Fast 3-D Velocity Estimation in 4-D Using a 62+62 Row-Column Addressed Array

被引:10
|
作者
Schou, Mikkel [1 ]
Jorgensen, Lasse Thurmann [1 ]
Beers, Christopher [2 ]
Traberg, Marie Sand [1 ]
Tomov, Borislav Gueorguiev [1 ]
Bo Stuart, Matthias [1 ]
Jensen, Jorgen Arendt [1 ]
机构
[1] Tech Univ Denmark, Dept Hlth Technol, Ctr Fast Ultrasound Imaging, DK-2800 Lyngby, Denmark
[2] BK Med, State Coll, PA 16803 USA
关键词
Estimation; Apertures; Blood; Focusing; Probes; Acoustics; Blood flow measurement; row column addressed; synthetic aperture flow imaging; tensor velocity imaging (TVI); three dimensional imaging; ultrasonic transducers;
D O I
10.1109/TUFFC.2020.3016991
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This article presents an imaging scheme capable of estimating the full 3-D velocity vector field in a volume using row-column addressed arrays (RCAs) at a high volume rate. A 62 + 62 RCA array is employed with an interleaved synthetic aperture sequence. It contains repeated emissions with rows and columns interleaved with B-mode emissions. The sequence contains 80 emissions in total and can provide continuous volumetric data at a volume rate above 125 Hz. A transverse oscillation cross correlation estimator determines all three velocity components. The approach is investigated using Field II simulations and measurements using a specially built 3-MHz 62 + 62 RCA array connected to the SARUS experimental scanner. Both the B-mode and flow sequences have a penetration depth of 14 cm when measured on a tissue-mimicking phantom (0.5-dB/[MHz . cm] attenuation). Simulations of a parabolic flow in a 12-mm-diameter vessel at a depth of 30 mm, beam-to-flow angle of 90 degrees, and xy-rotation of 45 degrees gave a standard deviation (SD) of (3.3, 3.4, 0.4)% and bias of (-3.3, -3.9, -0.1)%, for (v(x), v(y), and v(z)). Decreasing the beam-to-flow angle to 60 degrees gave an SD of (8.9, 9.1, 0.8)% and bias of (-7.6, -9.5, -7.2)%, showing a slight increase. Measurements were carried out using a similar setup, and pulsing at 2 kHz yielded comparable results at 90 degrees with an SD of (5.8, 5.5, 1.1)% and bias of (1.4, -6.4, 2.4)%. At 60 degrees, the SD was (5.2, 4.7 1.2)% and bias (-4.6, 6.9, -7.4)%. Results from measurements across all tested settings showed a maximum SD of 6.8% and a maximum bias of 15.8% for a peak velocity of 10 cm/s. A tissue-mimicking phantom with a straight vessel was used to introduce clutter, tissue motion, and pulsating flow. The pulsating velocity magnitude was estimated across ten pulse periods and yielded an SD of 10.9%. The method was capable of estimating transverse flow components precisely but underestimated the flow with small beam-to-flow angles. The sequence provided continuous data in both time and space throughout the volume, allowing for retrospective analysis of the flow. Moreover, B-mode planes can be selected retrospectively anywhere in the volume. This shows that tensor velocity imaging (full 3-D volumetric vector flow imaging) can be estimated in 4-D (x, y, z, and t) using only 62 channels in receive, making 4-D volumetric imaging implementable on current scanner hardware.
引用
收藏
页码:608 / 623
页数:16
相关论文
共 50 条
  • [31] Volumetric Ultrasound Imaging with Row-Column Addressed 2-D Arrays Using Spatial Matched Filter Beamforming
    Bouzari, Hamed
    Engholm, Mathias
    Christiansen, Thomas Lehrmann
    Stuart, Matthias Bo
    Nikolov, Svetoslav Ivanov
    Thomsen, Erik Vilain
    Jensen, Jorgen Arendt
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [32] A Row-Column (RC) Addressed 2D Capacitive Micromachined Ultrasonic Transducer (CMUT) Array on a Glass Substrate: Preliminary Results
    Sanders, Jean L.
    Wu, Xun
    Adelegan, Oluwafemi J.
    Mahmud, Marzana M.
    Yamaner, F. Yalcin
    Gallippi, Caterina M.
    Oralkan, Omer
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 3468 - 3471
  • [33] 4D in vivo ultrafast ultrasound imaging using a row-column addressed matrix and coherently-compounded orthogonal plane waves
    Flesch, M.
    Pernot, M.
    Provost, J.
    Ferin, G.
    Nguyen-Dinh, A.
    Tanter, M.
    Deffieux, T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (11): : 4571 - 4588
  • [34] Fast approximate 4-D/3-D discrete radon transform for lightfield refocusing
    Marichal-Hernandez, Jose G.
    Lueke, Jonas P.
    Rosa, Fernando L.
    Rodriguez-Ramos, Jose M.
    JOURNAL OF ELECTRONIC IMAGING, 2012, 21 (02)
  • [36] Deep Beamforming for Real-Time 3-D Passive Acoustic Mapping With Row-Column-Addressed Arrays
    Lian, Yihang
    Zeng, Yi
    Zhou, Suian
    Zhu, Hui
    Li, Fei
    Cai, Xiran
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2025, 72 (02) : 226 - 237
  • [37] Real-time 3-D image reconstruction of a 24x24 row-column addressing array-from raw data to image
    Li, Chunyu
    Yang, Jiali
    Li, Xu
    Zhong, Xiaoli
    Song, Junjie
    Ding, Mingyue
    Ming Yuchi
    MEDICAL IMAGING 2016: ULTRASONIC IMAGING AND TOMOGRAPHY, 2016, 9790
  • [38] 3-D Imaging Using Row-Column-Addressed Arrays With Integrated Apodization-Part II: Transducer Fabrication and Experimental Results
    Christiansen, Thomas Lehrmann
    Rasmussen, Morten Fischer
    Bagge, Jan Peter
    Moesner, Lars Nordahl
    Jensen, Jorgen Arendt
    Thomsen, Erik Vilain
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (05) : 959 - 971
  • [39] Image-Based 4-d Reconstruction Using 3-d Change Detection
    Ulusoy, Ali Osman
    Mundy, Joseph L.
    COMPUTER VISION - ECCV 2014, PT III, 2014, 8691 : 31 - 45
  • [40] Volume Registration Based on 3-D Phase Correlation for Tumor Motion Estimation in 4-D CT
    Zhang, Xiaoyong
    Homma, Noriyasu
    Abe, Makoto
    Sugita, Norihiro
    Takai, Yoshihiro
    Narita, Yuichiro
    Yoshizawa, Makoto
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 5095 - 5098