Four-dimensional radiotherapy planning for DMLC-based respiratory motion tracking

被引:236
|
作者
Keall, PJ
Joshi, S
Vedam, SS
Siebers, JV
Kini, VR
Mohan, R
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
[2] Univ N Carolina, Dept Radiat Oncol, Chapel Hill, NC 27514 USA
[3] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
[4] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
4D radiotherapy; deformable image registration; treatment planning;
D O I
10.1118/1.1879152
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Four-dimensional (4D) radiotherapy is the explicit inclusion of the temporal changes in anatomy during the imaging, planning, and delivery of radiotherapy. Temporal anatomic changes can occur for many reasons, though the focus of the current investigation is respiration motion for lung tumors. The aim of this study was to develop 4D radiotherapy treatment-planning methodology for DMLC-based respiratory motion tracking. A 4D computed tomography (CT) scan consisting of a series of eight 3D CT image sets acquired at different respiratory phases was used for treatment planning. Deformable image registration was performed to map each CT set from the peak-inhale respiration phase to the CT image sets corresponding to subsequent respiration phases. Deformable registration allows the contours defined on the peak-inhale CT to be automatically transferred to the other respiratory phase CT image sets. Treatment planning was simultaneously performed on each of the eight 3D image sets via automated scripts in which the MLC-defined beam aperture conforms to the PTV (which in this case equaled the GTV due to CT scan length limitations) plus a penumbral margin at each respiratory phase. The dose distribution from each respiratory phase CT image set was mapped back to the peak-inhale CT image set for analysis. The treatment intent of 4D planning is that the radiation beam defined by the DMLC tracks the respiration-induced target motion based on a feedback loop including the respiration signal to a real-time MLC controller. Deformation with respiration was observed for the lung tumor and normal tissues. This deformation was verified by examining the mapping of high contrast objects, such as the lungs and cord, between image sets. For the test case, dosimetric reductions for the cord, heart, and lungs were found for 4D planning compared with 3D planning. 4D radiotherapy planning for DMLC-based respiratory motion tracking is feasible and may offer tumor dose escalation and/or a reduction in treatment-related complications. However, 4D planning requires new planning tools, such as deformable registration and automated treatment planning on multiple CT image sets. (c) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:942 / 951
页数:10
相关论文
共 50 条
  • [41] Research on four-dimensional motion planning of on-orbit servicing autonomous spacecraft
    Wang, Ping
    Guo, Ji-Feng
    Shi, Xiao-Ning
    Cui, Nai-Gang
    Kongzhi yu Juece/Control and Decision, 2010, 25 (10): : 1519 - 1522
  • [42] Simultaneous Tracking and Four-Dimensional Radiotherapy Delivery: Accounting for Spatial and Morphological Tumor Changes
    Keall, P.
    Cho, B.
    Poulsen, P.
    Sawant, A.
    Suh, Y.
    Venkat, R.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [43] A novel four-dimensional radiotherapy method for lung cancer: imaging, treatment planning and delivery
    Alasti, H.
    Cho, Y. B.
    Vandermeer, A. D.
    Abbas, A.
    Norrlinger, B.
    Shubbar, S.
    Bezjak, A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (12): : 3251 - 3267
  • [44] Motion encoded beamlets for optimization and evaluation in four-dimensional (4D) radiotherapy
    Lu, W.
    Olivera, G.
    Chen, Q.
    Chen, M.
    Pan, Y.
    Schnarr, E.
    Ruchala, K.
    MEDICAL PHYSICS, 2006, 33 (06) : 2042 - 2043
  • [45] Experimental evaluation of four-dimensional Magnetic Resonance Imaging for radiotherapy planning of lung cancer
    Perkins, Terry
    Lee, Danny
    Simpson, John
    Greer, Peter
    Goodwin, Jonathan
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 32 - 35
  • [46] Both four-dimensional computed tomography and four-dimensional cone beam computed tomography under-predict lung target motion during radiotherapy
    Steiner, Elisabeth
    Shieh, Chun-Chien
    Caillet, Vincent
    Booth, Jeremy
    O'Brien, Ricky
    Briggs, Adam
    Hardcastle, Nicholas
    Jayamanne, Dasantha
    Szymura, Kathryn
    Eade, Thomas
    Keall, Paul
    RADIOTHERAPY AND ONCOLOGY, 2019, 135 : 65 - 73
  • [47] Determination of respiratory motion for distal esophagus cancer using four-dimensional computed tomography
    Yaremko, Brian P.
    Guerrero, Thomas M.
    McAleer, Mary F.
    Bucci, M. Kara
    Noyola-Martinez, Josue
    Nguyen, Linda T.
    Balter, Peter A.
    Guerra, Rudy
    Komaki, Ritsuko
    Liao, Zhongxing
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (01): : 145 - 153
  • [48] Four-Dimensional Respiratory Motion-Resolved Whole Heart Coronary MR Angiography
    Piccini, Davide
    Feng, Li
    Bonanno, Gabriele
    Coppo, Simone
    Yerly, Jerome
    Lim, Ruth P.
    Schwitter, Juerg
    Sodickson, Daniel K.
    Otazo, Ricardo
    Stuber, Matthias
    MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (04) : 1473 - 1484
  • [49] Multi-Objective Four-Dimensional Vehicle Motion Planning in Large Dynamic Environments
    Wu, Paul P. -Y.
    Campbell, Duncan
    Merz, Torsten
    IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS PART B-CYBERNETICS, 2011, 41 (03): : 621 - 634
  • [50] A four-dimensional random motion at finite speed
    Kolesnik, Alexander D.
    JOURNAL OF APPLIED PROBABILITY, 2006, 43 (04) : 1107 - 1118