COMPARISON ON TOTAL LUMINOUS FLUX MEASUREMENT OF SPECTROGONIOPHOTOMETER AND GONIOPHOTOMETER

被引:0
|
作者
Zhao, W. [1 ]
Hui, L. [1 ]
Jian, L. [2 ]
Haisu, Z. [1 ]
Tong, C. [3 ]
机构
[1] Natl Inst Metrol, Beijing, Peoples R China
[2] Beijing Normal Univ, Beijing 100875, Peoples R China
[3] Xiamen Inst Measurement & Testing, Xiamen, Peoples R China
关键词
spectrogoniophotometer; spectroradiometer; LED measurement;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A facility of spectrogoniophotometer in NIM has been setup, and we make a comparison between the spectrogoniophotometer and a traditional goniophotometer. The differences on total luminous flux measurement are similar to 0.1% for tungsten lamp and similar to 2% for LED (except blue). The facility shows the spectral mismatch correction factor F-star(phi,theta) spatial distribution for common white LED light source is nonuniform and we recommend to introduce this factor to the usage of a traditional goniophotometer for LED light source.
引用
收藏
页码:634 / 636
页数:3
相关论文
共 50 条
  • [41] Bilateral comparison of luminous flux (EURAMET.PR-K4.3)
    Pons, Alicia
    Campos, Joaquin
    Sametoglu, Ferhat
    METROLOGIA, 2019, 56 (1A)
  • [42] Estimation of the sampling interval error for LED measurement with a goniophotometer
    Zhao Weiqiang
    Liu Hui
    Liu Jian
    INTERNATIONAL CONFERENCE ON OPTICS IN PRECISION ENGINEERING AND NANOTECHNOLOGY (ICOPEN2013), 2013, 8769
  • [43] Study on LED'S luminous flux measurement using a compound parabolic concentrator
    Zhou, Xiao-Li
    Liu, Mu-Qing
    Zhu, Shao-Long
    Shen, Hai-Ping
    Guangdianzi Jiguang/Journal of Optoelectronics Laser, 2009, 20 (09): : 1137 - 1140
  • [44] The study of luminous flux method for rapid measurement of micro-aperture area
    Ming Ye
    Xu Dongjing
    Ni Zhiqiang
    MECHATRONICS AND INDUSTRIAL INFORMATICS, PTS 1-4, 2013, 321-324 : 671 - 675
  • [45] Loss formula on luminous flux of deep slit diffraction and the measurement on slit width
    Song, Wei
    Yingyong Jiguang/Applied Laser Technology, 2000, 20 (04): : 185 - 186
  • [47] A new approach to preparation of standard LEDs for luminous intensity and flux measurement of LEDs
    Park, Seung-Nam
    Park, Seongchong
    Lee, Dong-Hoon
    ADVANCED LEDS FOR SOLID STATE LIGHTING, 2006, 6355
  • [48] Traceability of Total Spectral Radiant Flux (TSRF) Scale Using Spectral Irradiance and Total Luminous Flux Scale at CSIR-NPL, India
    Manchalwar, Naval
    Saha, Shibu
    Jaiswal, V. K.
    Sharma, Parag
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2023, 38 (03): : 629 - 640
  • [49] Traceability of Total Spectral Radiant Flux (TSRF) Scale Using Spectral Irradiance and Total Luminous Flux Scale at CSIR-NPL, India
    Naval Vijeta
    Shibu Manchalwar
    V. K. Saha
    Parag Jaiswal
    MAPAN, 2023, 38 : 629 - 640
  • [50] THE IMPLEMENTATION AND ERROR ANALYSIS OF THE NOVEL DEVICE FOR LED'S TOTAL LUMINOUS FLUX MEASURING
    Gan, Yuanyuan
    Zhou, Xiaoli
    Liu, Muqing
    LIGHT & ENGINEERING, 2013, 21 (03): : 36 - 42