Comparison of three videolaryngoscopes: Pentax Airway Scope, C-MAC™, Glidescope® vs the Macintosh laryngoscope for tracheal intubation

被引:73
|
作者
Teoh, W. H. L. [1 ]
Saxena, S. [2 ]
Shah, M. K. [1 ]
Sia, A. T. H. [1 ]
机构
[1] KK Womens & Childrens Hosp, Dept Womens Anaesthesia, Singapore, Singapore
[2] Minist Hlth Holdings Pte Ltd, Singapore, Singapore
关键词
ENDOTRACHEAL INTUBATION; DIFFICULT LARYNGOSCOPY; VIDEO LARYNGOSCOPE; LARYNGEAL VIEW; OBESE-PATIENTS; STYLET USE; PERFORATION; MANNEQUINS; BLADE; SCORE;
D O I
10.1111/j.1365-2044.2010.06513.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
P>We compared the intubating characteristics of the Pentax Airway Scope, the C-MAC (TM) and the Glidescope (R) with those of the Macintosh laryngoscope in 400 patients without predictors of difficult intubation. We found shorter intubation times with the Airway Scope (mean (SD) 20.6 (11.5) s) compared with the C-MAC (31.9 (17.6) s) and Glidescope (31.2 (15.0) s), p < 0.001. The median (IQR [range]) score for ease of intubation was significantly lower (better) with the Airway Scope (0 (0-8.75 [0-60])) than with the C-MAC (10 (0-20 [0-90])) or Glidescope (0 (0-20 [0-80])), p < 0.001. Ease of blade insertion in the C-MAC (0 (0-0 [0-30])) was superior to that of the Airway Scope (0 (0-0 [0-70])), Macintosh (0 (0-0 [0-80])) or Glidescope (0 (0-10 [0-60])), p = 0.006. More patients had a grade 1 laryngeal view with the Airway Scope (97%) compared with the C-MAC (87%), Glidescope (78%), or Macintosh (58%), p < 0.001. There were no grade 3 laryngeal views with the Airway Scope. The three videolaryngoscopes had comparable first attempt successful intubation rates: 95% for the Airway Scope, 93% for the C-MAC and 91% for the Glidescope.
引用
收藏
页码:1126 / 1132
页数:7
相关论文
共 50 条
  • [1] A Randomized Comparison Simulating Face to Face Endotracheal Intubation of Pentax Airway Scope, C-MAC Video Laryngoscope, Glidescope Video Laryngoscope, and Macintosh Laryngoscope
    Choi, Hyun Young
    Oh, Young Min
    Kang, Gu Hyun
    Kang, Hyunggoo
    Jang, Yong Soo
    Kim, Wonhee
    Kim, Euichung
    Cho, Young Soon
    Choi, Hyukjoong
    Kim, Hyunjong
    Kim, Gyoung Yong
    [J]. BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [2] Comparison of the Pentax Airway Scope® and McGrath® Videolaryngoscope with the Macintosh laryngoscope in tracheal intubation by anaesthetists unfamiliar with videolaryngoscopes: a manikin study
    Sharma, D. -J.
    Weightman, W. M.
    Travis, A.
    [J]. ANAESTHESIA AND INTENSIVE CARE, 2010, 38 (01) : 39 - 42
  • [3] A Comparison of the GlideScope® Video Laryngoscope to the C-MAC® Video Laryngoscope for Tracheal Intubation in the Emergency Department
    Sakles, J. C.
    Mosier, J.
    Chiu, S.
    Patanwala, A.
    [J]. ANNALS OF EMERGENCY MEDICINE, 2012, 60 (04) : S63 - S63
  • [4] Tracheal intubation with restricted access: a randomised comparison of the Pentax-Airway Scope and Macintosh laryngoscope in a manikin
    Asai, T.
    [J]. ANAESTHESIA, 2009, 64 (10) : 1114 - 1117
  • [5] Randomized comparison of the Pentax AirWay Scope and Macintosh laryngoscope for tracheal intubation in patients with obstructive sleep apnoea
    Kim, M. K.
    Park, S. W.
    Lee, J. W.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2013, 111 (04) : 662 - 666
  • [6] A comparison of the Macintosh laryngoscope, McGrath video laryngoscope, and Pentax Airway Scope in paediatric nasotracheal intubation
    Ji Young Yoo
    Yun Jeong Chae
    Young Bok Lee
    Sujin Kim
    Jaemoon Lee
    Dae Hee Kim
    [J]. Scientific Reports, 8
  • [7] Comparison of the GlideScope® with the Macintosh laryngoscope for tracheal intubation in patients with simulated difficult airway
    Lim, Y
    Yeo, SW
    [J]. ANAESTHESIA AND INTENSIVE CARE, 2005, 33 (02) : 243 - 247
  • [8] A comparison of the Macintosh laryngoscope, McGrath video laryngoscope, and Pentax Airway Scope in paediatric nasotracheal intubation
    Yoo, Ji Young
    Chae, Yun Jeong
    Lee, Young Bok
    Kim, Sujin
    Lee, Jaemoon
    Kim, Dae Hee
    [J]. SCIENTIFIC REPORTS, 2018, 8
  • [10] Comparison of tracheal intubation with the Airway Scope and Macintosh laryngoscope in the lateral position
    Xue, Fu-Shan
    Liao, Xu
    Wang, Qiang
    Yuan, Yu-Jing
    Xiong, Jun
    Liu, Jian-Hua
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2011, 28 (06) : 464 - 465