Potential reduction of lung dose via VMAT with jaw tracking in the treatment of single-isocenter/two-lesion lung SBRT

被引:23
|
作者
Pokhrel, Damodar [1 ]
Sanford, Lana [1 ]
Halfman, Matthew [1 ]
Molloy, Janelle [1 ]
机构
[1] Univ Kentucky, Dept Radiat Med, Med Phys Grad Program, Lexington, KY 40506 USA
来源
关键词
lung SBRT; VMAT; jaw tracking; single-isocenter; two-lesion; BODY RADIATION-THERAPY; PNEUMONITIS; RADIOSURGERY; CARCINOMA; TRIAL; RISK;
D O I
10.1002/acm2.12580
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose/objectives Due to higher radiosensitivity, non-target normal tissue dose is a major concern in stereotactic body radiation therapy (SBRT) treatment. The aim of this report was to estimate the dosimetric impact, specifically the reduction of normal lung dose in the treatment of single-isocenter/two-lesion lung SBRT via volumetric modulated arc therapy with jaw tracking (JT-VMAT). Materials/methods Twelve patients with two peripherally located early-stage non-small-cell-lung cancer (NSCLC) lung lesions underwent single-isocenter highly conformal non-coplanar JT-VMAT SBRT treatment in our institution. The mean isocenter to tumors distance was 5.6 +/- 1.9 (range 4.3-9.5) cm. The mean combined planning target volume (PTV) was 38.7 +/- 22.7 (range 5.0-80.9) cc. A single isocenter was placed between the two lesions. Doses were 54 and 50 Gy in three and five fractions, respectively. Plans were optimized in Eclipse with AcurosXB algorithm utilizing jaw tracking options for the Truebeam with a 6 MV-FFF beam and standard 120 leaf millennium multi-leaf collimators. For comparison, the JT-VMAT plans were retrospectively re-computed utilizing identical beam geometry, objectives, and planning parameters, but without jaw tracking (no JT-VMAT). Both plans were normalized to receive the same target coverage. The conformity and heterogeneity indices, intermediate-dose spillage [D-2cm, R50, Gradient Index (GI), Gradient Distance (GD)], organs at risks (OAR) doses including normal lung as well as modulation factor (MF) were compared for both plans. Results For similar target coverage, GI, R50, GD, as well as the normal lung V5, V10, V20, mean lung dose (MLD), and maximum dose received by 1000 cc of lungs were statistically significant. Normal lung doses were reduced by 8%-11% with JT-VMAT. Normal lung dose increased as a function of tumor distance from isocenter. For the other OAR, up to 1%-16% reduction of non-target doses were observed with JT-VMAT. The MF and beam-on time were similar for both plans, however, MF increased as a function of tumors distance, consequently, delivering higher dose to normal lungs. Conclusion Utilizing jaw tracking options during optimization for single-isocenter/two-lesion lung SBRT VMAT plans reduced doses to the normal lung and other OAR, reduced intermediate-dose spillage and provided superior/similar target coverage. Application of jaw tracking did not affect delivery efficiency and provided excellent plan quality with similar MF and beam-on time. Jaw tracking is recommended for future clinical SBRT plan optimization.
引用
收藏
页码:55 / 63
页数:9
相关论文
共 50 条
  • [1] Evaluating the Plan Quality and Treatment Efficiency for Single-Isocenter/Two-Lesion Lung SBRT
    Sanford, L.
    Molloy, J.
    McGarry, R.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E529 - E530
  • [2] Feasibility of Using 6MV-FFF Halcyon RDS for Single-Isocenter/Two-Lesion Lung SBRT
    Webster, A.
    Visak, J.
    Bernard, M.
    McGarry, R.
    Kudrimoti, M.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [3] How much rotational error is clinically acceptable for single-isocenter/two-lesion lung SBRT treatment on halcyon ring delivery system (RDS)?
    Pokhrel, Damodar
    Mallory, Richard
    Bernard, Mark E.
    Kudrimoti, Mahesh
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (07):
  • [4] Evaluation of plan quality and treatment efficiency for single-isocenter/two-lesion lung stereotactic body radiation therapy
    Sanford, Lana
    Molloy, Janette
    Kumar, Sameera
    Randall, Marcus
    McGarry, Ronald
    Pokhrel, Damodar
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (01): : 118 - 127
  • [5] Dosimetric Comparison of Single-Isocenter and MultipleIsocenter Techniques for Two-Lesion Lung SBRT Using the RefleXion High-Speed Ring-Gantry System
    Oderinde, O. M.
    Voronenko, Y.
    Tian, S.
    Yang, X.
    Higgins, K. A.
    Da Silva, A.
    Shirvani, S. M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E139 - E140
  • [6] Feasibility of using ring-mounted Halcyon Linac for single-isocenter/two-lesion lung stereotactic body radiation therapy
    Pokhrel, Damodar
    Webster, Aaron
    Mallory, Richard
    Visak, Justin
    Bernard, Mark E.
    McGarry, Ronald C.
    Kudrimoti, Mahesh
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):
  • [7] Margin calculation for multiple metastases in single-isocenter lung SBRT
    van Timmeren, J.
    Hoogeman, M.
    Ehrbar, S.
    Mayinger, M.
    Andratschke, N.
    Guckenberger, M.
    Tanadini-Lang, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S62 - S63
  • [8] Improving treatment efficiency via photon optimizer (PO) MLC algorithm for synchronous single-isocenter/multiple-lesions VMAT lung SBRT
    Sanford, Lana
    Pokhrel, Damodar
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10): : 201 - 207
  • [9] Improving Treatment Efficiency Using Photon Optimizer MLC Algorithm for Single-Isocenter/Multiple-Lesions VMAT Lung SBRT
    Sanford, L.
    Molloy, J.
    Bernard, M.
    Randall, M.
    McGarry, R.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E265 - E265
  • [10] Selection Strategy of Jaw Tracking in VMAT Planning for Lung SBRT
    Sun, Wuji
    Shi, Yinghua
    Li, Yu
    Ge, Chao
    Yang, Xu
    Xia, Wenming
    Chen, Kunzhi
    Wang, Libo
    Dong, Lihua
    Wang, Huidong
    [J]. FRONTIERS IN ONCOLOGY, 2022, 12