Evaluation of spatial resolution in image acquisition by optical flatbed scanners for radiochromic film dosimetry

被引:7
|
作者
Asero, G. [1 ,2 ]
Greco, C. [3 ,4 ]
Gueli, A. M. [1 ,2 ,4 ]
Raffaele, L. [3 ,4 ]
Spampinato, S. [1 ,2 ,4 ]
机构
[1] Univ Catania, Dipartimento Fis & Astron, PH3DRA Labs PHys Dating Diagnost Dosimetry Res &, Via Santa Sofia 64, I-95123 Catania, Italy
[2] Ist Nazl Fis Nucl, Via Santa Sofia 64, I-95123 Catania, Italy
[3] Azienda Osped Univ Policlin Vittorio Emanuele, Via Santa Sofia 78, I-95123 Catania, Italy
[4] Univ Catania, Scuola Fac Med, Dipartimento Sci Med Chirurg & Tecnol Avanzate GF, Scuola Specializzaz Fis Med, Via Santa Sofia 78, I-95123 Catania, Italy
来源
关键词
Image reconstruction in medical imaging; Data processing methods; Dosimetry concepts and apparatus; X-ray radiography and digital radiography (DR); XR TYPE-T; ENERGY RESPONSE;
D O I
10.1088/1748-0221/11/03/P03024
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Introduction: Radiochromic films are two-dimensional dosimeters that do not require developing and give values of absorbed dose with accuracy and precision. Since this dosimeter colours directly after irradiation, it can be digitized with commercial optical flatbed scanners to obtain a calibration curve that links blackening of the film with dose. Although the film has an intrinsic high spatial resolution, the scanner determines the actual resolution of this dosimeter, in particular the "dot per inch" (dpi) parameter. The present study investigates the effective spatial resolution of a scanner used for Gafchromic (R) XR-QA2 film (designed for radiology Quality Assurance) analysis. Material and methods: The quantitative evaluation of the resolution was performed with the Modulation Transfer Function (MTF) method, comparing the nominal resolution with the experimental one. The analysis was performed with two procedures. First, the 1951 USAF resolution test chart, a tool that tests the performance of optical devices, was used. Secondly, a combined system of mammography X-ray tube, XR-QA2 film and a bar pattern object was used. In both cases the MTF method has been applied and the results were compared. Results: The USAF and the film images have been acquired with increasing dpi and a standard protocol for radiochromic analysis, to evaluate horizontal and vertical and resolution. The effective resolution corresponds to the value of the MTF at 50%. In both cases and for both procedures, it was verified that, starting from a dpi value, the effective resolution saturates. Conclusion: The study found that, for dosimetric applications, the dpi of the scanner have to be adjusted to a reasonable value because, if too high, it requires high scanning and computational time without providing additional information.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] EVALUATION OF A GLASSLESS FLATBED SCANNER FOR RADIOCHROMIC DYE FILM DOSIMETRY OF CLINICAL X-RAYS BEAMS
    Villarreal-Barajas, J. E.
    Lisa, G.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S145 - S145
  • [22] RADIOCHROMIC FILM DOSIMETRY USING A COMMERCIAL FLATBED DOCUMENT SCANNER: A FEASIBILITY STUDY
    Adamczyk, S.
    Litoborski, M.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S216 - S216
  • [23] Reference radiochromic film dosimetry in kilovoltage photon beams during CBCT image acquisition
    Tomic, Nada
    Devic, Slobodan
    DeBlois, Francois
    Seuntjens, Jan
    MEDICAL PHYSICS, 2010, 37 (03) : 1083 - 1092
  • [24] Comparison of the Epson Perfection V700 Photo Flatbed and the Vidar VXR-16 Dosimetry PRO AdvantageTM Film Scanners for Use with Radiochromic Film
    Matney, J.
    Parker, B.
    Neck, D.
    Rosen, I.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [25] Reference Dosimetry During CBCT Acquisition Using Radiochromic Film
    Tomic, N.
    Seuntjens, J.
    DeBlois, F.
    Devic, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [26] The origin of the flatbed scanner artifacts in radiochromic film dosimetry-key experiments and theoretical descriptions
    Schoenfeld, Andreas A.
    Wieker, Soeren
    Harder, Dietrich
    Poppe, Bjoern
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (21): : 7704 - 7724
  • [27] Characterization of Radiochromic Film GAFCHROMIC XR-RV2 by Reflection Densitometers, Transmission and Flatbed Scanners
    Lunelli, N.
    Rochas, M. S.
    Paz-Viera, J. E.
    Khoury, H. J.
    5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 535 - 538
  • [28] Lateral Optical Density Variation in Flatbed Scanners in Combination with Gafchromic Film
    van Battum, L. J.
    Heukelom, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 328 - 328
  • [29] Spatial resolution of a stacked radiochromic film dosimeter
    Butson, MJ
    Cheung, T
    Yu, PKN
    RADIOTHERAPY AND ONCOLOGY, 2001, 61 (02) : 211 - 213
  • [30] Temperature Effect During Radiation Exposure Period for Radiochromic Film Dosimetry Using Flatbed Scanner Densitometer
    Ren, T.
    Bourland, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E608 - E608