Amperometric hydrogen peroxide sensor using carbon electrode loaded with perovskite-type oxide

被引:2
|
作者
Takase, S [1 ]
Ishikawa, A [1 ]
Shimizu, Y [1 ]
机构
[1] Kyushu Inst Technol, Fac Engn, Dept Appl Chem, Kitakyushu, Fukuoka 8048550, Japan
关键词
H2O2; sensor; Perovskite-type oxide; amperometric sensing; solid polymer electrolyte;
D O I
10.5796/electrochemistry.69.272
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Amperometric sensing properties to H2O2 of the carbon electrodes loaded with various perovskite-type oxides have been investigated. It was found that the carbon electrodes loaded with perovskite-type oxides showed high activity to anodic reaction including H2O2. A limiting current was observed at the electrode potential between 0.7 and 1.0 V vs. SCE. and the current at a fixed potential increased with increasing H2O2 concentration between 10 and 1000 muM (M = mol dm(-3)). H2O2 responses of the elements loaded with La0.6Ca0.4BO3 (B = Fe and Cr) were hardly affected by dissolved oxygen. All solid-state H2O2 sensor element based on solid polymer electrolyte and La0.6Ca0.4FeO3-based electrodes also responded well to H2O2 at the concentration between 50 and 200 muM, with the 90% response time of as short as 20 s.
引用
收藏
页码:272 / 275
页数:4
相关论文
共 50 条
  • [41] Novel sensitive amperometric hydrogen peroxide sensor using layered hierarchical porous α-MoO3 and GO modified glass carbon electrode
    Li, Bo
    Song, Hai-Yan
    Deng, Zhao-Peng
    Huo, Li-Hua
    Gao, Shan
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2019, 288 : 641 - 648
  • [42] Ruthenium Oxide Hexacyanoferrate as an Effective Electrode Modifier for Amperometric Detection of Iodate and Hydrogen Peroxide
    Dodevska, Totka
    Shterev, Ivan
    Lazarova, Yanna
    Hadzhiev, Dobrin
    [J]. ACTA CHIMICA SLOVENICA, 2020, 67 (04) : 1216 - 1226
  • [43] NONSTOICHIOMETRY, DIFFUSION, AND ELECTRICAL-PROPERTIES OF PEROVSKITE-TYPE OXIDE ELECTRODE MATERIALS
    MIZUSAKI, J
    [J]. SOLID STATE IONICS, 1992, 52 (1-3) : 79 - 91
  • [44] A thick-film impedancemetric carbon monoxide sensor using layered perovskite-type cuprate
    Shimizu, Youichi
    Yamamoto, Shinyu
    Takase, Satoko
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2017, 249 : 667 - 672
  • [45] Perovskite-Type Oxide Thin Film Integrated Fiber Optic Sensor for High-Temperature Hydrogen Measurement
    Tang, Xiling
    Remmel, Kurtis
    Lan, Xinwei
    Deng, Jiangdong
    Xiao, Hai
    Dong, Junhang
    [J]. ANALYTICAL CHEMISTRY, 2009, 81 (18) : 7844 - 7848
  • [46] Tetravalent Dysprosium in a Perovskite-Type Oxide
    Han, Donglin
    Uda, Tetsuya
    Nose, Yoshitaro
    Okajima, Toshihiro
    Murata, Hidenobu
    Tanaka, Isao
    Shinoda, Kozo
    [J]. ADVANCED MATERIALS, 2012, 24 (15) : 2051 - 2053
  • [47] Bifunctional hydrogen peroxide electrode as an amperometric transducer for biosensors
    Gamburzev, S
    Atanasov, P
    Ghindilis, AL
    Wilkins, E
    Kaisheva, A
    Iliev, I
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1997, 43 (1-3) : 70 - 77
  • [48] Methylene Blue/Multiwall Carbon Nanotube Modified Electrode for the Amperometric Determination of Hydrogen Peroxide
    Pena, Roselyn C.
    Bertotti, Mauro
    Brett, Christopher M. A.
    [J]. ELECTROANALYSIS, 2011, 23 (10) : 2290 - 2296
  • [49] Catalytic destruction of dichloromethane using perovskite-type oxide catalysts
    Lou, JC
    Hung, CM
    Yang, BY
    [J]. JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 2004, 54 (06) : 727 - 740
  • [50] Sensitive Amperometric Sensing of Hydrogen Peroxide Using Ag Nanowire Array Electrode
    Kurowska-Tabor, E.
    Jaskula, M.
    Sulka, G. D.
    [J]. ELECTROANALYSIS, 2015, 27 (08) : 1968 - 1978