Direct DC Voltages Comparison between two Programmable Josephson Voltage Standards at SCL

被引:0
|
作者
Yang, S. [1 ]
Cular, S. [2 ,3 ]
Rufenacht, A. [2 ,3 ]
Burroughs, C. J. [2 ,3 ]
Dresselhaus, P. D. [2 ,3 ]
Benz, S. P. [2 ,3 ]
Ng, M. N. [1 ]
机构
[1] Stand & Calibrat Lab, Hong Kong, Peoples R China
[2] NIST, Gaithersburg, MD 20899 USA
[3] NIST, Boulder, CO 80305 USA
关键词
Programmable Josephson voltage standard; quantum voltage measurement; measurement uncertainty; Josephson voltage comparison;
D O I
10.1109/cpem49742.2020.9191745
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Standards and Calibration Laboratory (SCL) setup in 2018 a liquid helium based programmable Josephson voltage standard (PJVS) developed by the National Institute of Standards and Technology (NIST). The system was validated by direct comparison with a NIST transportable PJVS system at 1.018 V, 4 V, 6 V, 8 V and 10 V. The difference between the systems was within 0.5 nV, with an expanded uncertainty of less than 2.2 nV (k = 2). In this paper, the setup and the results of the direct comparison method are presented.
引用
收藏
页数:2
相关论文
共 50 条
  • [31] Comparison of the Josephson voltage standards of the SP and the BIPM
    Reymann, D
    Witt, TJ
    Eklund, G
    Pajander, H
    Nilsson, H
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 220 - 223
  • [32] Comparison of the Josephson voltage standards of VNIIM and PTB
    Behr, R
    Niemeyer, J
    Katkov, A
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 251 - 252
  • [33] Comparison of the Josephson voltage standards of the CEM and the BIPM
    Reymann, D
    Witt, TJ
    Balmisa, J
    Castejón, P
    Pérez, S
    METROLOGIA, 1999, 36 (01) : 59 - 62
  • [34] Comparison of the Josephson voltage standards of VNIIM and PTB
    Behr, R
    Niemeyer, J
    Katkov, AS
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (02) : 203 - 205
  • [35] NPL BIPM COMPARISON OF JOSEPHSON VOLTAGE STANDARDS
    HENDERSON, LCA
    REYMANN, D
    WITT, TJ
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (10) : 1011 - 1013
  • [36] Comparison of the Josephson voltage standards of the IEN and the BIPM
    Reymann, D
    Witt, TJ
    Andreone, D
    Cerri, R
    Godone, A
    METROLOGIA, 1998, 35 (01) : 21 - 24
  • [37] Comparison of the Josephson voltage standards of the MIKES and the BIPM
    Solve, Stephane
    Chayramy, Regis
    Immonen, Pekkra
    Nissila, Jaani
    Stock, Michael
    Manninen, Antti
    METROLOGIA, 2021, 58 (1A)
  • [38] Design and fabrication of 10 VSINIS Josephson arrays for programmable voltage standards
    Schulze, H
    Behr, R
    Kohlmann, J
    Müller, F
    Niemeyer, J
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2000, 13 (09): : 1293 - 1295
  • [39] Cryocooled Programmable and Pulse-driven Josephson Voltage Standards at INRiM
    Durandetto, Paolo
    Monticone, Eugenio
    Trinchera, Bruno
    Serazio, Danilo
    Sosso, Andrea
    2017 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2017, : 791 - 795
  • [40] Practical Operation of Cryogen-Free Programmable Josephson Voltage Standards
    Schwall, R. E.
    Zilz, D. P.
    Power, J.
    Burroughs, C. J.
    Dresselhaus, P. D.
    Benz, S. P.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2011, 21 (03) : 891 - 895