Repeatability of intraocular pressure and corneal biomechanical properties measurements by the ocular response analyser

被引:33
|
作者
Kynlgopoulos, M. [1 ]
Schlote, T. [1 ]
Kotecha, A. [2 ]
Tzamalis, A. [1 ]
Pajic, B. [1 ]
Haefliger, I. [3 ]
机构
[1] Clin Pallas, Dept Ophthalmol, CH-4600 Olten, Switzerland
[2] City Univ London, Dept Optometry, London EC1V 0HB, England
[3] Univ Eye Clin, Lab Ocular Pharmacol & Physiol, Basel, Switzerland
关键词
ocular response analyzer; corneal hysteresis; corneal resistance factor; corneal compensated IOP;
D O I
10.1055/s-2008-1027256
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: The Ocular Response Analyser (ORA, Reichert Ophthalmic Instruments) is a non-contact applanation tonometer, providing two measures of intraocular pressure (IOP) - IOPg which represents a Goldmann equivalent IOP measure and IOPcc, representing a measure of IOP independent of corneal effects. In addition, the device provides two measures believed to represent corneal biomechanical properties: corneal hysteresis (CH) and corneal resistance factor (CRF). The aim of this study was to assess the repeatability of these measurements. Patients and Methods: One randomly chosen eye from 49 healthy volunteers was measured four times consecutively with the ORA prior to Goldmann applanation tonometry (GAT). The repeatability coefficient (RC), the coefficient of variation (CV) and the intraclass correlation coefficient (ICC) were calculated as a measure of intrasession repeatability. Results: CH was the most variable and IOPg the most repeatable measure, with an RC of 2.61 and 1.97, respectively, and ICC of 0.86 and 0.92, respectively. CV ranged between 5.73% for IOPg and 12.38% for CH. ORA IOP measurements were higher than GAT (IOPcc=17.43 +/- 3.23; IOPg = 17.53 +/- 3.0; GAT = 15.75 +/- 2.77 mmHg). Conclusions: ORA measurements show good short-term repeatability in normal volunteers. Thus, this device appears to be applicable in clinical practice.
引用
收藏
页码:357 / 360
页数:4
相关论文
共 50 条
  • [1] Intraocular pressure and biomechanical corneal properties measure by ocular response analyser in patients with primary congenital glaucoma
    Perucho-Gonzalez, Lucia
    Martinez de la Casa, Jose Maria
    Morales-Fernandez, Laura
    Baneros-Rojas, Paula
    Saenz-Frances, Federico
    Garcia-Feijoo, Julian
    ACTA OPHTHALMOLOGICA, 2016, 94 (05) : e293 - e297
  • [2] Evaluation of the influence of corneal biomechanical properties on Intraocular pressure measurements using the Ocular Response Analyzer
    Medeiros, Felipe A.
    Weinreb, Robert N.
    JOURNAL OF GLAUCOMA, 2006, 15 (05) : 364 - 370
  • [3] Corneal biomechanical properties measured with the Ocular Response Analyser in a myopic population
    Plakitsi, Athina
    O'Donnell, Clare
    Miranda, Marco A.
    Charman, W. Neil
    Radhakrishnan, Hema
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2011, 31 (04) : 404 - 412
  • [4] Assessment of corneal biomechanical properties and intraocular pressure with the Ocular Response Analyzer in childhood myopia
    Chang, Pei-Yao
    Chang, Shu-Wen
    Wang, Jiun-Yi
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (07) : 877 - 881
  • [5] The influence of the tear film on the intraocular pressure and the corneal biomechanical properties analyzed with the Ocular Response Analyzer
    Da Silva, Filipe
    Linhares, Joao M. M.
    Lira, Madalena
    JOURNAL OF OPTOMETRY, 2024, 17 (02)
  • [6] Biomechanical properties of the cornea measured by the Ocular Response Analyzer and their association with intraocular pressure and the central corneal curvature
    Franco, Sandra
    Lira, Madalena
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2009, 92 (06) : 469 - 475
  • [7] Ocular response analyser measurements and central corneal thickness in ocular rosacea patients
    Palamar, Melis
    Degirmenci, Cumali
    Ertam, Ilgen
    Yagci, Ayse
    INTERNATIONAL OPHTHALMOLOGY, 2017, 37 (01) : 7 - 11
  • [8] Ocular response analyser measurements and central corneal thickness in ocular rosacea patients
    Melis Palamar
    Cumali Degirmenci
    Ilgen Ertam
    Ayse Yagci
    International Ophthalmology, 2017, 37 : 7 - 11
  • [9] The Effect of Corneal Edema on Ocular Response Analyser and Goldmann Tonometry Measurements
    Pye, D. C.
    Lau, W. M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [10] The Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using a Rebound Self-tonometer
    Brown, Lyndsay
    Foulsham, William
    Pronin, Savva
    Tatham, Andrew J.
    JOURNAL OF GLAUCOMA, 2018, 27 (06) : 511 - 518