A magnetic stereo-actuation mechanism for active capsule endoscope

被引:12
|
作者
Wang, Xiaona [1 ]
Meng, Max Q. -H. [1 ]
机构
[1] Chinese Univ Hong Kong, Dept Elect Engn, Hong Kong, Hong Kong, Peoples R China
关键词
D O I
10.1109/IEMBS.2007.4352913
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The commercial available capsule endoscope lacks a self-actuation mechanism and the passive locomotion mode results in a low diagnosis efficiency or even missing of the diseases. An actuation mechanism is necessary for an active capsule endoscope to carry out thorough examinations and further operations efficiently. In this paper, after measuring the force needed for driving the capsules in the small intestine, we introduced a wireless stereo-actuation mechanism taking advantage of magnetic force and torque. A small permanent magnet was enclosed into the capsule as a seed to be actuated by multiple current coils placed around the abdomen area. An optimization algorithm was designed for selection of the most economical coil currents. Experiments and results show the current level for six orthogonal coils of reasonable size to actuate the capsule in the pre-determined position and direction.
引用
收藏
页码:2811 / 2814
页数:4
相关论文
共 50 条
  • [21] Resistance model of an active capsule endoscope in a peristaltic intestine
    Huang, Yi
    Liang, Liang
    Tang, Puhua
    Guo, Zhiming
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2024, 238 (05) : 529 - 536
  • [22] NON-INVASIVE ACTIVE CAPSULE ENDOSCOPE INTEGRATED TARGETING BIOPSY FUNCTION BASED ON ELECTRO-MANGETIC ACTUATION SYSTEM
    Viet Ha Le
    Kang, Byungjeon
    Kim, Jayoung
    Lee, Kyungmin
    Manh Cuong Hoang
    Van Du Nguyen
    Choi, Eunpyo
    Kim, Chang-Sei
    Park, Jong-Oh
    [J]. 2017 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND BIOMIMETICS (IEEE ROBIO 2017), 2017, : 2309 - 2313
  • [23] Stopping mechanism for capsule endoscope using electrical stimulus
    Woo, Sang Hyo
    Kim, Tae Wan
    Cho, Jin Ho
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2010, 48 (01) : 97 - 102
  • [24] Stopping mechanism for capsule endoscope using electrical stimulus
    Sang Hyo Woo
    Tae Wan Kim
    Jin Ho Cho
    [J]. Medical & Biological Engineering & Computing, 2010, 48 : 97 - 102
  • [25] Dipole modeling of magnetic marker for capsule endoscope localization
    Wang, Xiaona
    Meng, Max Q. -H.
    [J]. WCICA 2006: SIXTH WORLD CONGRESS ON INTELLIGENT CONTROL AND AUTOMATION, VOLS 1-12, CONFERENCE PROCEEDINGS, 2006, : 5382 - +
  • [26] Experimental Investigation of the Spiral Structure of a Magnetic Capsule Endoscope
    Yang, Wanan
    Li, Zhen
    He, Yong
    Dai, Houde
    Wu, Zhouxing
    [J]. INTERNATIONAL JOURNAL OF ADVANCED ROBOTIC SYSTEMS, 2016, 13
  • [27] The application of intelligent capsule endoscope with magnetic guided attitude
    Du, Fan Du
    Du, Tie-Yi Yang
    Huang, Qin-Ya Huangya
    Du, Ling Yao
    [J]. JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2013, 28 : 26 - 26
  • [28] A wireless tattooing capsule endoscope using external electromagnetic actuation and chemical reaction pressure
    Hoang, Manh Cuong
    Le, Viet Ha
    Kim, Jayoung
    Choi, Eunpyo
    Kang, Byungjeon
    Park, Jong-Oh
    Kim, Chang-Sei
    [J]. PLOS ONE, 2019, 14 (07):
  • [29] Closed Loop Control of a Tethered Magnetic Capsule Endoscope
    Taddese, Addisu Z.
    Slawinski, Piotr R.
    Obstein, Keith L.
    Valdastri, Pietro
    [J]. ROBOTICS: SCIENCE AND SYSTEMS XII, 2016,
  • [30] Medical Microrobot - A Drug Delivery Capsule Endoscope with Active Locomotion and Drug Release Mechanism: Proof of Concept
    Kim Tien Nguyen
    Manh Cuong Hoang
    Choi, Eunpyo
    Kang, Byungjeon
    Park, Jong-Oh
    Kim, Chang-Sei
    [J]. INTERNATIONAL JOURNAL OF CONTROL AUTOMATION AND SYSTEMS, 2020, 18 (01) : 65 - 75