Effects of bismuth breast shielding on iodine quantification in dual-energy computed tomography: an experimental phantom study

被引:2
|
作者
Jang, Hyunsik [1 ]
Hur, Jin [1 ]
Choi, Byoung Wook [1 ]
Im, Dong Jin [1 ]
Hong, Yoo Jin [1 ]
Kim, Young Jin [1 ]
Lee, Hye-Jeong [1 ]
机构
[1] Yonsei Univ, Severance Hosp, Coll Med, Dept Radiol,Res Inst Radiol Sci, Seoul, South Korea
基金
新加坡国家研究基金会;
关键词
Bismuth; dual-energy computed tomography; iodine; radiation protection; radiation dosage; THORACIC CT; IONIZING-RADIATION; IMAGE QUALITY; TUBE-CURRENT; CANCER; REDUCTION;
D O I
10.1177/0284185118768098
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Although the bismuth breast shield can reduce radiation exposure to the breast during dual-energy computed tomography (DECT), it can potentially affect material quantification on DECT due to artifacts. Purpose: To evaluate the effects of bismuth breast shielding on iodine quantification and radiation exposure in DECT. Material and Methods: Small balloons were made with 0.2%, 0.6% and 1.0% blended iodinated contrast (370 mg/mL of iodine) with water. The balloons were located at both anterior and posterior lungs in an adult anthropomorphic chest phantom. DECT was performed with and without breast shielding. Afterwards, iodine concentration values were measured for each balloon on the iodine maps. Absorbed radiation doses in the breast were measured with the optically stimulated luminescence dosimeter. Results: After shielding, we obtained significantly decreased iodine quantification for all three concentrations with 0.78 +/- 0.13 to 0.46 +/- 0.13 mg/mL, 2.31 +/- 0.17 to 1.68 +/- 0.19 mg/mL, and 3.82 +/- 0.10 to 2.84 +/- 0.20 mg/mL at the anterior location, and 0.72 +/- 0.11 to 0.48 +/- 0.09 mg/mL, 2.24 +/- 0.13 to 1.87 +/- 0.21 mg/mL, and 3.75 +/- 0.16 to 3.15 +/- 0.14 mg/mL at the posterior location for the 0.2%, 0.6%, and 1.0% balloons, respectively (P = 0.001 for all). After shielding, absorbed radiation doses to the breast significantly decreased by 14.8% (4.32 +/- 0.33 to 3.68 +/- 0.30 mGy; P= 0.005). Conclusion: Although using the bismuth breast shield may decrease radiation exposure to the breast on DECT, it may also significantly affect iodine quantification.
引用
收藏
页码:1475 / 1481
页数:7
相关论文
共 50 条
  • [1] Accuracy of dual-energy computed tomography for the quantification of iodine in a soft tissue-mimicking phantom
    Li, Jung-Hui
    Du, Yeh-Ming
    Huang, Hsuan-Ming
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05): : 418 - 426
  • [2] Iodine Quantification Using Dual-Energy Computed Tomography for Differentiating Thymic Tumors
    Yan, Wei-Qiang
    Xin, Yong-Kang
    Jing, Yong
    Li, Gang-Feng
    Wang, Shu-Mei
    Rong, Wei-Cheng
    Xiao, Gang
    Lei, Xue-Bin
    Li, Bo
    Hu, Yu-Chuan
    Cui, Guang-Bin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 873 - 880
  • [3] Computation of Calcium Score With Dual-Energy Computed Tomography: A Phantom Study
    Kumar, Vidhya
    Min, James K.
    He, Xin
    Raman, Subha V.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (01) : 156 - 158
  • [4] Effects of scan parameters on the accuracies of iodine quantification and hounsfield unit values in dual layer dual-energy head and neck computed tomography: A phantom study conducted in a hospital in Japan
    Sakai, Y.
    Shirasaka, T.
    Hioki, K.
    Yamane, S.
    Kinoshita, E.
    Kato, T.
    RADIOGRAPHY, 2023, 29 (05) : 838 - 844
  • [5] Accuracy of iodine quantification using dual-energy computed tomography with focus on low concentrations
    Ahnfelt, Anders
    Dahlman, Par
    Segelsjo, Monica
    Magnusson, Mats O.
    Magnusson, Anders
    ACTA RADIOLOGICA, 2022, 63 (05) : 623 - 631
  • [6] Dual-Energy Computed Tomography Quantification of Extravasated Iodine and Hemorrhagic Transformation after Thrombectomy
    Baik, Minyoul
    Cha, Jihoon
    Ahn, Sung Soo
    Lee, Seung-Koo
    Kim, Young Dae
    Nam, Hyo Suk
    Jeon, Soyoung
    Lee, Hye Sun
    Heo, Ji Hoe
    JOURNAL OF STROKE, 2022, 24 (01) : 152 - +
  • [7] Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study
    Jacobsen, Megan C.
    Schellingerhout, Dawid
    Wood, Cayla A.
    Tamm, Eric P.
    Godoy, Myrna C.
    Sun, Jia
    Cody, Dianna D.
    RADIOLOGY, 2018, 287 (01) : 224 - 234
  • [8] Iodine Quantification With Dual-Energy CT: Phantom Study and Preliminary Experience With Renal Masses
    Chandarana, Hersh
    Megibow, Alec J.
    Cohen, Benjamin A.
    Srinivasan, Ramya
    Kim, Danny
    Leidecker, Christianne
    Macari, Michael
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) : W693 - W700
  • [9] Iodine Quantification Using Dual-Energy Multidetector Computed Tomography Imaging Phantom Study Assessing the Impact of Iterative Reconstruction Schemes and Patient Habitus on Accuracy
    Feuerlein, Sebastian
    Heye, Tobias J.
    Bashir, Mustafa R.
    Boll, Daniel T.
    INVESTIGATIVE RADIOLOGY, 2012, 47 (11) : 656 - 661
  • [10] Accessory spleen versus lymph node: Value of iodine quantification with dual-energy computed tomography
    Winklhofer, Sebastian
    Lin, Wei-Ching
    Lambert, Jack W.
    Yeh, Benjamin M.
    EUROPEAN JOURNAL OF RADIOLOGY, 2017, 87 : 53 - 58