Optimizing MRI sequences and images for MRI-based stereotactic radiosurgery treatment planning

被引:13
|
作者
Taghizadeh, Somayeh [1 ,2 ,4 ,5 ]
Labuda, Cecille [4 ,5 ]
Yang, Claus Chunli [1 ]
Morris, Bart [1 ]
Kanakamedala, Madhava R. [1 ]
Vijayakumar, Srinivasan [1 ]
Rey-Dios, Roberto [1 ,3 ]
Duggar, William N. [1 ]
Florez, Edward [2 ]
Fatemi, Ali [1 ,2 ]
机构
[1] Univ Mississippi, Med Ctr, Dept Radiat Oncol, 2500 North State St, Jackson, MS 39216 USA
[2] Univ Mississippi, Med Ctr, Dept Radiol, Jackson, MS 39216 USA
[3] Univ Mississippi, Med Ctr, Dept Neurosurg, Jackson, MS 39216 USA
[4] Univ Mississippi, Natl Ctr Phys Acoust, Jackson, MS 38655 USA
[5] Univ Mississippi, Dept Phys & Astron, Jackson, MS 38655 USA
关键词
MRI; Stereotactic radiosurgery planning; Tumor; COMPLETE DISTORTION CORRECTION; POINT-SPREAD FUNCTION; GEOMETRIC DISTORTION;
D O I
10.1016/j.rpor.2018.09.010
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim: Development of MRI sequences and processing methods for the production of images appropriate for direct use in stereotactic radiosurgery (SRS) treatment planning. Background: MRI is useful in SRS treatment planning, especially for patients with brain lesions or anatomical targets that are poorly distinguished by CT, but its use requires further refinement. This methodology seeks to optimize MRI sequences to generate distortion-free and clinically relevant MR images for MRI-only SRS treatment planning. Materials and methods: We used commercially available SRS MRI-guided radiotherapy phantoms and eight patients to optimize sequences for patient imaging. Workflow involved the choice of correct MRI sequence(s), optimization of the sequence parameters, evaluation of image quality (artifact free and clinically relevant), measurement of geometrical distortion, and evaluation of the accuracy of our offline correction algorithm. Results: CT images showed a maximum deviation of 1.3 mm and minimum deviation of 0.4 mm from true fiducial position for SRS coordinate definition. Interestingly, uncorrected MR images showed maximum deviation of 1.2 mm and minimum of 0.4 mm, comparable to CT images used for SRS coordinate definition. After geometrical correction, we observed a maximum deviation of 1.1 mm and minimum deviation of only 0.3 mm. Conclusion: Our optimized MRI pulse sequences and image correction technique show promising results; MR images produced under these conditions are appropriate for direct use in SRS treatment planning. Published by Elsevier Sp. z o.o. on behalf of Greater Poland Cancer Centre.
引用
收藏
页码:12 / 19
页数:8
相关论文
共 50 条
  • [41] Impact of MRI resolution for Linac-based stereotactic radiosurgery
    Huang, Yimei
    Liang, Evan
    Schaff, Eric M.
    Zhao, Bo
    Snyder, Karen C.
    Chetty, Indrin J.
    Shah, Mira M.
    Siddiqui, Salim M.
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [42] USE OF MRI FUSION FOR CT BASED INTRACRANIAL STEREOTACTIC RADIOSURGERY
    Caglar, H. B.
    Altinok, A. Yildirim
    Kucukmorkoc, E.
    Acar, H.
    Kucuk, N.
    NEURO-ONCOLOGY, 2014, 16
  • [43] Feasibility of stereotactic MRI-based image guidance for the treatment of vascular malformations: a phantom study
    Marius Schwalbe
    Axel Haine
    Marc Schindewolf
    Hendrik von Tengg-Kobligk
    Tom Williamson
    Stefan Weber
    Iris Baumgartner
    Torsten Fuss
    International Journal of Computer Assisted Radiology and Surgery, 2016, 11 : 2207 - 2215
  • [44] Feasibility of stereotactic MRI-based image guidance for the treatment of vascular malformations: a phantom study
    Schwalbe, Marius
    Haine, Axel
    Schindewolf, Marc
    von Tengg-Kobligk, Hendrik
    Williamson, Tom
    Weber, Stefan
    Baumgartner, Iris
    Fuss, Torsten
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2016, 11 (12) : 2207 - 2215
  • [45] Stereotactic radiosurgery planning of vestibular schwannomas: Is MRI at 3 Tesla geometrically accurate?
    Schmidt, M. A.
    Wells, E. J.
    Davison, K.
    Riddell, A. M.
    Welsh, L.
    Saran, F.
    MEDICAL PHYSICS, 2017, 44 (02) : 375 - 381
  • [46] CT and MRI fusion to minimize contouring uncertainties in Stereotactic Radiosurgery (SRS) planning
    Payanova, I.
    Georgiev, R.
    Lazarov, R.
    Spasova, Z.
    Ivanov, V.
    Iliev, L.
    Tsenov, P.
    Antonova, A.
    Georgieva, T.
    Georgieva, S.
    Avgerinova, D.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S983 - S984
  • [47] Geometric accuracy of MRI for stereotactic radiosurgery planning of Acoustic Neuromas at 3 Tesla
    Schmidt, M. A.
    Wells, E.
    Davison, K.
    Riddell, A.
    Welsh, L.
    Saran, F.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S899 - S900
  • [48] MRI-based treatment planning for liver stereotactic body radiotherapy: validation of a deep learning-based synthetic CT generation method
    Liu, Yingzi
    Lei, Yang
    Wang, Tonghe
    Kayode, Oluwatosin
    Tian, Sibo
    Liu, Tian
    Patel, Pretesh
    Curran, Walter J.
    Ren, Lei
    Yang, Xiaofeng
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1100):
  • [49] MRI-Based Treatment Planning with Pseudo CT Generated Through Atlas Registration
    Uh, J.
    Merchant, T.
    Hua, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [50] Evaluation of MRI-based treatment planning for prostate cancer using the AcQPlan system
    Chen, L
    Price, R
    Li, J
    Wang, L
    Qin, L
    Ma, C
    MEDICAL PHYSICS, 2003, 30 (06) : 1507 - 1507