Improving treatment efficiency via photon optimizer (PO) MLC algorithm for synchronous single-isocenter/multiple-lesions VMAT lung SBRT

被引:12
|
作者
Sanford, Lana [1 ,2 ]
Pokhrel, Damodar [1 ,2 ]
机构
[1] Univ Kentucky, Med Phys Grad Program, Lexington, KY 40506 USA
[2] Univ Kentucky, Dept Radiat Med, Lexington, KY 40506 USA
来源
关键词
lung SBRT; MLC; photon optimizer; single-isocenter; multi-lesions; VMAT; BODY RADIATION-THERAPY; MODULATED ARC THERAPY; PLAN QUALITY;
D O I
10.1002/acm2.12721
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Elderly patients with multiple primary or oligometastases (<5 lesions) lesions with associated co-morbidities may not retain their treatment position for the traditional long SBRT treatment time with individual isocenters for each lesion. Treating multiple lesions synchronously using a single-isocenter volumetric arc therapy (VMAT) plan would be more efficient with the use of the most recently adopted photon optimizer (PO) MLC algorithm and improve the patient comfort. Herein, we quantified the clinical performance of PO versus its predecessor progressive resolution optimizer (PRO) algorithm for single-isocenter/multiple-lesions VMAT lung SBRT. Materials and methods Fourteen patients with metastatic non-small-cell lung cancer lesions (two to five, both uni- and bilateral lungs) received a highly conformal single-isocenter co/non-coplanar VMAT (2-6 arcs) SBRT treatment plan. Patients were treated with a 6X-FFF beam and Acuros algorithm with a single-isocenter placed between/among the lesions, using PO for MLC optimization. Average isocenter to tumor distance was 5.5 +/- 1.9 cm. Mean combined PTV derived from 4D-CT scans was 38.7 +/- 22.7 cc. Doses were 54 Gy/50 Gy in 3/5 fractions prescribed to 70%-80% isodose line so that at least 95% of the PTV receives 100% of prescribed dose. Plans were re-optimized using PRO algorithm. Plans were compared via ROTG-0915 protocol criteria for target conformity, heterogeneity and gradient indices, and dose to organs-at-risk (OAR). Additionally, total number of monitor units (MU), modulation factor (MF) and beam-on time were compared. Results All plans met SBRT protocol requirements for target coverage and OAR doses. Comparison of target coverage and dose to the OAR showed no statistical significance between the two plans. PO had 1042 +/- 753 (P < 0.001) less MU than PRO resulting in a beam-on time of about 0.75 +/- 0.5 min (P < 0.001) less, on average. For similar dose distribution, a significant reduction of beam delivery complexity was observed with PO (average MF = 3.7 +/- 0.7) vs PRO MLC algorithm (average MF = 4.4 +/- 1.3) (P < 0.001). Conclusions PO MLC algorithm improved treatment efficiency without compromising plan quality when compared to PRO algorithm for single-isocenter/multi-lesions VMAT lung SBRT. Shorter beam-on time can potentially reduce intrafraction motion errors and improve patient compliance. PO MLC algorithm is recommended for future clinical lung SBRT plan optimization.
引用
收藏
页码:201 / 207
页数:7
相关论文
共 19 条
  • [1] Improving Treatment Efficiency Using Photon Optimizer MLC Algorithm for Single-Isocenter/Multiple-Lesions VMAT Lung SBRT
    Sanford, L.
    Molloy, J.
    Bernard, M.
    Randall, M.
    McGarry, R.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E265 - E265
  • [2] Improving the Plan Quality and Treatment Efficiency Via Flattening Filter Free (FFF) Beams for Single-Isocenter/multiple-Lesions VMAT Lung SBRT
    Sanford, L.
    Molloy, J.
    Bernard, M.
    Randall, M.
    McGarry, R.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E659 - E660
  • [3] On the use of single-isocenter VMAT plans for SBRT treatment of synchronous multiple lung lesions: Plan quality, treatment efficiency, and early clinical outcomes
    Pokhrel, Damodar
    Sanford, Lana
    Larkin, Shilpa
    Dhanireddy, Bhaswanth
    Bernard, Mark E.
    Randall, Marcus
    McGarry, Ronald C.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (08): : 160 - 167
  • [4] Impact of MLC Uncertainty in Single Isocenter Multiple-Lesions SRS VMAT Plans
    Xu, J.
    Chacko, E.
    Zhang, B.
    MacFarlane, M.
    Jiang, K.
    Mishra, M.
    Guerrero, M.
    Chen, S.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [5] Clinical evaluation of photon optimizer (PO) MLC algorithm for stereotactic, single-dose of VMAT lung SBRT
    Visak, Justin
    McGarry, Ronald C.
    Pokhrel, Damodar
    [J]. MEDICAL DOSIMETRY, 2020, 45 (04) : 321 - 326
  • [6] Plan Evaluation of Lung SBRT with Multiple Lesions: Single-Isocenter Versus Multiple-Isocenters
    Chanyavanich, V.
    Carter, D.
    Bailey, J.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E330 - E331
  • [7] Potential reduction of lung dose via VMAT with jaw tracking in the treatment of single-isocenter/two-lesion lung SBRT
    Pokhrel, Damodar
    Sanford, Lana
    Halfman, Matthew
    Molloy, Janelle
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05): : 55 - 63
  • [8] Evaluating the Plan Quality and Treatment Efficiency for Single-Isocenter/Two-Lesion Lung SBRT
    Sanford, L.
    Molloy, J.
    McGarry, R.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E529 - E530
  • [9] VMAT for Single Isocenter Lung SBRT for Multiple Lesions: Rotation Error Effects on Plan Clinical Acceptability
    Alani, S.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E493 - E493
  • [10] A Novel Restricted Single-Isocenter Stereotactic Body Radiotherapy (RESIST) Treatment for Synchronous Multiple Lung Lesions That Minimizes Setup Uncertainties
    Critchfield, L.
    Bernard, M.
    Randall, M.
    McGarry, R.
    Pokhrel, D.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E388 - E388