Impact of respiratory motion artifact on coronary image quality of one beat coronary CT angiography

被引:3
|
作者
Shen, Wenting [1 ,2 ]
Chen, Yang [1 ]
Qian, Wen [1 ]
Liu, Wangyan [1 ]
Zhu, Yinsu [1 ]
Xu, Yi [1 ]
Zhu, Xiaomei [1 ]
机构
[1] Nanjing Med Univ, Dept Radiol, Affiliated Hosp 1, Guangzhou Rd, Nanjing, Jiangsu, Peoples R China
[2] Liyang Peoples Hosp, Dept Radiol, Jianshe West Rd, Liyang, Jiangsu, Peoples R China
关键词
Motion artifacts; computed tomography angiography; coronary angiography; breath holding; FRACTIONAL FLOW RESERVE; CORRECTION ALGORITHM; ITERATIVE RECONSTRUCTION; PULMONARY ANGIOGRAPHY; COMPUTED-TOMOGRAPHY; DIAGNOSTIC-ACCURACY;
D O I
10.3233/XST-200812
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
BACKGROUND: Accuracy of CT-derived fractional flow reserve depends on good image quality. Thus, improving image quality during coronary CT angiography (CCTA) is important. OBJECTIVE: To investigate impact of respiratory motion artifact on coronary image quality focusing on vessel diameter and territory during one beat CCTA by a 256-row detector. METHODS: We retrospectively reviewed patients who underwent CCTA under free-breathing (n = 100) and breath-holding (n = 100), respectively. Coronary image quality is defined as 4-1 from excellent to poor (non-diagnostic) and respiratory motion artifact severity is also scored on a 4-point scale from no artifact to severe artifact. Coronary image quality and respiratory motion artifact severity of all images were evaluated by two radiologists independently. RESULTS: Compared with free-breathing group, the image qualities are significantly higher in per-segment, per-vessel and per-patient levels (P < 0.001) and proportion of segments with excellent image quality also improves significantly (73.6% vs 60.1%, P < 0.001) in breath-holding group. The image quality improvement occurs in medium-sized coronary arterial segments. Coronary image quality improves with respiratory motion artifacts decreasing in both groups, respectively. CONCLUSION: During one heartbeat CCTA, breath-holding is still recommended to improve coronary image quality due to improvement of the image quality in the medium-sized coronary arteries.
引用
收藏
页码:287 / 296
页数:10
相关论文
共 50 条
  • [21] Coronary MR Angiography in patients with coronary artery disease using image-based respiratory motion compensation
    Markus Henningsson
    Konstantinos Bratis
    Eike Nagel
    Rene Botnar
    [J]. Journal of Cardiovascular Magnetic Resonance, 17 (Suppl 1)
  • [22] Impact of coronary calcium score as gatekeeper for coronary CT angiography
    Otsuka, M.
    Kokura, I.
    Nakashima, T.
    Gunji, T.
    Ozaki, Y.
    Takenaka, S.
    Nakamura, M.
    Tsurumi, Y.
    [J]. EUROPEAN HEART JOURNAL, 2013, 34 : 974 - 974
  • [23] Impact on image quality and radiation exposure in coronary CT angiography: 100 kVp versus 120 kVp
    Ripsweden, Jonaz
    Brismar, Torkel B.
    Holm, Jon
    Melinder, Annika
    Mir-Akbari, Habib
    Nilsson, Tage
    Nyman, Ulf
    Rasmussen, Elsbeth
    Ruck, Andreas
    Cederlund, Kerstin
    [J]. ACTA RADIOLOGICA, 2010, 51 (08) : 903 - 909
  • [24] Image assessment with multislice CT coronary angiography
    Cademartiri, F
    Luccichenti, G
    Runza, G
    Belgrano, M
    Brambilla, L
    Midiri, M
    Marano, R
    Mollet, N
    Galia, M
    Gualerzi, M
    Coruzzi, P
    [J]. RADIOLOGIA MEDICA, 2005, 109 (03): : 198 - 207
  • [25] One-beat whole heart coronary CT-angiography using the 256-multislice CT
    Kurata, Akira
    Mochizuki, Teruhito
    Mori, Shin-ichiro
    Kdo, Teruhito
    Tanada, Shuji
    Endo, Masahiro
    Katada, Kazuhiro
    Higashino, Hiroshi
    Saito, Makoto
    Ogimoto, Alkiyoshi
    Ohtsuka, Tomoaki
    Okayama, Hideki
    Higaki, Jitsuo
    [J]. CIRCULATION, 2006, 114 (18) : 449 - 449
  • [26] Image Quality of CT Angiography of Coronary Arteries Dependig on the Degree of Coronary Calcification Using a Dual Source CT Scanner
    Artmann, A.
    Enayati, S.
    Ratzenboeck, M.
    Schwarz, M.
    Weber, T.
    Eber, B.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (09): : 863 - 869
  • [27] 64 multidetector CT coronary angiography: Effect of sublingual nitroglycerin on coronary luminal diameter and image quality
    Vanhoenacker, Piet K.
    Decramer, Isabel
    Samo, Giovanna
    Van Hoe, Lieven R.
    Wijns, William
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 179A - 179A
  • [28] Impact of PVCs noted during coronary calcium scan on image quality and accuracy in subsequent coronary dual-source CT angiography
    Eicher, Wolfgang
    Kau, Thomas
    Niedermayer, Martin
    Senft, Birgit
    Sinzig, Maria
    Strozyk, Birgit
    Hausegger, Klaus Armin
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (04): : 601 - 610
  • [29] Impact of PVCs noted during coronary calcium scan on image quality and accuracy in subsequent coronary dual-source CT angiography
    Wolfgang Eicher
    Thomas Kau
    Martin Niedermayer
    Birgit Senft
    Maria Sinzig
    Birgit Strozyk
    Klaus Armin Hausegger
    [J]. The International Journal of Cardiovascular Imaging, 2011, 27 : 601 - 610
  • [30] Advanced Respiratory Motion Compensation for Coronary MR Angiography
    Henningsson, Markus
    Botnar, Rene M.
    [J]. SENSORS, 2013, 13 (06) : 6882 - 6899