Validation of a flexible endoscopy simulator

被引:71
|
作者
Felsher, JJ
Olesevich, M
Farres, H
Rosen, M
Fanning, A
Dunkin, BJ
Marks, JM [1 ]
机构
[1] Case Western Reserve Univ, Dept Surg, Cleveland, OH 44106 USA
[2] Cleveland Clin Fdn, Minimally Invas Surg Ctr, Cleveland, OH 44195 USA
[3] Univ Miami, Sch Med, Dept Surg, Miami, FL 33152 USA
[4] Hillcrest Hosp, Mayfield Heights, OH 44124 USA
来源
AMERICAN JOURNAL OF SURGERY | 2005年 / 189卷 / 04期
关键词
simulators; colonoscopy; validation;
D O I
10.1016/j.amjsurg.2005.01.008
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Virtual reality (VR) simulation is a rapidly proliferating adjunct of surgical training. Numerous devices have evolved as educational tools in a variety of fields. Whether these tools can be used for validation of physicians' skills has yet to be determined. The objective of this study Was to determine whether the GI Mentor (Simbionix, Lod, Israel) flexible endoscopy simulator construct could distinguish experienced endoscopists from beginners. Methods: Seventy-five surgical attendings, fellows, and residents were recruited for participation in the study. Two cohorts were used and these groups were selected from 2 separate scientific sessions. Participants completed a standardized questionnaire documenting their endoscopic training and experience. Physicians subsequently were designated as experienced or beginner after their endoscopic training and experience were evaluated. All participants completed 1 of 2 colonoscopic simulations. The GI Mentor objectively evaluated performance on the basis of programmed data points, including the time to reach the cecum, the percentage of mucosa visualized, the completed polypectomy rate, the percentage of time spent in clear view through the lumen, the percentage of time that the patient was in pain, and overall efficiency. Results: In both simulations, experienced endoscopists were more efficient than beginners (.32%/s vs. .26%/s, P = .02; and .53%/s vs. .37%/s, P = .03) and achieved a greater polypectomy rate (78% vs. 43%, P = .03; and 87% vs. 48%, P = .01). Furthermore, experienced endoscopists visualized more of the colonic surface (86% vs. 82%, P = .02) and spent a greater proportion of the time in clear view of the lumen (55% vs. 47%, P = .05) than beginners completing the first simulation. In the second simulation, experienced participants reached the cecum more rapidly than beginners (175 vs. 262 s, P = .01). Conclusions: The GI Mentor VR colonoscopy construct appears valid. Significant performance differences were shown between the experienced and beginner cohorts. The beginner participants in this study were all physicians with some degree of endoscopic experience. Therefore, the GI Mentor distinguished endoscopists of varying experience and exposure. Further validation studies are needed to evaluate the breadth of programs inherent to this simulator and to determine whether if may be used in the future for qualification and certification purposes. (c) 2005 Excerpta Medica Inc. All rights reserved.
引用
收藏
页码:497 / 500
页数:4
相关论文
共 50 条
  • [1] GI endoscopy training: Validation of a computer-based endoscopy simulator
    Tse, Frances
    Liu, Louis
    Marshall, John K.
    Armstrong, David
    GASTROINTESTINAL ENDOSCOPY, 2007, 65 (05) : AB316 - AB316
  • [2] Development of Simulator Guidelines for Resident Assessment in Flexible Endoscopy
    Sutton, Erica
    Chase, Sheree Carter
    Klein, Rosemary
    Zhu, Yue
    Godinez, Carlos
    Youssef, Yassar
    Park, Adrian
    AMERICAN SURGEON, 2013, 79 (01) : 14 - 22
  • [3] Validation of a Realistic Simulator for Veterinary Gastrointestinal Endoscopy Training
    Uson-Gargallo, Jesus
    Uson-Casaus, Jesus M.
    Perez-Merino, Eva M.
    Soria-Galvez, Federico
    Morcillo, Esther
    Enciso, Silvia
    Sanchez-Margallo, Francisco M.
    JOURNAL OF VETERINARY MEDICAL EDUCATION, 2014, 41 (03) : 209 - 217
  • [4] Haptic display for all degrees of freedom of a simulator for flexible endoscopy
    Körner, O
    Rieger, K
    Männer, R
    MEDICAL SIMULATION, PROCEEDINGS, 2004, 3078 : 161 - 167
  • [5] Implementation of a haptic interface for a virtual reality simulator for flexible endoscopy
    Körner, O
    Männer, R
    11TH SYMPOSIUM ON HAPTIC INTERFACES FOR VIRTUAL ENVIRONMENT AND TELEOPERATOR SYSTEMS - HAPTICS 2003, PROCEEDINGS, 2003, : 278 - 284
  • [6] Fabrication and Validation of a Cost-Effective Upper Endoscopy Simulator
    Deere, Matthew J.
    Jones, Mark W.
    Jelinek, Leandra A.
    Henning, Werner H.
    JSLS-JOURNAL OF THE SOCIETY OF LAPAROENDOSCOPIC SURGEONS, 2020, 24 (04)
  • [7] Flexible endoscopy
    Hoetker, M. S.
    Malek, N. P.
    Goetz, M.
    GASTROENTEROLOGE, 2014, 9 (05): : 441 - 447
  • [8] Flexible endoscopy [Flexible endoskopie]
    Hoetker M.S.
    Malek N.P.
    Götz M.
    Der Gastroenterologe, 2014, 9 (5): : 441 - 447
  • [9] Evaluation of a virtual endoscopy simulator for training in gastrointestinal endoscopy
    Ferlitsch, A
    Glauninger, P
    Gupper, A
    Schillinger, M
    Haefner, M
    Gangl, A
    Schoefl, R
    ENDOSCOPY, 2002, 34 (09) : 698 - 702
  • [10] AN ENDOSCOPY SIMULATOR FOR TEACHING COLONOSCOPY
    WILLIAMS, CB
    GILLIES, DF
    HARITSIS, A
    GUT, 1991, 32 (10) : A1232 - A1232