Magnesium and calcium determination in desalted crude oil by direct sampling graphite furnace atomic absorption spectrometry

被引:14
|
作者
Seeger, Tassia S. [1 ]
Muller, Edson I. [1 ]
Mesko, Marcia F. [2 ]
Duarte, Fabio A. [1 ]
机构
[1] Univ Fed Santa Maria, Dept Quim, BR-97105900 Santa Maria, RS, Brazil
[2] Univ Fed Pelotas, Ctr Ciencias Quim Farmaceut & Alimentos, BR-96160000 Capao Do Leao, RS, Brazil
关键词
Direct sampling; Graphite furnace atomic absorption spectrometry; Magnesium; Calcium; Petroleum; MICROWAVE-INDUCED COMBUSTION; PLASMA-MASS SPECTROMETRY; ICP-MS; ELEMENT DETERMINATION; CHLORIDE SALTS; TRACE-ELEMENTS; DIGESTION; EXTRACTION; PETROLEUM; MG;
D O I
10.1016/j.fuel.2018.09.108
中图分类号
TE [石油、天然气工业]; TK [能源与动力工程];
学科分类号
0807 ; 0820 ;
摘要
Direct sampling graphite furnace atomic absorption spectrometry (DS-GF AAS) was proposed for Mg and Ca determination in desalted crude oil. A Zeeman-effect background correction system with variable magnetic field was used and measurements were carried out with low and high sensitivity. Pyrolysis and atomization temperatures, calibration with aqueous standards and the effect of sample mass to be introduced into the atomizer were evaluated. The use of chemical modifiers was not necessary for both elements. For Mg determination, two sensitivity conditions (high and low, respectively) were used: i) main absorption line (285.2 nm), 2-field mode (0.8 T) and gas stop flow during the atomization and ii) secondary absorption line (202.6 nm), 2-field mode (0.8 T) and gas stop flow during the atomization. For Ca determination, three low-sensitivity conditions using the main absorption line (422.7 nm) and variable magnetic field were used: i) 3-field dynamic mode (0.6-0.8 T) and gas stop flow during the atomization, ii) 3-field dynamic mode (0.4-0.8 T) and gas stop flow during the atomization and iii) 3-field dynamic mode (0.6-0.8 T) and gas flow during the atomization. The characteristic mass and limit of quantification for the highest sensitivity condition were 0.3 pg and 1.2 ng g(-1) for Mg and 4.8 pg and 0.1 mu g g(-1) for Ca, respectively. For all sensitivity conditions, good linearity (R-2 above 0.997) and relative standard deviation below 14% were achieved. Accuracy for DS-GF AAS method was evaluated by results comparison with those obtained by GF AAS after microwave-assisted digestion, which were in agreement. Additionally, recovery tests were carried out and recoveries for both analytes ranged from 92 to 107%. The use of DS-GF AAS provides some advantages such as simplicity, low risk of contamination and up to twelve measurements can be performed per hour.
引用
收藏
页码:1483 / 1488
页数:6
相关论文
共 50 条
  • [1] Direct sampling graphite furnace atomic absorption spectrometry - feasibility of Na and K determination in desalted crude oil
    Seeger, Tassia S.
    Machado, Eduarda Q.
    Flores, Erico M. M.
    Mello, Paola A.
    Duarte, Fabio A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2018, 141 : 28 - 33
  • [2] Determination of arsenic and cadmium in crude oil by direct sampling graphite furnace atomic absorption spectrometry
    de Jesus, Alexandre
    Zmozinski, Ariane Vanessa
    Ferreira Damin, Isabel Cristina
    Silva, Marcia Messias
    Rodrigues Vale, Maria Goreti
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2012, 71-72 : 86 - 91
  • [3] Comparison of direct sampling and emulsion analysis using a filter furnace for the determination of lead in crude oil by graphite furnace atomic absorption spectrometry
    Damin, Isabel C. F.
    Dessuy, Morgana B.
    Castilhos, Tamara S.
    Silva, Marcia M.
    Vale, Maria Goreti R.
    Welz, Bernhard
    Katskov, Dmitri A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2009, 64 (06) : 530 - 536
  • [4] Direct determination of lead in smoke from burned crude oil by impaction graphite furnace atomic absorption spectrometry
    Noble, CO
    Hammond, JL
    Beck, JN
    Proffitt, CE
    Sneddon, J
    MICROCHEMICAL JOURNAL, 1997, 57 (03) : 361 - 369
  • [5] Simultaneous determination of Cr, Fe, Ni and V in crude oil by emulsion sampling graphite furnace atomic absorption spectrometry
    Luz, Maciel Santos
    Oliveira, Pedro Vitoriano
    ANALYTICAL METHODS, 2011, 3 (06) : 1280 - 1283
  • [6] Direct determination of nickel in petroleum by solid sampling–graphite furnace atomic absorption spectrometry
    Geisamanda Pedrini Brandão
    Reinaldo Calixto de Campos
    Eustáquio Vinicius Ribeiro de Castro
    Honério Coutinho de Jesus
    Analytical and Bioanalytical Chemistry, 2006, 386 : 2249 - 2253
  • [7] Determination of trace elements in paints by direct sampling graphite furnace atomic absorption spectrometry
    Bentlin, Fabrina R. S.
    Pozebon, Dirce
    Mello, Paola A.
    Flores, Trico M. M.
    ANALYTICA CHIMICA ACTA, 2007, 602 (01) : 23 - 31
  • [8] Solid sampling determination of magnesium in lithium niobate crystals by graphite furnace atomic absorption spectrometry
    Dravecz, Gabriella
    Laczai, Nikoletta
    Hajdara, Ivett
    Bencs, Laszlo
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2016, 126 : 1 - 5
  • [9] DIRECT DETERMINATION OF CALCIUM IN METALS BY ATOMIC-ABSORPTION SPECTROMETRY WITH A GRAPHITE-FURNACE
    ISHIBASHI, W
    KIKUCHI, R
    YAMAMOTO, K
    BUNSEKI KAGAKU, 1979, 28 (06) : 394 - 399
  • [10] Direct determination of iron in sand using solid sampling graphite furnace atomic absorption spectrometry
    Cíntia S. Silva
    Cassiana S. Nomura
    Joaquim A. Nóbrega
    Pedro V. Oliveira
    Microchimica Acta, 2008, 161 : 109 - 114