A Simple Method for In Vivo Measurement of Implant Rod Three-Dimensional Geometry During Scoliosis Surgery

被引:4
|
作者
Salmingo, Remel A. [2 ]
Tadano, Shigeru [1 ]
Fujisaki, Kazuhiro [1 ]
Abe, Yuichiro [3 ]
Ito, Manabu [4 ]
机构
[1] Hokkaido Univ, Fac Engn, Div Human Mech Syst & Design, Kita Ku, Sapporo, Hokkaido 0608628, Japan
[2] Hokkaido Univ, Grad Sch Engn, Div Human Mech Syst & Design, Kita Ku, Sapporo, Hokkaido 0608628, Japan
[3] Eniwa Hosp, Dept Orthopaed Surg, Eniwa 0611449, Japan
[4] Hokkaido Univ, Grad Sch Med, Dept Orthopaed Surg, Kita Ku, Sapporo, Hokkaido 0608638, Japan
关键词
Scoliosis; implant fixation; three-dimensional rod geometry; projective geometry; biomechanics; bone; INSTRUMENTATION;
D O I
10.1115/1.4006687
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Scoliosis is defined as a spinal pathology characterized as a three-dimensional deformity of the spine combined with vertebral rotation. Treatment for severe scoliosis is achieved when the scoliotic spine is surgically corrected and fixed using implanted rods and screws. Several studies performed biomechanical modeling and corrective forces measurements of scoliosis correction. These studies were able to predict the clinical outcome and measured the corrective forces acting on screws, however, they were not able to measure the intraoperative three-dimensional geometry of the spinal rod. In effect, the results of biomechanical modeling might not be so realistic and the corrective forces during the surgical correction procedure were intra-operatively difficult to measure. Projective geometry has been shown to be successful in the reconstruction of a three-dimensional structure using a series of images obtained from different views. In this study, we propose a new method to measure the three-dimensional geometry of an implant rod using two cameras. The reconstruction method requires only a few parameters, the included angle theta between the two cameras, the actual length of the rod in mm, and the location of points for curve fitting. The implant rod utilized in spine surgery was used to evaluate the accuracy of the current method. The three-dimensional geometry of the rod was measured from the image obtained by a scanner and compared to the proposed method using two cameras. The mean error in the reconstruction measurements ranged from 0.32 to 0.45 mm. The method presented here demonstrated the possibility of intra-operatively measuring the three-dimensional geometry of spinal rod. The proposed method could be used in surgical procedures to better understand the biomechanics of scoliosis correction through real-time measurement of three-dimensional implant rod geometry in vivo. [DOI: 10.1115/1.4006687]
引用
收藏
页数:5
相关论文
共 50 条
  • [31] Geometry-related Treatments for Three-dimensional Meshless Method
    Lee, Ming-Hsiao
    Chen, Wen-Hwa
    CMES-COMPUTER MODELING IN ENGINEERING & SCIENCES, 2010, 61 (03): : 249 - 271
  • [32] Calibration Method for Line-Structured Light Three-Dimensional Measurement Based on a Simple Target
    Li, Xuexing
    Zhang, Wenhui
    Song, Guanglei
    PHOTONICS, 2022, 9 (04)
  • [33] Computational Simulation of Three-dimensional Tumor Geometry during Radiotherapy
    Takao, S.
    Tadano, S.
    Taguchi, H.
    Shirato, H.
    13TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, VOLS 1-3, 2009, 23 (1-3): : 2083 - +
  • [34] Determination of observing optical geometry for three-dimensional measurement of objects in liquid
    Kaneko, T
    Nakayama, D
    Kubo, T
    OPTICS IN COMPUTING 98, 1998, 3490 : 453 - 456
  • [35] Analysis Method of Sucker Rod Centralizer Spacing in Three-dimensional Hole
    Ren, Tao
    Nan, Zhenjiang
    Sun, Wen
    Kang, Xiaoqing
    PROCEEDINGS OF THE 5TH INTERNATIONAL CONFERENCE ON MECHANICAL ENGINEERING, MATERIALS AND ENERGY (5TH ICMEME2016), 2016, 90 : 163 - 167
  • [36] A three-dimensional measurement method for medical electric endoscope
    Zhou, Tingting
    Tao, Pei
    Yuan, Bo
    Wang, Liqiang
    INTERNATIONAL CONFERENCE ON INNOVATIVE OPTICAL HEALTH SCIENCE, 2017, 0245
  • [37] Active stereo method for three-dimensional shape measurement
    Wang, Heng
    Hu, Jiasheng
    OPTICAL ENGINEERING, 2012, 51 (06)
  • [38] Color pattern projection method for three-dimensional measurement
    Wakayama, Toshitaka
    Yoshizawa, Toru
    DIMENSIONAL OPTICAL METROLOGY AND INSPECTION FOR PRACTICAL APPLICATIONS, 2011, 8133
  • [39] Method for three-dimensional measurement of dynamic deformable surfaces
    Shi, Jinlong
    Liu, Ye
    Chen, Yan Qiu
    JOURNAL OF ELECTRONIC IMAGING, 2012, 21 (03)
  • [40] Three-dimensional friction measurement during hip simulation
    Sonntag, Robert
    Braun, Steffen
    Al-Salehi, Loay
    Reinders, Joern
    Mueller, Ulrike
    Kretzer, J. Philippe
    PLOS ONE, 2017, 12 (09):