Effect of Wall Exhaust and Spill Locations on Indoor Air Quality in a Chemical Laboratory

被引:0
|
作者
Morcos, Ayman A. Shaaban Samy M. [1 ]
Khalil, Essam E. [1 ]
Fouad, Mahmoud A. [1 ]
机构
[1] Cairo Univ, Fac Engn, Mech Power Dept, Giza 12211, Egypt
来源
关键词
CONTAMINANT DISTRIBUTION; MIXING VENTILATION; ROOM AIR; ENVIRONMENTS; INLET;
D O I
暂无
中图分类号
O414.1 [热力学];
学科分类号
摘要
Indoor air quality inside chemical laboratories subjected to gaseous contaminants was investigated numerically throughout the current research using Ansys Fluent 13. The lab is 4.8 m (L) * 4.3 m (W) * 2.73 m (H). The model was built and mesh was generated using Gambit 2.2.30 yielding around 1.4 million cells. To ensure the reliability of the Computational Fluid Dynamics (CFD) model validation was done against experimental data of three cases done by Jin et al. [1]. The model could simulate accurately contaminant mole fraction to the order of 10(-7) with average absolute error of 9% using Re-normalization group (RNG) k-epsilon turbulence model. Parametric study is done to investigate the effect of wall exhaust and spill locations on the efficiency of the ventilation system. Wall exhaust location had strong impact on the contaminant distribution within the laboratory. Using wall exhaust in same half where the spill occurs was a very efficient solution for contaminant removal in accidents. The spill location did not have strong impact on the indoor air quality inside chemical laboratories. Average absolute deviation percentage in the average contaminant concentration in the studied cases because of the change of the spill location is 5% with maximum deviation percentage of 14%. This was attributed to the good mixing achieved by high Air Changes per Hour (ACH) normally used in laboratories (6 - 12).
引用
收藏
页数:8
相关论文
共 50 条
  • [11] Impact of condensate cooled air purging on indoor air quality in an air conditioned laboratory
    Palanisamy, Dhamodharan
    Ayalur, Bakthavatsalam Kannappan
    BUILDING AND ENVIRONMENT, 2021, 188
  • [12] Effect of quality and maintenance of the ventilation to the indoor air quality
    Niskanen, J
    Valkeapää, A
    Anttonen, H
    PROGRESS IN MODERN VENTILATION, VOL 2, PROCEEDINGS, 2000, : 168 - 171
  • [13] Indoor air quality modelling and measurements of a studio apartment with a mechanical exhaust system
    Szczepanik-Scislo, Nina
    Flaga-Maryanczyk, Agnieszka
    10TH CONFERENCE ON INTERDISCIPLINARY PROBLEMS IN ENVIRONMENTAL PROTECTION AND ENGINEERING EKO-DOK 2018, 2018, 44
  • [14] Microbiological and chemical quality of indoor air in kindergartens in Slovenia
    Rejc, Tanja
    Kukec, Andreja
    Bizjak, Mirko
    GodicTorkar, Karmen
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL HEALTH RESEARCH, 2020, 30 (01) : 49 - 62
  • [15] TAP WATER AND INDOOR AIR CONTAMINATION DUE TO AN UNINTENTIONAL CHEMICAL SPILL IN SOURCE WATER
    Omur-Ozbek, Pinar
    Akalp, Duygu
    Whelton, Andrew
    PROCEEDINGS OF INTERNATIONAL STRUCTURAL ENGINEERING AND CONSTRUCTION: INTERACTION BETWEEN THEORY AND PRACTICE IN CIVIL ENGINEERING AND CONSTRUCTION, 2016, : 597 - 602
  • [16] Analysis of Indoor Air Quality in Different Exhaust Conditions in Residential Buildings in Winter
    Wang, Yueren
    Wu, Nan
    Qi, Wei
    PROGRESS IN ENVIRONMENTAL SCIENCE AND ENGINEERING, PTS 1-4, 2013, 610-613 : 2823 - 2826
  • [17] Studies of possible chemical emissions from PVC articles used in indoor applications and the effect on indoor air quality
    Howick, CJ
    McCarthy, SA
    JOURNAL OF VINYL & ADDITIVE TECHNOLOGY, 1996, 2 (02): : 134 - 142
  • [18] PARTIAL PIT EXHAUST IMPROVES INDOOR AIR QUALITY AND EFFECTIVENESS OF AIR CLEANING IN LIVESTOCK HOUSING: A REVIEW
    Zhang, G.
    Bjerg, B.
    Zong, C.
    APPLIED ENGINEERING IN AGRICULTURE, 2017, 33 (02) : 243 - 256
  • [19] Effect of air purifier on indoor air quality and atopic dermatitis
    Park, Hyun Chul
    Kim, Young Hoon
    Kim, Jeong Eun
    Ko, Joo Yeon
    Goung, Sun Ju Nam
    Lee, Cheol Min
    Kim, Yoon Shin
    Ro, Young Suck
    ALLERGY ASTHMA & RESPIRATORY DISEASE, 2013, 1 (03): : 248 - 256
  • [20] The control of nitrogen oxides from motor vehicle exhaust for improved indoor air quality
    Muller, Chris
    PROCEEDINGS OF THE 5TH INTERNATIONAL SYMPOSIUM ON HEATING, VENTILATING AND AIR CONDITIONING, VOLS I AND II, 2007, : 384 - 388