Comparative accuracy of optical sensor-based wearable system for non-invasive measurement of blood glucose concentration

被引:21
|
作者
Rodin, Dmitry [1 ]
Kirby, Michael [1 ]
Sedogin, Natallia [3 ]
Shapiro, Yair [2 ]
Pinhasov, Albert [1 ]
Kreinin, Anatoly [3 ]
机构
[1] Ariel Univ, Dept Mol Biol, Ariel, Israel
[2] Ariel Univ, Dept Hlth Syst Management, Ariel, Israel
[3] Bruce Rappaport Med Fac, Maale Carmel Mental Hlth Ctr, Technion, Tirat Carmel, Israel
关键词
Diabetes; Glucose level; Non-invasive glucometry; Optical biosensor;
D O I
10.1016/j.clinbiochem.2018.12.014
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Non-invasive biosensors for indirect evaluation of routinely-measured blood components by sweat analysis have broad potential clinical applications. This trial tested a wrist-borne non-invasive glucose monitor (NIGM) to measure blood glucose (BG) levels using photoplethysmographic (PPG) optical sensors. Our aim was to determine the accuracy of the device in comparison with a standard, invasive clinical method for blood glucose monitoring. Adult participants (n = 200) of both sexes from 18 to 75 were recruited for the study. Exclusion criteria: hemophilia and other serious coagulation disorders, impaired venous access, other serious medical conditions. A biosensor was placed on the right wrist of each participant for a non-invasive indirect BG measurement. In parallel, blood from the antecubital vein was collected and glucose levels were assessed with YSI 2300 Bioanalyzer. The measurements were performed twice: before and after food intake, with a 1-h interval between measurements. There were no limitations to food type and quantity. In both anteprandial (rho = 0.8994, p < 0.0001) and postprandial (rho = 0.9382, p < 0.0001) glucose measurements, NIGM correlated with values obtained by the YSI 2300 reference device - there was no significant difference between the two methods. Plotted on a Parkes Error Grid for Type II diabetes, NIGM readings did not deviate from those of the YSI 2300 in any clinically-significant way, with the majority of correlated readings falling within Parkes zone A. Very few readings fell within Parkes zone B. In anteprandial measurements, the mean bias between methods for all patient volunteers was 3.705 +/- 7.838. In postprandial measurements gave a mean bias of 1.362 +/- 10.15 for all patient glucose data. The mean absolute relative difference of currently available glucometer models ranges from 5.6% to 20.8%. The NIGM falls in the lower end of this error range at 7.40-7.54%, indicating that PPG-chemochrome sensors are capable of producing results comparable with those of direct measure glucometers. Data presented here demonstrates the reliability and accuracy of the NIGM system as an adjunctive, and perhaps substitutive, non-invasive tool for blood glucose monitoring.
引用
收藏
页码:15 / 20
页数:6
相关论文
共 50 条
  • [21] Non-invasive Detection System Design for Blood Glucose Based on Optical Rotation Properties
    Liu, Shengping
    Li, Ersong
    Zhou, Qi
    [J]. 2010 3RD INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2010), VOLS 1-7, 2010, : 1532 - 1535
  • [22] Verification of Non-Invasive Blood Glucose Measurement Method Based on Pulse Wave Signal Detected by FBG Sensor System
    Kurasawa, Shintaro
    Koyama, Shouhei
    Ishizawa, Hiroaki
    Fujimoto, Keisaku
    Chino, Shun
    [J]. SENSORS, 2017, 17 (12)
  • [23] An Optimized Non-Invasive Blood Glucose and Temperature Body Measurement System
    Ibrahim, Hagar
    Darweesh, M. Saeed
    Soltan, Ahmed
    [J]. 2022 11TH INTERNATIONAL CONFERENCE ON MODERN CIRCUITS AND SYSTEMS TECHNOLOGIES (MOCAST), 2022,
  • [24] A novel non-invasive wearable sensor for intraocular pressure measurement
    Campigotto, Angelica
    Lai, Yongjun
    [J]. Medical Devices and Sensors, 2020, 3 (03):
  • [25] Comparative Study of Different Measurement Sites using NIR Based Non-invasive Glucose Measurement system
    Yadav, Jyoti
    Rani, Asha
    Singh, Viajnder
    Murari, Bhaskar Moahn
    [J]. PROCEEDINGS OF THE 4TH INTERNATIONAL CONFERENCE ON ECO-FRIENDLY COMPUTING AND COMMUNICATION SYSTEMS, 2015, 70 : 469 - 475
  • [26] Towards a Wearable Non-invasive Blood Glucose Monitoring Device
    Andrews, Joseph Thomas
    Solanki, J.
    Choudhary, Om P.
    Chouksey, S.
    Malvia, N.
    Chaturvedi, P.
    Sen, P.
    [J]. INTERNATIONAL CONFERENCE ON RECENT TRENDS IN PHYSICS (ICRTP 2012), 2012, 365
  • [27] New biosensor for non-invasive glucose concentration measurement
    Gourzi, M
    Rouane, A
    McHugh, MB
    Guelaz, R
    Nadi, M
    [J]. PROCEEDINGS OF THE IEEE SENSORS 2003, VOLS 1 AND 2, 2003, : 1343 - 1347
  • [28] The discussion about the complexity of samples and the prediction accuracy of glucose concentration in non-invasive blood glucose sensing
    Gu, XY
    Liu, R
    Chen, WL
    Xu, KX
    [J]. Complex Dynamics and Fluctuations in Biomedical Photonics II, 2005, 5696 : 222 - 231
  • [29] Comparative sensor analysis for an electronic wearable and non-invasive respiratory signal acquisition system
    Lopez-Meyer, Paulo
    Sazonov, Edward
    Cheung, Brian
    [J]. 2012 SIXTH INTERNATIONAL CONFERENCE ON SENSING TECHNOLOGY (ICST), 2012, : 805 - 808
  • [30] Light Wavelength for Measurement of Non-Invasive Blood Glucose Level with Optical Device
    Jenie, Renan Prasta
    Nurdin, Naufal Muharram
    Husein, Irzaman
    Alatas, Husin
    [J]. ANNALS OF NUTRITION AND METABOLISM, 2019, 75 : 187 - 187