Chirp-Encoded Excitation for Dual-Frequency Ultrasound Tissue Harmonic Imaging

被引:31
|
作者
Shen, Che-Chou [1 ]
Lin, Chin-Hsiang [1 ]
机构
[1] Natl Taiwan Univ Sci & Technol, Dept Elect Engn, Taipei, Taiwan
关键词
CODED EXCITATION; MEDICAL ULTRASOUND; DISTORTION; SIGNALS; SYSTEM;
D O I
10.1109/TUFFC.2012.2474
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Dual-frequency (DF) transmit waveforms comprise signals at two different frequencies. With a DF transmit waveform operating at both fundamental frequency (f(0)) and second-harmonic frequency (2f(0)), tissue harmonic imaging can be simultaneously performed using not only the conventional 2f(0) second-harmonic signal but also using the f(0) frequency-difference harmonic signal. Nonetheless, when chirp excitation is incorporated into the DF transmit waveform for harmonic SNR improvement, a particular waveform design is required to maintain the bandwidth of the f(0) harmonic signal. In this study, two different DF chirp waveforms are proposed to produce equal harmonic bandwidth at both the f(0) and 2f(0) frequencies to achieve speckle reduction by harmonic spectral compounding and to increase harmonic SNR for enhanced penetration and sensitivity. The UU13 waveform comprises an up-sweeping f(0) chirp and an up-sweeping 2f(0) chirp with triple bandwidth, whereas the UD11 waveform includes an up-sweeping f(0) chirp and a down-sweeping 2f(0) chirp with equal bandwidth. Experimental results indicate that the UU13 tends to suffer from a high range side lobe level resulting from 3f(0) interference. Consequently, the 2f(0) harmonic envelopes of the UD11 and the UU13 waveforms have compression qualities of 87% and 77%, respectively, when the signal bandwidth is 30%. When the bandwidth increases to 50%, the compression quality of the 2f(0) harmonic envelope degrades to 78% and 54%, respectively, for the UD11 and the UU13 waveforms. The compression quality value of the f(0) harmonic envelope remains similar between the two DF transmit waveforms for all signal bandwidths. B-mode harmonic images also show that the UD11 is less contaminated by range side lobe artifacts than is the UU13. Compared with a short pulse with equal bandwidth, the UD11 waveform not only preserves the same spatial resolution after compression but also improves the image SNR by about 10 dB. Moreover, the image contrast-to-noise ratio (CNR), defined as the ratio of the mean to the standard deviation of image intensity in the speckle region, can be increased from 1.0 to about 1.2 when DF spectral compounding is performed. Therefore, it is concluded that the UD11 waveform is a potential solution for chirp-encoded DF harmonic imaging.
引用
收藏
页码:2420 / 2430
页数:11
相关论文
共 50 条
  • [41] Ultrasound second harmonic imaging with a weighted chirp signal
    Kim, DY
    Lee, JC
    Kwon, SJ
    Song, TK
    [J]. 2001 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2001, : 1477 - 1480
  • [42] An electrical impedance matching method of dual-frequency transducers for ultrasound internal imaging
    Li, Jiaqi
    Huang, Wenchang
    Shao, Weiwei
    Li, Xiangxin
    Shen, Zhitian
    Xu, Jie
    Shen, Jun
    Cui, Yaoyao
    [J]. MEASUREMENT, 2023, 221
  • [43] Dual-Element Intravascular Ultrasound Transducer for Tissue Harmonic Imaging and Frequency Compounding: Development and Imaging Performance Assessment
    Lee, Junsu
    Chang, Jin Ho
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (11) : 3146 - 3155
  • [44] A dual-frequency excitation technique for enhancing the sub-harmonic emission from encapsulated microbubbles
    Zhang, Dong
    Xi, Xiaoyu
    Zhang, Zhe
    Gong, Xiufen
    Chen, Gong
    Wu, Junru
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (13): : 4257 - 4272
  • [45] Parameter optimization and experiment of chirp excitation ultrasound imaging
    Gu Jinhong
    Wang Supin
    [J]. ISTM/2007: 7TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1-7, CONFERENCE PROCEEDINGS, 2007, : 1286 - 1289
  • [46] Intensification of leaching process by dual-frequency ultrasound
    Swamy, KM
    Narayana, KL
    [J]. ULTRASONICS SONOCHEMISTRY, 2001, 8 (04) : 341 - 346
  • [47] Dual-Frequency Ultrasound Imaging and Therapeutic Bilaminar Array Using Frequency Selective Isolation Layer
    Azuma, Takashi
    Ogihara, Makoto
    Kubota, Jun
    Sasaki, Akira
    Umemura, Shin-ichiro
    Furuhata, Hiroshi
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2010, 57 (05) : 1211 - 1224
  • [48] Dual-frequency ultrasound for detecting and sizing bubbles
    Buckey, JC
    Knaus, DA
    Alvarenga, DL
    Kenton, MA
    Magari, PJ
    [J]. ACTA ASTRONAUTICA, 2005, 56 (9-12) : 1041 - 1047
  • [49] Color-coded dual-frequency intravascular ultrasound imaging with use of radio-frequency ultrasound signals: A novel modality of tissue characterization of coronary plaque
    Hashimoto, Genta
    Hiro, Takafumi
    Fujii, Takashi
    Okamura, Takayuki
    Yamada, Jutaro
    Fukumoto, Yusaku
    Fujimura, Tatsuhiro
    Kanou, Hiroko
    Matsuzaki, Masunori
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 49 (09) : 171A - 171A
  • [50] A high frequency endoscopic ultrasound imaging method combining chirp coded excitation and compressed sensing
    Wang, Ninghao
    Li, Xinze
    Xu, Jie
    Jiao, Yang
    Cui, Yaoyao
    Jian, Xiaohua
    [J]. ULTRASONICS, 2022, 121