摘要
Purpose The study aims at a novel dosimetry methodology to reconstruct a 3D dose distribution as imparted to a virtual cylindrical phantom using an electronic portal imaging device (EPID). Methods A deep learning-based signal processing strategy, referred to as 3DosiNet, is utilized to learn a mapping from an EPID image to planar dose distributions at given depths. The network was trained with the volumetric dose exported from the clinical treatment planning system (TPS). Given the latent inconsistency between measurements and corresponding TPS calculations, unsupervised learning is formulated in 3DosiNet to capture abstractive image features that are less sensitive to the potential variations. Results Validation experiments were performed using five regular fields and three clinical intensity-modulated radiation therapy (IMRT) cases. The measured dose profiles and percentage depth dose (PDD) curves were compared with those measured using standard tools in terms of the 1D gamma index. The mean gamma pass rates (2%/2 mm) over the regular fields are 100% and 97.3% for the dose profile and PDD measurements, respectively. The measured volumetric dose was compared to the corresponding TPS calculation in terms of the 3D gamma index. The mean 2%/2 mm gamma pass rates are 97.9% for square fields and 94.9% for the IMRT fields. Conclusions The system promises to be a practical 3D dosimetric tool for pre-treatment patient-specific quality assurance and further developed for in-treatment patient dose monitoring.
引用
收藏
页码:6810 / 6819
页数:10
相关论文
共 50 条
  • [21] EPID-based dosimetry and its relation to other 2D and 3D dose measurement techniques in radiation therapy
    Mijnheer, B.
    [J]. 9TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, 2017, 847
  • [22] 3D DOSE VERIFICATION FOR LUNG CANCER TREATMENTS BY IN AQUA VIVO EPID DOSIMETRY
    Wendling, M.
    McDermott, L.
    Mans, A.
    Sonke, J. J.
    van Herk, M.
    Mijnheer, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S160 - S160
  • [23] Error detection during VMAT delivery using EPID-based 3D transit dosimetry
    Mijnheer, B.
    Jomehzadeh, A.
    Gonzalez, P.
    Olaciregui-Ruiz, I
    Rozendaal, R.
    Shokrani, P.
    Spreeuw, H.
    Tielenburg, R.
    Mans, A.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 54 : 137 - 145
  • [24] A Novel Method for EPID-Based In-Vivo Exit Dose Verification for Intensity Modulated Radiotherapy
    Wu, Z.
    Wang, J.
    Peng, J.
    Wang, J.
    Hu, W.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3514 - 3515
  • [25] 3D EPID-Based in Vivo Dosimetry: A Major Step Forward Towards Optimal Quality and Safety in Radiation Oncology Practice
    Mijnheer, B.
    Mans, A.
    Olaciregui-Ruiz, I.
    Rozendaal, R.
    Spreeuw, H.
    van Herk, M.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 457 - 457
  • [26] Automated prioritised 3D dose-based MLC segment generation for step-and-shoot IMRT
    Schipaanboord, B. W. K.
    Breedveld, S.
    Rossi, L.
    Keijzer, M.
    Heijmen, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (16):
  • [27] Towards real-time EPID-based 3D in vivo dosimetry for IMRT with Deep Neural Networks: A feasibility study
    Martins, Juliana Cristina
    Maier, Joscha
    Gianoli, Chiara
    Neppl, Sebastian
    Dedes, George
    Alhazmi, Abdulaziz
    Veloza, Stella
    Reiner, Michael
    Belka, Claus
    Kachelriess, Marc
    Parodi, Katia
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 114
  • [28] EPID-Based 4D Patient Dose Verification for Lung SBRT Treatment
    Lin, M.
    Li, J.
    Wang, L.
    Ma, C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [29] A simple video-based EPID technique for pre-treatment verification of step-and-shoot intensity modulation
    Coughlin, M
    Leszczynski, K
    Provost, D
    [J]. MEDICAL PHYSICS, 2003, 30 (07) : 1940 - 1940
  • [30] 3D dose verification of breathing adapted radiotherapy using gel dosimetry
    Maansson, S.
    Gustavsson, H.
    Karlsson, A.
    Wittgren, L.
    Back, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S151 - S151