Cardiac-gated dual-energy imaging of the chest: Design and performance evaluation of a cardiac trigger based on a fingertip pulse oximeter

被引:0
|
作者
Shkumat, N.
Siewerdsen, J.
Dhanantwari, A.
Williams, D.
Pau, N.
Yorkston, J.
Van Metter, R.
机构
[1] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
[2] Princess Margaret Hosp, Ontario Canc Inst, Toronto, ON M4X 1K9, Canada
[3] Princess Margaret Hosp, Dept Med Imaging, Toronto, ON M4X 1K9, Canada
[4] Eastman Kodak Co, Rochester, NY USA
[5] Eastman Kodak Co, Washington, DC USA
关键词
D O I
10.1118/1.2760492
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A research prototype for high‐performance dual‐energy (DE) imaging of the chest is under development. This paper discusses the development and characterization of a cardiac gating system designed to precisely trigger the imaging system according to cardiac phase and minimize anatomical misregistration due to heart motion. Method and Materials: A fingertip pulse oximeter was employed to measure the peripheral pulse waveform and trigger x‐ray exposures during the quiescent phase of the heart (diastole). Temporal delays accounted in the timing scheme include physiological pulse propagation, waveform processing, and imaging system delays (filter‐wheel, bucky‐grid, and flat‐panel detector). An empirical model of the diastolic period allows calculation of the implemented delay, timp, required to trigger correctly at any patient heart‐rate. Performance was evaluated in terms of accuracy and precision of diastole‐trigger coincidence and expert assessment of cardiac motion artifact in gated and ungated DE images. Results: The model suggests a triggering scheme characterized by two heart‐rate (HR) regimes: below a HR‐threshold, sufficient time exists to expose on the same heartbeat [formula omitted]; above the HR‐threshold, a characteristic timp(HR) delays exposure to the subsequent heartbeat, accounting for all fixed and variable system delays. Initial implementation indicated 83% accuracy in diastole‐trigger coincidence. By modifying the HR estimation method (reduced temporal smoothing of the pulse waveform), trigger accuracy of 100% was achieved. Cardiac‐gated DE patient images demonstrate significantly reduced cardiac motion as assessed by expert radiologists. Conclusion: A pulse oximeter combined with a cardiac model provides accurate x‐ray triggering and significantly reduces heart motion artifacts. A simple fingertip clip presents logistic, cost, and workflow advantages compared to ECG. The system has been implemented in a clinical research trial, with gated and ungated arms allowing characterization of the impact of cardiac motion artifact on diagnostic performance. Conflict of Interest: Research sponsored in part by Eastman Kodak. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2364 / 2364
页数:1
相关论文
共 50 条
  • [1] Cardiac gating with a pulse oximeter for dual-energy imaging
    Shkumat, N. A.
    Siewerdsen, J. H.
    Dhanantwari, A. C.
    Williams, D. B.
    Paul, N. S.
    Yorkston, J.
    Van Metter, R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (21): : 6097 - 6112
  • [2] Development and implementation of a high-performance, cardiac-gated dual-energy imaging system
    Shkumat, N. A.
    Siewerdsen, J. H.
    Dhanantwari, A. C.
    Williams, D. B.
    Richard, S.
    Tward, D. J.
    Paul, N. S.
    Yorkston, J.
    Van Metter, R.
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [3] Evaluation of a commercial cardiac motion phantom for dual-energy chest radiography
    Hsieh, Ching-Yi
    Gladish, Gregory
    Willis, Charles E.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 235 - 251
  • [4] QUANTIFICATION TECHNIQUES FOR DUAL-ENERGY CARDIAC IMAGING
    MOLLOI, SY
    MISTRETTA, CA
    MEDICAL PHYSICS, 1989, 16 (02) : 209 - 217
  • [5] Dual-Energy Computed Tomography in Cardiac Imaging
    Boettcher, Benjamin
    Zsarnoczay, Emese
    Varga-Szemes, Akos
    Schoepf, Uwe Joseph
    Meinel, Felix G.
    van Assen, Marly
    De Cecco, Carlo N.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 995 - 1009
  • [6] Cardiac-gated multidetector CT coronary angiography in evaluation of emergency department chest pain patients
    Mahr, C. W.
    Flaherty Jr, J. J.
    Mylonas, I
    Salanitri, J.
    ANNALS OF EMERGENCY MEDICINE, 2006, 48 (04) : S16 - S16
  • [7] Effects of the technique parameters on the imaging performance of the dual-energy chest radiography
    Kim, Changsoo
    Yang, Sung-Hee
    Kim, Junwoo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2021, 78 (09) : 849 - 859
  • [8] Effects of the technique parameters on the imaging performance of the dual-energy chest radiography
    Changsoo Kim
    Sung-Hee Yang
    Junwoo Kim
    Journal of the Korean Physical Society, 2021, 78 : 849 - 859
  • [9] The Feasibility of Dual-Energy Computed Tomography in Cardiac Contusion Imaging for Mildest Blunt Cardiac Injury
    Sade, Recep
    Kantarci, Mecit
    Ogul, Hayri
    Bayraktutan, Ummugulsum
    Uzkeser, Mustafa
    Aslan, Sahin
    Aksakal, Enbiya
    Becit, Necip
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (03) : 354 - 359
  • [10] Evaluation of respiratory and cardiac motion correction schemes in dual gated PET/CT cardiac imaging
    Lamare, F.
    Le Maitre, A.
    Dawood, M.
    Schaefers, K. P.
    Fernandez, P.
    Rimoldi, O. E.
    Visvikis, D.
    MEDICAL PHYSICS, 2014, 41 (07)