Construction of an automated gas chromatography/mass spectrometry system for the analysis of ambient volatile organic compounds with on-line internal standard calibration

被引:34
|
作者
Su, Yuan-Chang [1 ]
Chang, Chih-Chung [2 ]
Wang, Jia-Lin [1 ]
机构
[1] Natl Cent Univ, Dept Chem, Chungli 320, Taiwan
[2] Acad Sinica, Res Ctr Environm Changes, Taipei 11529, Taiwan
关键词
volatile organic compounds (VOCs); gas chromatography/mass spectrometry (GC/MS); internal standard (IS);
D O I
10.1016/j.chroma.2008.03.071
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An automated sampling and enrichment apparatus coupled with a gas chromatography/mass spectrometry (GC/MS) technique was constructed for the analysis of ambient volatile organic compounds (VOCs). A sorbent trap was built within the system to perform on-line enrichment and thermal desorption of VOCs onto GC/MS. In order to improve analytical precision, calibration accuracy, and to safe-guard the long-term stability of this system, a mechanism to allow on-line internal standard (I.S.) addition to the air sample stream was configured within the sampling and enrichment apparatus. A sub-ppm (v/v) level standard gas mixture containing 1,4-fluorobenzene, chloropentafluorobenzene, 1-bromo-4-fluorobenzene was prepared from their pure forms. A minute amount of this I.S. gas was volumetrically mixed into the sample stream at the time of on-line enrichment of the air sample to compensate for measurement uncertainties. To assess the performance of this VOC GC/MS system, a gas mixture containing numerous VOCs at sub-ppb (v/v) level served as the ambient air sample. Various internal standard methods based on total ion count (TIC) and selective ion monitoring (SIM) modes were attempted to assess the improvement in analytical precision and accuracy. Precision was improved from 7-8% RSD without I.S. to 2-3% with I.S. for the 14 target VOCs. Uncertainties in the calibration curves were also improved with the adoption of I.S. by reducing the relative standard deviation of the slope (Sm%) by an average a factor of 4, and intercept (Sb%) by a factor of 2 for the 14 target VOCs. (c) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:134 / 140
页数:7
相关论文
共 50 条
  • [1] In-line sampling with gas chromatography-mass spectrometry to monitor ambient volatile organic compounds
    Wang, Jia-Lin
    Chang, Chih-Chung
    Lee, Kun-Zhang
    JOURNAL OF CHROMATOGRAPHY A, 2012, 1248 : 161 - 168
  • [2] An improved, automated whole air sampler and gas chromatography mass spectrometry analysis system for volatile organic compounds in the atmosphere
    Lerner, Brian M.
    Gilman, Jessica B.
    Aikin, Kenneth C.
    Atlas, Elliot L.
    Goldan, Paul D.
    Graus, Martin
    Hendershot, Roger
    Isaacman-VanWertz, Gabriel A.
    Koss, Abigail
    Kuster, William C.
    Lueb, Richard A.
    McLaughlin, Richard J.
    Peischl, Jeff
    Sueper, Donna
    Ryerson, Thomas B.
    Tokarek, Travis W.
    Warneke, Carsten
    Yuan, Bin
    de Gouw, Joost A.
    ATMOSPHERIC MEASUREMENT TECHNIQUES, 2017, 10 (01) : 291 - 313
  • [3] Analysis of the trace volatile organic compounds in landfill gas using automated thermal desorption gas chromatography mass spectrometry
    Allen, MR
    Braithwaite, A
    Hills, CC
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1996, 62 (01) : 43 - 52
  • [4] An automated purge and trap gas chromatography-mass spectrometry system for the sensitive shipboard analysis of volatile organic compounds in seawater
    Hashimoto, S
    Tanaka, T
    Yamashita, N
    Maeda, T
    JOURNAL OF SEPARATION SCIENCE, 2001, 24 (02): : 97 - 103
  • [5] Full-range analysis of ambient volatile organic compounds by a new trapping method and gas chromatography/mass spectrometry
    Su, Yuan-Chang
    Liu, Wen-Tzu
    Liao, Wei-Cheng
    Chiang, Sen-Wei
    Wang, Jia-Lin
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (34) : 5733 - 5742
  • [6] Cryogen free automated gas chromatography for the measurement of ambient volatile organic compounds
    Wang, JL
    Chen, WL
    Lin, YH
    Tsai, CH
    JOURNAL OF CHROMATOGRAPHY A, 2000, 896 (1-2) : 31 - 39
  • [7] Multicomponent analysis of volatile organic compounds in water by automated purge and trap coupled to gas chromatography-mass spectrometry
    Martínez, E
    Lacorte, S
    Llobet, I
    Viana, P
    Barceló, D
    JOURNAL OF CHROMATOGRAPHY A, 2002, 959 (1-2) : 181 - 190
  • [8] Automatic on-line monitoring of atmospheric volatile organic compounds: Gas chromatography-mass spectrometry and gas chromatography-flame ionization detection as complementary systems
    de Blas, Maite
    Navazo, Marino
    Alonso, Lucio
    Durana, Nieves
    Iza, Jon
    SCIENCE OF THE TOTAL ENVIRONMENT, 2011, 409 (24) : 5459 - 5469
  • [9] Determination of Oxygenated Volatile Organic Compounds in Ambient Air Using Canister Collection–Gas Chromatography/Mass Spectrometry
    Jun-ya Hoshi
    Masato Higuchi
    Yuko Sasaki
    Takashi Korenaga
    Analytical Sciences, 2007, 23 : 987 - 992
  • [10] Proton Transfer Reaction-Mass Spectrometry for On-line Analysis of Trace Volatile Organic Compounds
    Zhan Xue-Fang
    Duan Yi-Xiang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 39 (10) : 1611 - 1618