Impact of respiratory motion artifact on coronary image quality of one beat coronary CT angiography

被引:3
|
作者
Shen, Wenting [1 ,2 ]
Chen, Yang [1 ]
Qian, Wen [1 ]
Liu, Wangyan [1 ]
Zhu, Yinsu [1 ]
Xu, Yi [1 ]
Zhu, Xiaomei [1 ]
机构
[1] Nanjing Med Univ, Dept Radiol, Affiliated Hosp 1, Guangzhou Rd, Nanjing, Jiangsu, Peoples R China
[2] Liyang Peoples Hosp, Dept Radiol, Jianshe West Rd, Liyang, Jiangsu, Peoples R China
关键词
Motion artifacts; computed tomography angiography; coronary angiography; breath holding; FRACTIONAL FLOW RESERVE; CORRECTION ALGORITHM; ITERATIVE RECONSTRUCTION; PULMONARY ANGIOGRAPHY; COMPUTED-TOMOGRAPHY; DIAGNOSTIC-ACCURACY;
D O I
10.3233/XST-200812
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
BACKGROUND: Accuracy of CT-derived fractional flow reserve depends on good image quality. Thus, improving image quality during coronary CT angiography (CCTA) is important. OBJECTIVE: To investigate impact of respiratory motion artifact on coronary image quality focusing on vessel diameter and territory during one beat CCTA by a 256-row detector. METHODS: We retrospectively reviewed patients who underwent CCTA under free-breathing (n = 100) and breath-holding (n = 100), respectively. Coronary image quality is defined as 4-1 from excellent to poor (non-diagnostic) and respiratory motion artifact severity is also scored on a 4-point scale from no artifact to severe artifact. Coronary image quality and respiratory motion artifact severity of all images were evaluated by two radiologists independently. RESULTS: Compared with free-breathing group, the image qualities are significantly higher in per-segment, per-vessel and per-patient levels (P < 0.001) and proportion of segments with excellent image quality also improves significantly (73.6% vs 60.1%, P < 0.001) in breath-holding group. The image quality improvement occurs in medium-sized coronary arterial segments. Coronary image quality improves with respiratory motion artifacts decreasing in both groups, respectively. CONCLUSION: During one heartbeat CCTA, breath-holding is still recommended to improve coronary image quality due to improvement of the image quality in the medium-sized coronary arteries.
引用
收藏
页码:287 / 296
页数:10
相关论文
共 50 条
  • [31] Coronary MR angiography: Respiratory motion correction with BACSPIN
    Hardy, CJ
    Zhao, L
    Zong, XL
    Saranathan, M
    Yucel, EK
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (02) : 170 - 176
  • [32] Impact of coronary CT image quality on the accuracy of the FFRCT Planner
    Andreini, Daniele
    Belmonte, Marta
    Penicka, Martin
    Van Hoe, Lieven
    Mileva, Niya
    Paolisso, Pasquale
    Nagumo, Sakura
    Norgaard, Bjarne L.
    Ko, Brian
    Otake, Hiromasa
    Koo, Bon-Kwon
    Jensen, Jesper Moller
    Mizukami, Takuya
    Munhoz, Daniel
    Updegrove, Adam
    Taylor, Charles
    Leipsic, Jonathon
    Sonck, Jeroen
    De Bruyne, Bernard
    Collet, Carlos
    [J]. EUROPEAN RADIOLOGY, 2024, 34 (04) : 2677 - 2688
  • [33] Impact of coronary CT image quality on the accuracy of the FFRCT Planner
    Daniele Andreini
    Marta Belmonte
    Martin Penicka
    Lieven Van Hoe
    Niya Mileva
    Pasquale Paolisso
    Sakura Nagumo
    Bjarne L. Nørgaard
    Brian Ko
    Hiromasa Otake
    Bon-Kwon Koo
    Jesper Møller Jensen
    Takuya Mizukami
    Daniel Munhoz
    Adam Updegrove
    Charles Taylor
    Jonathon Leipsic
    Jeroen Sonck
    Bernard De Bruyne
    Carlos Collet
    [J]. European Radiology, 2024, 34 : 2677 - 2688
  • [34] Adaptive Statistical Iterative Reconstruction: Assessment of Image Noise and Image Quality in Coronary CT Angiography
    Leipsic, Jonathon
    LaBounty, Troy M.
    Heilbron, Brett
    Min, James K.
    Mancini, G. B. John
    Lin, Fay Y.
    Taylor, Carolyn
    Dunning, Allison
    Earls, James P.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (03) : 649 - 654
  • [35] Influence of virtual monochromatic spectral image at different energy levels on motion artifact correction in dual-energy spectral coronary CT angiography
    Jia, Yongjun
    Zhai, Bingying
    He, Taiping
    Yu, Yong
    Yu, Nan
    Duan, Haifeng
    Yang, Chuangbo
    Zhang, Xirong
    [J]. JAPANESE JOURNAL OF RADIOLOGY, 2019, 37 (09) : 636 - 641
  • [36] Influence of virtual monochromatic spectral image at different energy levels on motion artifact correction in dual-energy spectral coronary CT angiography
    Yongjun Jia
    Bingying Zhai
    Taiping He
    Yong Yu
    Nan Yu
    Haifeng Duan
    Chuangbo Yang
    Xirong Zhang
    [J]. Japanese Journal of Radiology, 2019, 37 : 636 - 641
  • [37] Impact of left atrial enlargement on image quality in coronary computed tomography angiography
    Nguyen, Tien C.
    Ohyama, Yoshiaki
    Taketomi-Takahashi, Ayako
    Nguyen, Huong T.
    Sumiyoshi, Hisako
    Nakamura, Tetsuya
    Kurabayashi, Masahiko
    Tsushima, Yoshito
    [J]. CLINICAL IMAGING, 2022, 84 : 87 - 92
  • [38] CT coronary angiography: indications, image acquisition, and interpretation
    Schoepf, U. J.
    Thilo, C.
    Fernandez, M. J.
    Costello, P.
    [J]. RADIOLOGIA, 2008, 50 (02): : 113 - 130
  • [39] Impact of a New Motion-Correction Algorithm on Image Quality of Low-Dose Coronary CT Angiography in Patients with Insufficient Heart Rate Control
    Fuchs, Tobias A.
    Stehli, Julia
    Dougoud, Svetlana
    Fiechter, Michael
    Sah, Bert-Ram
    Buechel, Ronny R.
    Bull, Sacha
    Gaemperli, Oliver
    Kaufmann, Philipp A.
    [J]. ACADEMIC RADIOLOGY, 2014, 21 (03) : 312 - 317
  • [40] Assessment of image quality and radiation dose in prospective ECG-triggered coronary CT angiography compared with retrospective ECG-gated coronary CT angiography
    Ko, Sung Min
    Kim, Na Ra
    Kim, Dong Hun
    Song, Meong Gun
    Kim, Jin Hwan
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2010, 26 : 93 - 101