Stability of Prototype Two-Piece Zirconia and Titanium Implants after Artificial Aging: An In Vitro Pilot Study

被引:54
|
作者
Kohal, Ralf-Joachim [1 ]
Finke, Hans Christian [1 ]
Klaus, Gerold [1 ]
机构
[1] Univ Freiburg, Sch Dent, Dept Prosthodont, Univ Clin Freiburg, D-79106 Freiburg, Germany
关键词
all-ceramic crowns; artificial mouth; fracture strength; two-piece zirconia implant; LOW-TEMPERATURE DEGRADATION; COMMERCIALLY PURE TITANIUM; ALL-CERAMIC RESTORATIONS; FRACTURE-RESISTANCE; FAILURE MODE; INTERFACE; STRENGTH; ALUMINA; TISSUES; SURFACE;
D O I
10.1111/j.1708-8208.2008.00116.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: Zirconia oral implants are a new topic in implant dentistry. So far, no data are available on the biomechanical behavior of two-piece zirconia implants. Therefore, the purpose of this pilot investigation was to test in vitro the fracture strength of two-piece cylindrical zirconia implants after aging in a chewing simulator. Materials and Methods: This laboratory in vitro investigation comprised three different treatment groups. Each group consisted of 16 specimens. In group 1, two-piece zirconia implants were restored with zirconia crowns (zirconia copings veneered with Triceram (R); Esprident, Ispringen, Germany), and in group 2 zirconia implants received Empress (R) 2 single crowns (Ivoclar Vivadent AG, Schaan, Liechtenstein). The implants, including the abutments, in the two zirconia groups were identical. In group 3, similar titanium implants were reconstructed with porcelain-fused-to-metal crowns. Eight samples of each group were submitted to artificial aging with a long-term load test in the artificial mouth (chewing simulator). Subsequently, all not artificially aged samples and all artificially aged samples that survived the long-term loading of each group were submitted to a fracture strength test in a universal testing machine. For the pairwise comparisons in the different test groups with or without artificial loading and between the different groups at a given artificial loading condition, the Wilcoxon rank-sum test for independent samples was used. The significance level was set at 5%. Results: One sample of group 1 (veneer fracture), none of group 2, and six samples of group 3 (implant abutment screw fractures) failed while exposed to the artificial mouth. The values for the fracture strength after artificial loading with 1.2 million cycles for group 1 were between 45 and 377 N (mean: 275.7 N), in group 2 between 240 and 314 N (mean: 280.7 N), and in the titanium group between 45 and 582 N (mean: 165.7 N). The fracture strength results without artificial load for group 1 amounted to between 270 and 393 N (mean: 325.1 N), for group 2 between 235 and 321 N (mean: 281.8 N), and between 474 and 765 N (mean: 595.2 N) for the titanium group. The failure mode during the fracture testing in the zirconia implant groups was a fracture of the implant head and a bending/fracture of the abutment screw in the titanium group. Conclusions: Within the limits of this pilot investigation, the biomechanical stability of all tested prototype implant groups seems to be - compared with the possibly exerted occlusal forces - borderline for clinical use. A high number of failures occurred already during the artificial loading in the titanium group at the abutment screw level. The zirconia implant groups showed irreparable implant head fractures at relatively low fracture loads. Therefore, the clinical use of the presented prototype implants has to be questioned.
引用
收藏
页码:323 / 329
页数:7
相关论文
共 50 条
  • [41] Early bone formation around immediately placed two-piece tissue-level zirconia implants with a modified surface: an experimental study in the miniature pig mandible
    Glauser, Roland
    Schupbach, Peter
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2022, 8 (01)
  • [42] Early bone formation around immediately placed two-piece tissue-level zirconia implants with a modified surface: an experimental study in the miniature pig mandible
    Roland Glauser
    Peter Schupbach
    International Journal of Implant Dentistry, 8
  • [43] Validation offinite-element simulations with synchrotron radiography - A descriptive study of micromechanics in two-piece dental implants
    Wiest, Wolfram
    Rack, Alexander
    Zabler, Simon
    Schaer, Alex
    Swain, Michael
    Nelson, Katja
    HELIYON, 2018, 4 (02):
  • [44] Minimally Invasive Treatment of Lateral Incisors with Guided One-Piece or Two-Piece Titanium-Made Narrow Diameter Implants: A Retrospective Comparative Study with Up to Two Years Follow-Up
    Zadrozny, Lukasz
    Gorski, Bartlomiej
    Baldoni, Edoardo
    Lumbau, Aurea Immacolata
    Meloni, Silvio Mario
    Pisano, Milena
    Tallarico, Marco
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (11)
  • [45] Comparative analysis of stress distribution in one-piece and two-piece implants with narrow and extra-narrow diameters: A finite element study
    Barbosa, Fabricia Teixeira
    Silveira Zanatta, Luiz Carlos
    Rendohl, Edelcio de Souza
    Gehrke, Sergio Alexandre
    PLOS ONE, 2021, 16 (02):
  • [46] Assessing the Impact of Various Decontamination Instruments on Titanium and Zirconia Dental Implants: An In Vitro Study
    Vierling, Louisa
    Liu, Chun Ching
    Wiedemeier, Daniel
    Gubler, Andrea
    Schmidlin, Patrick R.
    DENTISTRY JOURNAL, 2024, 12 (05)
  • [47] Osseointegration of one-piece zirconia implants compared with a titanium implant of identical design: a histomorphometric study in the dog
    Koch, F. P.
    Weng, D.
    Kraemer, S.
    Biesterfeld, S.
    Jahn-Eimermacher, A.
    Wagner, W.
    CLINICAL ORAL IMPLANTS RESEARCH, 2010, 21 (03) : 350 - 356
  • [48] Soft Tissue Healing at One-Piece Zirconia Implants Compared to Titanium and PEEK Implants of Identical Design: A Histomorphometric Study in the Dog
    Koch, Felix Peter
    Weng, Dietmar
    Kraemer, Sonja
    Wagner, Wilfried
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2013, 33 (05) : 669 - U127
  • [49] In vitro synchrotron-based radiography of micro-gap formation at the implant-abutment interface of two-piece dental implants
    Rack, A.
    Rack, T.
    Stiller, M.
    Riesemeier, H.
    Zabler, S.
    Nelson, K.
    JOURNAL OF SYNCHROTRON RADIATION, 2010, 17 : 289 - 294
  • [50] Biomechanical evaluation of one-piece and two-piece small-diameter dental implants: In-vitro experimental and three-dimensional finite element analyses
    Wu, Aaron Yu-Jen
    Hsu, Jui-Ting
    Chee, Winston
    Lin, Yun-Te
    Fuh, Lih-Jyh
    Huang, Heng-Li
    JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2016, 115 (09) : 794 - 800