Metal Allergy as a Cause of Implant Failure in Shoulder Arthroplasty

被引:5
|
作者
Ko, Jia-Wei Kevin [1 ]
Nicholson, Thema A. [1 ]
Hoffler, C. Edward [1 ]
Williams, Gerald, Jr. [1 ]
Getz, Charles [1 ]
机构
[1] Rothman Inst, 925 Chestnut St, Philadelphia, PA 19107 USA
关键词
QUALITY-OF-LIFE; JOINT REPLACEMENT; SENSITIVITY; HYPERSENSITIVITY; VALIDITY; HEMIARTHROPLASTY; OSTEOARTHRITIS; RELIABILITY; PREVALENCE; OUTCOMES;
D O I
10.3928/01477447-20170719-01
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Metal allergy is an uncommon and poorly understood cause of failure of orthopedic implants. To the authors' knowledge, there have been no reports of the management of shoulder arthroplasty patients with metal allergy. The authors present their experience with the diagnosis and management of patients with metal allergy. Patients with metal allergy undergoing shoulder arthroplasty were identified through retrospective chart review from January 1, 2012, to January 31, 2015. Case characteristics collected included patient risk factors (age, sex, prior cutaneous reactions to metal), metal allergy factors (type of metal allergy, method of diagnosis), and surgery factors (implant type, primary/revision, type of shoulder arthroplasty). Outcomes measured included American Shoulder and Elbow Surgeons score, Penn Shoulder Score, and Single Assessment Numeric Evaluation score. Eleven patients were identified with metal allergy. Five were diagnosed prior to the index arthroplasty, and 6 were diagnosed after shoulder replacement. The diagnosis was made through skin patch testing, memory lymphocyte immunostimulation assay, or clinical history. Patients identified after implantation presented with progressive pain and stiffness, but none had cutaneous manifestations. Patients with metal allergy had better results undergoing primary shoulder arthroplasty than undergoing revision. Metal allergy is rare but may be a clinically significant cause of unsatisfactory shoulder arthroplasty. Given the superior results of primary shoulder arthroplasty compared with revision, screening for metal allergy by clinical history is recommended.
引用
下载
收藏
页码:E844 / E848
页数:5
相关论文
共 50 条
  • [41] ARTHROPLASTY AND ACUTE SHOULDER TRAUMA - REASONS FOR SUCCESS AND FAILURE
    COMPITO, CA
    SELF, EB
    BIGLIANI, LU
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1994, (307) : 27 - 36
  • [42] ALLERGY TO FOODS - COMMON CAUSE OF GROWTH FAILURE
    SANDBERG, DH
    CONLY, PW
    BERNSTEIN, CW
    CLEVELAND, WW
    PEDIATRIC RESEARCH, 1973, 7 (04) : 297 - 297
  • [43] Implant positioning in reverse shoulder arthroplasty has an impact on acromial stresses
    Wong, Murray T.
    Langohr, G. Daniel G.
    Athwal, George S.
    Johnson, James A.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2016, 25 (11) : 1889 - 1895
  • [44] Biometrical analysis of the shoulder joint regarding glenoid implant dimensions for arthroplasty
    Kircher, Joern
    Bittersohl, Bernd
    Zilkens, Christoph
    Hedtmann, Achim
    Krauspe, Rudiger
    SURGICAL AND RADIOLOGIC ANATOMY, 2014, 36 (04) : 321 - 325
  • [45] Secondary posttraumatic shoulder arthroplasty. Implant choice and clinical outcome
    Koenigshausen, M.
    Gessmann, J.
    Schildhauer, T. A.
    Seybold, D.
    TRAUMA UND BERUFSKRANKHEIT, 2014, 16 : 107 - 111
  • [46] Reverse shoulder arthroplasty for proximal humerus fractures: Is the glenoid implant problematic?
    Lignel, Arthur
    Berhouet, Julien
    Loirat, Marc-Alexandre
    Collin, Philippe
    Thomazeau, Herve
    Gallinet, David
    Boileau, Pascal
    Favard, Luc
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2018, 104 (06) : 773 - 777
  • [47] Biometrical analysis of the shoulder joint regarding glenoid implant dimensions for arthroplasty
    Jörn Kircher
    Bernd Bittersohl
    Christoph Zilkens
    Achim Hedtmann
    Rüdiger Krauspe
    Surgical and Radiologic Anatomy, 2014, 36 : 321 - 325
  • [48] Risk of Thromboembolism in Shoulder Arthroplasty: Effect of Implant Type and Traumatic Indication
    Navarro, Ronald A.
    Inacio, Maria C. S.
    Burke, Mary F.
    Costouros, John G.
    Yian, Edward H.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (05) : 1576 - 1581
  • [49] Common Factors in Shoulder and Hip Arthroplasty Implant Failures: A Historical Review
    Andrews, Reed
    Layuno-Matos, Josue G.
    Frankle, Mark A.
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (08)
  • [50] Gentamicin allergy as an unexpected 'hidden' cause of complications in knee arthroplasty
    Wittmann, Diana
    Summer, Burkhard
    Thomas, Benjamin
    Halder, Andreas
    Thomas, Peter
    CONTACT DERMATITIS, 2018, 78 (04) : 293 - 294