Investigation of MR-Based Attenuation Correction and Motion Compensation for Hybrid PET/MR

被引:29
|
作者
Buerger, Christian [1 ]
Tsoumpas, Charalampos [1 ]
Aitken, Andrew [1 ]
King, Andrew Peter [1 ]
Schleyer, Paul [1 ]
Schulz, Volkmar [2 ]
Marsden, Paul K. [1 ]
Schaeffter, Tobias [1 ]
机构
[1] Kings Coll London, Kings Hlth Partners, St Thomas Hosp, Dept Biomed Engn,Div Imaging Sci & Biomed Engn, London SE1 7EH, England
[2] Philips Res, Mol Imaging Syst, D-22335 Aachen, Germany
基金
英国工程与自然科学研究理事会; 英国惠康基金;
关键词
Attenuation correction; cortical bone; motion correction; motion model; MR; PET simulation; ultrashort echo time; RESPIRATORY MOTION; IMAGE REGISTRATION; CT IMAGES; RECONSTRUCTION; FIELDS; MODEL;
D O I
10.1109/TNS.2012.2209127
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In hybrid PET/MR systems, attenuation maps can be derived from MR to correct for attenuation in PET. However, MR-based attenuation correction (AC) in abdominal applications remains challenging (i) because of poor signal from important tissue types in common MR sequences (e. g., cortical bone) and (ii) because of respiratory motion which results in misalignments between the derived attenuation maps and the PET emissions. Furthermore, respiratory motion also leads to motion-blurring artefacts in the final PET reconstructions. In this paper, we compute an MR-based 4D attenuation map including cortical bone by combining an Ultrashort Echo Time (UTE) acquisition with a subject-specific motion model derived from a second near real-time 3D MR image acquisition. This model allows us to create attenuation maps at any respiratory position which are used for AC in the reconstruction of different respiratory resolved PET images. The inverse of the model is used for motion compensation (MC) of these images. We demonstrate our approach on MR data from 5 healthy volunteers including 3 manually inserted artificial lesions. The impact of bone tissue and respiratory motion on AC is investigated in PET simulations (i) by misclassifying bone to soft tissue in the attenuation maps leading to errors of up to 26.0% in mean uptake for lesions close to bone, and (ii) by using a non-moving attenuation map leading to errors of up to 24.2%. The impact of respiratory motion on MC showed errors of up to 50.7% in areas of strong motion if MC was not performed. The results show that the effect of motion has to be considered both for attenuation correction and for motion-compensating PET emissions. This additive effect of motion is larger than the effect of a wrong AC.
引用
收藏
页码:1967 / 1976
页数:10
相关论文
共 50 条
  • [1] Influence and Compensation of Truncation Artifacts in MR-Based Attenuation Correction in PET/MR
    Schramm, G.
    Langner, J.
    Hofheinz, F.
    Petr, J.
    Lougovski, A.
    Beuthien-Baumann, B.
    Platzek, I.
    van den Hoff, J.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (11) : 2056 - 2063
  • [2] MR-Based PET Attenuation Correction for PET/MR Imaging
    Bezrukov, Ilja
    Mantlik, Frederic
    Schmidt, Holger
    Schoelkopf, Bernhard
    Pichler, Bernd J.
    [J]. SEMINARS IN NUCLEAR MEDICINE, 2013, 43 (01) : 45 - 59
  • [3] Dynamic MR-based Respiratory Motion Compensation for Hybrid PET/MR System
    Miao, Shun
    Liao, Rui
    Moran, Gerald
    Butler, John
    Pan, Li
    Wang, Z. Jane
    [J]. PROCEEDINGS OF THE 2014 9TH IEEE CONFERENCE ON INDUSTRIAL ELECTRONICS AND APPLICATIONS (ICIEA), 2014, : 1915 - +
  • [4] Optimization Of MR-Based Attenuation Correction For Cardiac PET/MR
    Vergani, V.
    Robson, P.
    Trivieri, M.
    Fayad, Z. A.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S699 - S700
  • [5] A hybrid method for generation of attenuation map for MR-based attenuation correction of PET data in prostate PET/MR imaging
    Mehdi Shirin Shandiz
    Mohammad Hadi Arabi
    Pardis Ghafarian
    Mehrdad Bakhshayesh Karam
    Hamidreza Saligheh Rad
    Mohammad Reza Ay
    [J]. EJNMMI Physics, 1 (Suppl 1)
  • [6] MR-based truncation and attenuation correction in integrated PET/MR hybrid imaging using HUGE with continuous table motion
    Lindemann, Maike E.
    Oehmigen, Mark
    Blumhagen, Jan O.
    Gratz, Marcel
    Quick, Harald H.
    [J]. MEDICAL PHYSICS, 2017, 44 (09) : 4559 - 4572
  • [7] Accelerated MR-Based Motion Field Measurement for PET Motion Correction in PET/MR
    Han, Paul
    Djebra, Yanis
    Petibon, Yoann
    Marin, Thibault
    Ouyang, Jinsong
    El Fakhri, Georges
    Ma, Chao
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [8] MR-based motion compensation in simultaneous cardiac PET-MR
    Petibon, Yoann
    Ouyang, Jinsong
    Zhu, Xuping
    Reese, Timothy
    Dagher, Joseph
    Catana, Ciprian
    El Fakhri, Georges
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [9] Errors in MR-based attenuation correction for brain imaging with PET/MR scanners
    Kops, Elena Rota
    Herzog, Hans
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 702 : 104 - 107
  • [10] MR-based Attenuation Correction for Whole-body PET/MR System
    Hu, Z.
    Renisch, S.
    Schweizer, B.
    Blaffert, T.
    Ojha, N.
    Guo, T.
    Tang, J.
    Tung, C.
    Kaste, J.
    Schulz, V.
    Torres, I.
    Shao, L.