Cardiac-gated dual-energy imaging of the chest: Design and performance evaluation of a cardiac trigger based on a fingertip pulse oximeter

被引:0
|
作者
Shkumat, N.
Siewerdsen, J.
Dhanantwari, A.
Williams, D.
Pau, N.
Yorkston, J.
Van Metter, R.
机构
[1] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
[2] Princess Margaret Hosp, Ontario Canc Inst, Toronto, ON M4X 1K9, Canada
[3] Princess Margaret Hosp, Dept Med Imaging, Toronto, ON M4X 1K9, Canada
[4] Eastman Kodak Co, Rochester, NY USA
[5] Eastman Kodak Co, Washington, DC USA
关键词
D O I
10.1118/1.2760492
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A research prototype for high‐performance dual‐energy (DE) imaging of the chest is under development. This paper discusses the development and characterization of a cardiac gating system designed to precisely trigger the imaging system according to cardiac phase and minimize anatomical misregistration due to heart motion. Method and Materials: A fingertip pulse oximeter was employed to measure the peripheral pulse waveform and trigger x‐ray exposures during the quiescent phase of the heart (diastole). Temporal delays accounted in the timing scheme include physiological pulse propagation, waveform processing, and imaging system delays (filter‐wheel, bucky‐grid, and flat‐panel detector). An empirical model of the diastolic period allows calculation of the implemented delay, timp, required to trigger correctly at any patient heart‐rate. Performance was evaluated in terms of accuracy and precision of diastole‐trigger coincidence and expert assessment of cardiac motion artifact in gated and ungated DE images. Results: The model suggests a triggering scheme characterized by two heart‐rate (HR) regimes: below a HR‐threshold, sufficient time exists to expose on the same heartbeat [formula omitted]; above the HR‐threshold, a characteristic timp(HR) delays exposure to the subsequent heartbeat, accounting for all fixed and variable system delays. Initial implementation indicated 83% accuracy in diastole‐trigger coincidence. By modifying the HR estimation method (reduced temporal smoothing of the pulse waveform), trigger accuracy of 100% was achieved. Cardiac‐gated DE patient images demonstrate significantly reduced cardiac motion as assessed by expert radiologists. Conclusion: A pulse oximeter combined with a cardiac model provides accurate x‐ray triggering and significantly reduces heart motion artifacts. A simple fingertip clip presents logistic, cost, and workflow advantages compared to ECG. The system has been implemented in a clinical research trial, with gated and ungated arms allowing characterization of the impact of cardiac motion artifact on diagnostic performance. Conflict of Interest: Research sponsored in part by Eastman Kodak. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2364 / 2364
页数:1
相关论文
共 50 条
  • [41] Quantitative Analysis of a Whole Cardiac Mass Using Dual-Energy Computed Tomography: Comparison with Conventional Computed Tomography and Magnetic Resonance Imaging
    Yoo Jin Hong
    Jin Hur
    Kyunghwa Han
    Dong Jin Im
    Young Joo Suh
    Hye-Jeong Lee
    Young Jin Kim
    Byoung Wook Choi
    Scientific Reports, 8
  • [42] Design of Wireless Dual-Energy Dual-Source Versatile Pediatric Imaging System Based on CMOS Flat-Panel Detectors
    Qi, Yihong
    Zhou, Zhiyu
    Wang, Yuliang
    Chu, Jianping
    Li, Ziping
    Wang, Kai
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [43] Angle Dependence of Electrode Lead-Related Artifacts in Single- and Dual-Energy Cardiac ECG-Gated CT Scanning-A Phantom Study
    Tarkowski, Piotr
    Siek, Elzbieta
    Staskiewicz, Grzegorz
    Bielecki, Dennis K.
    Czekajska-Chehab, Elzbieta
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (13)
  • [44] Design optimization of X-ray security scanner based on dual-energy transmission imaging with variable tube voltage
    Park, Junsung
    Ana, Geunyoung
    Seo, Hee
    JOURNAL OF INSTRUMENTATION, 2023, 18 (01):
  • [45] Design and fabrication of a sandwich detector for material discrimination and contrast cancellation in dual-energy based x-ray imaging
    Alikunju, Rimcy Palakkappilly
    Buchanan, Ian
    Esposito, Michela
    Morehen, Jason
    Khan, Asmar
    Stamatis, Yiannis
    Iacovou, Nicolas
    Bullard, Edward
    Anaxagoras, Thalis
    Brodrick, James
    Olivo, Alessandro
    JOURNAL OF APPLIED PHYSICS, 2024, 135 (15)
  • [46] A Cascaded Systems Model for Imaging Performance and Task-Based Optimization in Dual-Energy Cone-Beam CT
    Gang, G.
    Prakash, P.
    Zbijewski, W.
    Stayman, J.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2011, 38 (06) : 3833 - +
  • [47] Myocardial perfusion imaging using adenosine-induced stress dual-energy computed tomography of the heart: comparison with cardiac magnetic resonance imaging and conventional coronary angiography
    Ko, Sung Min
    Choi, Jin Woo
    Song, Meong Gun
    Shin, Je Kyoun
    Chee, Hyun Kun
    Chung, Hyun Woo
    Kim, Dong Hun
    EUROPEAN RADIOLOGY, 2011, 21 (01) : 26 - 35
  • [48] Myocardial perfusion imaging using adenosine-induced stress dual-energy computed tomography of the heart: comparison with cardiac magnetic resonance imaging and conventional coronary angiography
    Ko, S.
    Song, M. G.
    Shin, J. K.
    Chee, H. K.
    EUROPEAN HEART JOURNAL, 2010, 31 : 581 - 581
  • [49] Myocardial perfusion imaging using adenosine-induced stress dual-energy computed tomography of the heart: comparison with cardiac magnetic resonance imaging and conventional coronary angiography
    Sung Min Ko
    Jin Woo Choi
    Meong Gun Song
    Je Kyoun Shin
    Hyun Kun Chee
    Hyun Woo Chung
    Dong Hun Kim
    European Radiology, 2011, 21 : 26 - 35
  • [50] The development and characterization of a dual-energy subtraction imaging system for chest radiography based on CsI:Tl amorphous silicon flat-panel technology
    Sabol, JM
    Avinash, GB
    Nicolas, F
    Claus, B
    Zhao, JG
    Dobbins, JT
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 399 - 408