The performance of a fluoroscopic electronic portal imaging device modified for portability

被引:3
|
作者
Kirby, MC [1 ]
Atherton, S [1 ]
Carson, P [1 ]
McDonagh, C [1 ]
Sykes, JR [1 ]
Williams, PC [1 ]
机构
[1] Christie Hosp NHS Trust, Manchester M20 4BX, Lancs, England
来源
BRITISH JOURNAL OF RADIOLOGY | 1999年 / 72卷 / 862期
关键词
D O I
10.1259/bjr.72.862.10673953
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Advances in external beam therapy technology have made routine, efficient conformal therapy a reality. With it comes the increasing need for online treatment verification, which is only achievable at present through the use of electronic portal imaging devices (EPIDs). For a large radiotherapy centre, the provision of one EPID per treatment machine proves extremely expensive. This paper details modifications to the design of a commercial fluoroscopic EPID (the SRI-100) to produce a portable system, capable of providing quick, high quality imaging on more than one treatment machine. We describe the necessary hardware and software changes made to the system, as well as the variety of mechanical and quality control checks performed for testing the stability and quality of the imaging. The modified system has been found to be both electronically and mechanically robust, with associated image quality, scaling, distortion and movement similar to other EPIDs in the department. Although the modification was designed specifically to allow for the acquisition of images from multiple treatment machines, it may also enable the operation of the EPID for other uses such as total body irradiation (TBI) treatment verification and a further range of quality control procedures on the linear accelerator itself.
引用
收藏
页码:1000 / 1005
页数:6
相关论文
共 50 条
  • [1] Dosimetry with a fluoroscopic electronic portal imaging device
    Pasma, KL
    MEDICAL PHYSICS, 2000, 27 (02) : 423 - 423
  • [2] Dosimetric properties of the Theraview™ fluoroscopic electronic portal imaging device
    Glendinning, AG
    Bonnett, DE
    BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (869): : 517 - 530
  • [3] Characterization of a high-elbow, fluoroscopic electronic portal imaging device for portal dosimetry
    de Boer, JCJ
    Heijmen, BJM
    Pasma, KL
    Visser, AG
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (01): : 197 - 216
  • [4] Verification of compensator thicknesses using a fluoroscopic electronic portal imaging device
    Department of Clinical Physics, Daniel den Hoed Cancer Center, University Hospital Rotterdam, Groene Hilledijk 301, 3075 EA Rotterdam, Netherlands
    Med. Phys., 8 (1524-1529):
  • [5] Verification of compensator thicknesses using a fluoroscopic electronic portal imaging device
    Pasma, KL
    Kroonwijk, M
    van Dieren, EB
    Visser, AG
    Heijmen, BJM
    MEDICAL PHYSICS, 1999, 26 (08) : 1524 - 1529
  • [6] Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation
    Pasma, KL
    Kroonwijk, M
    de Boer, JCJ
    Visser, AG
    Heijmen, BJM
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08): : 2047 - 2060
  • [7] Improved match performance by programming electronic portal imaging device positions
    Richards, EM
    Damhuis, HGM
    Steggink, JEM
    Ong, F
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S465 - S466
  • [8] Detection of organ movement in cervix cancer patients using a fluoroscopic electronic portal imaging device and radiopaque markers
    Kaatee, RSJP
    Olofsen, MJJ
    Verstraate, MBJ
    Quint, S
    Heijmen, BJM
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (02): : 576 - 583
  • [9] An electronic portal imaging device for transit dosimetry
    SymondsTayler, JRN
    Partridge, M
    Evans, PM
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11): : 2273 - 2283
  • [10] Preclinical Application of Electronic Portal Imaging Device
    Anvari, Akbar
    Sawant, Amit
    MEDICAL PHYSICS, 2020, 47 (06) : E830 - E830