikeda, H.
    VOX SANGUINIS, 2009, 97 : 39 - 39
  • [36] MEASUREMENT OF VOLATILE MERCAPTANS IN WHOLE-BLOOD
    DOIZAKI, WM
    ZIEVE, L
    JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1973, 82 (04): : 674 - 681
  • [37] More on the measurement of ionized magnesium in whole blood
    Ritter, C
    Ghahramani, M
    Marsoner, HJ
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1996, 56 : 275 - 280
  • [38] Evaluation of a biosensor for the measurement of lactate in whole blood
    de Keijzer, MH
    Brandts, RW
    Brans, PGW
    CLINICAL BIOCHEMISTRY, 1999, 32 (02) : 109 - 112
  • [39] Arginine measurement in dry and whole blood samples
    Manan, Z.
    Makhdoom, Az
    Jain, S.
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2025, 369 : S299 - S300
  • [40] APPLICATION OF THE COLORIMETRIC WHOLE-BLOOD METHOD TO THE MEASUREMENT OF BOVINE RED-BLOOD-CELL CHOLINESTERASE ACTIVITY
    ROBBINS, WE
    HOPKINS, TL
    ROTH, AR
    JOURNAL OF ECONOMIC ENTOMOLOGY, 1958, 51 (03) : 326 - 329