Construction of an automated gas chromatography/mass spectrometry system for the analysis of ambient volatile organic compounds with on-line internal standard calibration

被引:34
|
作者
Su, Yuan-Chang [1 ]
Chang, Chih-Chung [2 ]
Wang, Jia-Lin [1 ]
机构
[1] Natl Cent Univ, Dept Chem, Chungli 320, Taiwan
[2] Acad Sinica, Res Ctr Environm Changes, Taipei 11529, Taiwan
关键词
volatile organic compounds (VOCs); gas chromatography/mass spectrometry (GC/MS); internal standard (IS);
D O I
10.1016/j.chroma.2008.03.071
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An automated sampling and enrichment apparatus coupled with a gas chromatography/mass spectrometry (GC/MS) technique was constructed for the analysis of ambient volatile organic compounds (VOCs). A sorbent trap was built within the system to perform on-line enrichment and thermal desorption of VOCs onto GC/MS. In order to improve analytical precision, calibration accuracy, and to safe-guard the long-term stability of this system, a mechanism to allow on-line internal standard (I.S.) addition to the air sample stream was configured within the sampling and enrichment apparatus. A sub-ppm (v/v) level standard gas mixture containing 1,4-fluorobenzene, chloropentafluorobenzene, 1-bromo-4-fluorobenzene was prepared from their pure forms. A minute amount of this I.S. gas was volumetrically mixed into the sample stream at the time of on-line enrichment of the air sample to compensate for measurement uncertainties. To assess the performance of this VOC GC/MS system, a gas mixture containing numerous VOCs at sub-ppb (v/v) level served as the ambient air sample. Various internal standard methods based on total ion count (TIC) and selective ion monitoring (SIM) modes were attempted to assess the improvement in analytical precision and accuracy. Precision was improved from 7-8% RSD without I.S. to 2-3% with I.S. for the 14 target VOCs. Uncertainties in the calibration curves were also improved with the adoption of I.S. by reducing the relative standard deviation of the slope (Sm%) by an average a factor of 4, and intercept (Sb%) by a factor of 2 for the 14 target VOCs. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:134 / 140
页数:7
相关论文
共 50 条
  • [41] Volatile organic compounds emitted after leaf wounding: On-line analysis by proton-transfer-reaction mass spectrometry
    Fall, R
    Karl, T
    Hansel, A
    Jordan, A
    Lindinger, W
    JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 1999, 104 (D13) : 15963 - 15974
  • [42] APPLICATION OF GAS CHROMATOGRAPHY AND MASS-SPECTROMETRY FOR ANALYSIS OF WINE VOLATILE AND SEMI-VOLATILE COMPOUNDS
    Sagandykova, Gulyaim
    Sergazina, Meruyert
    Alimzhanova, Mereke
    Nauryzbaev, Mikhail
    ECOLOGY, ECONOMICS, EDUCATION AND LEGISLATION, VOL II, 2015, : 51 - 58
  • [43] Development and validation of a portable gas phase standard generation and calibration system for volatile organic compounds
    Veres, P.
    Gilman, J. B.
    Roberts, J. M.
    Kuster, W. C.
    Warneke, C.
    Burling, I. R.
    de Gouw, J.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2010, 3 (03) : 683 - 691
  • [44] Techniques of On-Line Monitoring Volatile Organic Compounds in Ambient Air with Optical Spectroscopy
    Du Zhen-hui
    Zhai Ya-qiong
    Li Jin-yi
    Hu Bo
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29 (12) : 3199 - 3203
  • [45] Analysis of volatile organic compounds in the ambient air of Algiers by gas chromatography with a β-cyclodextrin capillary column
    Yassaa, N
    Meklati, BY
    Cecinato, A
    JOURNAL OF CHROMATOGRAPHY A, 1999, 846 (1-2) : 287 - 293
  • [46] ANALYSIS OF ORGANIC COMPOUNDS IN HUMAN BREATH BY GAS CHROMATOGRAPHY-MASS SPECTROMETRY
    JANSSON, BO
    LARSSON, BT
    JOURNAL OF LABORATORY AND CLINICAL MEDICINE, 1969, 74 (06): : 961 - &
  • [47] Analysis of Volatile Organic Compounds in Human Saliva by a Static Sorptive Extraction Method and Gas Chromatography-Mass Spectrometry
    Helena A. Soini
    Iveta Klouckova
    Donald Wiesler
    Elisabeth Oberzaucher
    Karl Grammer
    Sarah J. Dixon
    Yun Xu
    Richard G. Brereton
    Dustin J. Penn
    Milos V. Novotny
    Journal of Chemical Ecology, 2010, 36 : 1035 - 1042
  • [48] Analysis of volatile organic compounds using gas chromatography
    Dewulf, J
    Van Langenhove, H
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2002, 21 (9-10) : 637 - 646
  • [49] Use of a presolvent to include volatile organic analytes in the application range of on-line solid-phase extraction gas chromatography mass spectrometry
    Hankemeier, T
    van Leeuwen, SPJ
    Vreuls, RJJ
    Brinkman, UAT
    JOURNAL OF CHROMATOGRAPHY A, 1998, 811 (1-2) : 117 - 133
  • [50] ANALYSIS OF TRACE VOLATILE ORGANIC-COMPOUNDS IN COFFEE BY HEADSPACE CONCENTRATION AND GAS-CHROMATOGRAPHY MASS-SPECTROMETRY
    WANG, TH
    SHANFIELD, H
    ZLATKIS, A
    CHROMATOGRAPHIA, 1983, 17 (08) : 411 - 417