MR-based truncation and attenuation correction in integrated PET/MR hybrid imaging using HUGE with continuous table motion

被引:32
|
作者
Lindemann, Maike E. [1 ]
Oehmigen, Mark [1 ]
Blumhagen, Jan O. [2 ]
Gratz, Marcel [1 ,3 ]
Quick, Harald H. [1 ,3 ]
机构
[1] Univ Duisburg Essen, Univ Hosp Essen, High Field & Hybrid MR Imaging, Essen, Germany
[2] Siemens Healthcare GmbH, Magnet Resonance, Erlangen, Germany
[3] Univ Duisburg Essen, Erwin L Hahn Inst Magnet Resonance Imaging, Essen, Germany
关键词
attenuation correction; continuous table motion; FOV extension; PET quantification; truncation correction; whole-body PET/MR; WHOLE-BODY PET/MR; EMISSION TOMOGRAPHY/COMPUTED TOMOGRAPHY; QUANTITATIVE-EVALUATION; SYSTEMS; PHANTOM; CANCER;
D O I
10.1002/mp.12449
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The objective of this study was to introduce and evaluate a method for MR-based attenuation and truncation correction in phantom and patient measurements to improve PET quantification in PET/MR hybrid imaging. Methods: The fully MR-based method HUGE (B0 Homogenization using gradient enhancement) provides field-of-view extension in MR imaging, which can be used for truncation correction and improved PET quantification in PET/MR hybrid imaging. The HUGE method in this recent implementation is combined with continuously moving table data acquisition to provide a seamless nontruncated whole-body data set of the outer patient contours to complete the established standard MRbased Dixon-VIBE data for attenuation correction. The method was systematically evaluated in NEMA standard phantom experiments to investigate the impact of HUGE truncation correction on PET quantification. The method was then applied to 24 oncologic patients in whole-body PET/MR hybrid imaging. The impact of MR-based truncation correction with HUGE on PET data was compared to the impact of the established PET-based MLAA algorithm for contour detection. Results: In phantom and in all patient measurements, the standard Dixon-VIBE attenuation correction data show geometric distortions and signal truncations at the edges of the MR imaging field-ofview. In contrast, the Dixon-VIBE-based attenuation correction data additionally extended by applying HUGE shows significantly less distortion and truncations and due to the continuously moving table acquisition robustly provides smooth outer contours of the patient arms. In the investigated patient cases, MLAA frequently showed an overestimation of arm volume and associated artifacts limiting contour detection. When applying HUGE, an average relative increase in SUVmean in patients' lesion of 4.2% and for MLAA of 4.6% were measured, when compared to standard DixonVIBE only. In specific lesions maximal differences in SUVmean up to 13% for HUGE and 14% for MLAA were measured. Quantification in truncated regions showed maximal differences up to 40% for both, MLAA and HUGE. Average differences in those regions in SUVmean for HUGE are 13.3% and 14.6% for MLAA. In a patient with I-124 radiotracer PET-based MLAA contour detection completely failed in this specific case, whereas HUGE as MR-based approach provided accurate truncation correction. Conclusions: The HUGE method for truncation correction combined with continuous table movement extends the lateral MR field-of-view and effectively reduces truncations along the outer contours of the patient's arms in whole-body PET/MR imaging. HUGE as a fully MR-based approach is independent of the choice of radiotracer, thus also offering robust truncation correction in patients that are not injected with Fluordesoxyglucose (FDG) as radiotracer. Therefore, this method improves the standard Dixon MR-based attenuation correction and PET image quantification in whole-body PET/MR imaging applications. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4559 / 4572
页数:14
相关论文
共 50 条
  • [1] MR-based truncation correction using an advanced HUGE method to improve attenuation correction in PET/MR imaging of obese patients
    Lindemann, Maike E.
    Gratz, Marcel
    Blumhagen, Jan Ole
    Jakoby, Bjoern
    Quick, Harald H.
    MEDICAL PHYSICS, 2022, 49 (02) : 865 - 877
  • [2] MR-Based PET Attenuation Correction for PET/MR Imaging
    Bezrukov, Ilja
    Mantlik, Frederic
    Schmidt, Holger
    Schoelkopf, Bernhard
    Pichler, Bernd J.
    SEMINARS IN NUCLEAR MEDICINE, 2013, 43 (01) : 45 - 59
  • [3] Investigation of MR-Based Attenuation Correction and Motion Compensation for Hybrid PET/MR
    Buerger, Christian
    Tsoumpas, Charalampos
    Aitken, Andrew
    King, Andrew Peter
    Schleyer, Paul
    Schulz, Volkmar
    Marsden, Paul K.
    Schaeffter, Tobias
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (05) : 1967 - 1976
  • [4] MR-based correction of respiratory motion in oncological PET imaging using an integrated PET/MR system
    Wuerslin, Christian
    Schmidt, Holger
    Schwenzer, Nina
    Martirosian, Petros
    Claussen, Claus
    Stegger, Lars
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [5] Influence and Compensation of Truncation Artifacts in MR-Based Attenuation Correction in PET/MR
    Schramm, G.
    Langner, J.
    Hofheinz, F.
    Petr, J.
    Lougovski, A.
    Beuthien-Baumann, B.
    Platzek, I.
    van den Hoff, J.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (11) : 2056 - 2063
  • [6] Field of view extension and truncation correction for MR-based human attenuation correction in simultaneous MR/PET imaging
    Blumhagen, Jan O.
    Braun, Harald
    Ladebeck, Ralf
    Fenchel, Matthias
    Faul, David
    Scheffler, Klaus
    Quick, Harald H.
    MEDICAL PHYSICS, 2014, 41 (02)
  • [7] A hybrid method for generation of attenuation map for MR-based attenuation correction of PET data in prostate PET/MR imaging
    Mehdi Shirin Shandiz
    Mohammad Hadi Arabi
    Pardis Ghafarian
    Mehrdad Bakhshayesh Karam
    Hamidreza Saligheh Rad
    Mohammad Reza Ay
    EJNMMI Physics, 1 (Suppl 1)
  • [8] Errors in MR-based attenuation correction for brain imaging with PET/MR scanners
    Kops, Elena Rota
    Herzog, Hans
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 702 : 104 - 107
  • [9] MR-based attenuation correction for hybrid PET-MR brain imaging systems using deformable image registration
    Schreibmann, Eduard
    Nye, Jonathon A.
    Schuster, David M.
    Martin, Diego R.
    Votaw, John
    Fox, Tim
    MEDICAL PHYSICS, 2010, 37 (05) : 2101 - 2109
  • [10] Optimization Of MR-Based Attenuation Correction For Cardiac PET/MR
    Vergani, V.
    Robson, P.
    Trivieri, M.
    Fayad, Z. A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S699 - S700